Voltage - Lets you enter a voltage value with the numeric keys. You can use the Voltage knob to
adjust the value in this field. Press Enter to enter the value.
Range - There is only one voltage range.
Current Limit - You can specify a current limit that limits the input current when in voltage priority
mode. It is programmable up to the rated current, with an over-range of 2%.
Refer to
Current Priority Mode
Resistance Priority Mode
In this mode, the load will sink a current linearly proportional to the voltage in accordance with the
programmed resistance value.
Resistance - Lets you enter a resistance value with the numeric keys. Press Enter to enter the value.
Range - Lets you select from three overlapping resistance ranges. The lower ranges provide better
resolution at low resistance settings. The following ranges can be selected:
High resistance range
Medium resistance range
Low resistance range
Current Limit - the current limit is fixed at the specified setting.
When changing ranges, the load input will be turned off, then back on again.
Refer to
Current Priority Mode
Power Priority Mode
In this mode, the module will maintain the output power at the specified programmed power level.
Note that the module has an independent power limit loop that regulates the output power at the limit
setting - either 100 W or 200 W, with an over-range of 10%.
Keysight N6705C User's Guide
for descriptions of the Under Voltage Inhibit and Enable short controls.
N6791A
from 30 Ω to 8 kΩ
from 2 Ω to 100 Ω
from 0.08 Ω to 3 Ω
for descriptions of the Under Voltage Inhibit and Enable short controls.
3 Using the Source/Load Functions
N6792A
from 15 Ω to 8 kΩ
from 2 Ω to 100 Ω
from 0.04 Ω to 3 Ω
95