To set the front panel sense terminals to remote sensing:
VOLT:SENS:SOUR EXT,(@1)
To query the setting of the front panel sense terminals (INT=local sensing; EXT=remote sensing):
VOLT:SENS:SOUR?,(@1)
To set the output turn-off mode to the high impedance setting:
OUTP:TMOD HIGHZ,(@1)
To set the output voltage bandwidth of output 1 to the default setting:
VOLT:BWID LOW,(@1)
Programming N679xA Load Settings
Use the Source Settings window to access the specialized operating modes of the
Keysight N679xA Load modules
to as "Outputs" throughout this document.
Priority Modes
The Operating In dropdown list lets you select the priority mode of the Keysight N679xA Load module.
Use the navigation keys to select one of the four priority modes. Then press Enter.
When switching between Priority modes, the load input is turned off and the Load
settings revert to their Power-on or RST values.
Current Priority Mode
In this mode, the Load module will sink a current in accordance with the programmed value regardless
of the input voltage. Note that a programmable voltage limit is not available. If the DUT imposes a
voltage above the voltage rating of the module, the overvoltage protection will trip.
Keysight N6705C User's Guide
3 Using the Source/Load Functions
. Load module input terminals are referred