6 Advanced Source, Measurement, and Control
From the front panel:
To configure the trigger input function, press the Menu key, scroll down and select the Utilities item,
then select Digital I/O. Then press Enter.
Select pin 6 from the Pin dropdown list. Then select the On Couple function for the pin the from the
Function dropdown list.
Select pin 7 from the Pin dropdown list. Then select the Off Couple function for the pin the from the
Function dropdown list.
Repeat these steps for mainframes #2 and #3.
From the remote interface:
To configure pin 6 of mainframe #1 as the ON control:
DIG:PIN6:FUNC ONC
To configure pin 7 of mainframe #1 as the OFF control:
DIG:PIN7:FUNC OFFC
Repeat these commands for mainframes #2 and #3.
Operation
Once configured and enabled, turning the output on or off on any coupled output will cause all coupled
outputs on all configured mainframes to turn on or off according to their user-programmed delays.
This applies to the front panel On and Off keys, the Web server, and to SCPI commands.
Turning the outputs on or off using the front panel All Outputs On and Off keys will cause all coupled
outputs as well as non-coupled outputs on that mainframe to turn on or off.
Keysight N6705C User's Guide
233