4 Using the Measurement Functions
From the front panel:
To view the meter properties, press the Meter View key, then press Properties.
Note that the Meter Properties dialog only displays the voltage or current ranges that have been set in
the
Source Settings
dialogs. Also, Load modules have a High and a Low current meter ranges;
however, these cannot be selected in the Meter properties window. The reason for this is because
measurement ranges are coupled with the priority mode range settings.
The voltage measurement range is set to 61.2 V for all priority modes.
For the Voltage priority mode, the current measurement range is set to High.
For the Power priority low range, the current measurement range is set to Low.
For the Power priority high range, the current measurement range is set to High.
For the Resistance priority high range, the current range is set to Low.
For the Resistance priority medium and low ranges, the current range is set to High.
When changing priority modes, all ranges will be reset to the default (*RST) settings. Measurement
ranges are not auto-ranging.
Line Frequency – Specify the ac mains frequency (50 Hz or 60 Hz).
NPLC - Specify the number of power line cycles (NPLC) that the measurement will span. Increase the
NPLC for improved accuracy and noise reduction on low current and voltage measurements.
Duration – Indicates the duration of the measurement based on the Frequency and NPLC settings.
From the remote interface:
To trigger and return a measurement:
MEAS:CURR? (@1)
MEAS:VOLT? (@1)
To return a previous measurement:
FETC:CURR? (@1)
FETC:VOLT? (@1)