The I
and I
parameters shown apply
0
1
to the Current Arb type.
The V
, V
, P
, P
, R
, and R
0
1
0
1
0
parameters apply to the respective
Voltage, Power and Resistance Arb
types.
Parameter
Description
Amplitude (I 0 , V 0 , P 0 , or
The amplitude or peak value.
R 0 )
Offset (I 1 , V 1 , P 1 , or R 1 )
The offset from zero. For power modules that do not generate negative values, the offset must be ≥ Amplitude.
Frequency (f)
The frequency of the sine wave.
Step 3. Configure the final steps common to all Arbs.
Refer to
Configuring Parameters Common to all Arbs
From the remote interface:
Select the Arb Type and Shape:
ARB:FUNC:TYPE VOLT,(@1)
ARB:FUNC:SHAP SIN,(@1)
Enter the values for the voltage amplitude, offset from zero, and frequency:
ARB:VOLT:SIN:AMPL 5,(@1)
ARB:VOLT:SIN:OFFS 5,(@1)
ARB:VOLT:SIN:FREQ 1,(@1)
Configuring Pulse Arbs
Step 1. Select the Arb Type and Shape. Power and resistance types apply to
Press the Arb key twice or press the Arb key and then Properties to access the Arb Selection window.
In the Arb Type dropdown, select the Arb type. Then select Pulse.
Keysight N6705C User's Guide
1
3 Using the Source/Load Functions
at the end of this section.
:
109