3 Using the Source/Load Functions
Configuring Step Arbs
Step 1. Select the Arb Type and Shape. Power and resistance types apply to
Press the Arb key twice or press the Arb key and then Properties to access the Arb Selection window.
In the Arb Type dropdown, select the Arb type. Then select Step.
Step 2. Configure the Step Properties.
Press the Properties key or select the Arb Properties button. Then program the following parameters:
The I
and I
parameters shown apply
0
1
to the Current Arb type.
The V
, V
, P
, P
, R
, and R
0
1
0
1
0
parameters apply to the respective
Voltage, Power and Resistance Arb
types.
Parameter
Start Setting (I 0 , V 0 , P 0 , or R 0 )
End Setting (I 1 , V 1 , P 1 , or R 1 )
Delay (T 0 )
End Time (T 1 )
Step 3. Configure the final steps common to all Arbs.
Refer to
Configuring Parameters Common to all Arbs
From the remote interface:
Select the Arb Type and Shape:
ARB:FUNC:TYPE VOLT,(@1)
ARB:FUNC:SHAP STEP,(@1)
102
1
Description
The setting before the step.
The setting after the step.
The delay after the trigger is received but before the step occurs.
The time the output remains at the end setting after the step occurs.
at the end of this section.
Keysight N6705C User's Guide
: