The I
and I
parameters shown apply
0
1
to the Current Arb type.
The V
, V
, P
, P
, R
, and R
0
1
0
1
0
parameters apply to the respective
Voltage, Power and Resistance Arb
types.
Parameter
Start Setting (I 0 , V 0 , P 0 , or R 0 )
End Setting (I 1 , V 1 , P 1 , or R 1 )
Delay (T 0 )
Time (T 1 )
Time Constant (T c )
Step 3. Configure the final steps common to all Arbs.
Refer to
Configuring Parameters Common to all Arbs
From the remote interface:
Select the Arb Type and Shape:
ARB:FUNC:TYPE VOLT,(@1)
ARB:FUNC:SHAP EXP,(@1)
Enter the values for the level before and after a voltage exponential:
ARB:VOLT:EXP:STAR 0,(@1)
ARB:VOLT:EXP:END 5,(@1)
Enter the time before the exponential, the exponential time, and the time constant of the exponential:
ARB:VOLT:EXP:STAR:TIM 0.25,(@1)
ARB:VOLT:EXP:TIM 0.75,(@1)
ARB:VOLT:EXP:TCON 0.1,(@1)
Configuring Constant-Dwell Arbs
Constant-Dwell (CD) Arbs are a unique type that have some useful differences from other Arbs. CD
Arbs are not limited to 511 points, they have up to 65,535 points. Unlike other Arbs, they do not have
separate dwell values for each point - a single dwell value applies to ALL points. Also, the minimum
dwell time of a CD Arb is 10 .24 microseconds, instead of the 1 microsecond resolution of other Arbs.
Keysight N6705C User's Guide
1
Description
The setting before the waveform.
The end setting of the waveform.
The delay after the trigger is received but before the waveform starts.
Time for the amplitude to go from the start setting to the end setting.
The time constant of the curve.
3 Using the Source/Load Functions
at the end of this section.
113