Meter View
Press Meter
View
This key
toggles
between
multiple and
single output
views.
1. Output
Identifies the output. When an output is selected, the background becomes highlighted. The selected output is displayed in an
Identifier
enlarged format in single output view.
2. Output
Off: the output is off
Status
CV: the output is in constant voltage mode
CC: the output is in constant current mode
Unr: the output is unregulated
CP+, CP–: a positive or negative power limit
CL+, CL–: a positive or negative current limit
VL+, VL–: a positive or negative voltage limit
OV: over-voltage protection tripped
3. Output
Displays the actual output voltage and current. Displays power in single output view.
Meter
4. Output
Displays the present output voltage and current settings. Turn the front panel voltage or current knob to adjust these settings.
Setting
Can also be changed using the numeric keypad.
5. Remote
Error – an error has occurred (press the Menu key, select Utilities, then Error Log)
Status
LAN – the LAN is connected and has been configured
IO – there is activity on one of the remote interfaces
6. Model
Identifies the model number of the power module connected to this output.
Number
7. Arb, Delay,
Displays the Arb waveform that is presently configured for this output. If no Arb is configured, no waveform will be displayed.
& Slew Rate
Also displays the Output On and Output Off delay settings as well as the slew rate setting.
8. Polarity
Indicates that the output and sense polarities are reversed.
Reverse
9. Ratings &
Displays the maximum voltage and current ratings of the output. Also displays the present over-voltage protection setting
Protection
and whether over-current protection is on or off.
10. Other
Displays the actual voltage, current, and status of the other outputs.
Outputs
Keysight N6705C User's Guide
Multiple Output View
Single Output View
OV–: negative voltage protection tripped
OC: over-current protection tripped
OT: over-temperature protection tripped
PF: a power-fail condition occurred
Inh: an external inhibit signal is received
Osc: the oscillation protection tripped
Prot: a coupled output protect occurred
SH: the load input is shorted (N679xA)
UVI: an under-voltage inhibit occurred (N679xA)
1 Quick Reference
19