Enter the values for the level before and after a voltage step:
ARB:VOLT:STEP:STAR 0,(@1)
ARB:VOLT:STEP:END 5,(@1)
Enter the delay time before the step:
ARB:VOLT:STEP:STAR:TIM 0.01,(@1)
Configuring Ramp Arbs
Step 1. Select the Arb Type and Shape. Power and resistance types apply to
Press the Arb key twice or press the Arb key and then Properties to access the Arb Selection window.
In the Arb Type dropdown, select the Arb type. Then select Ramp.
Step 2. Configure the Ramp Properties.
Press the Properties key or select the Arb Properties button. Then program the following parameters:
The I
and I
parameters shown apply
0
1
to the Current Arb type.
The V
, V
, P
, P
, R
, and R
0
1
0
1
0
parameters apply to the respective
Voltage, Power and Resistance Arb
types.
Parameter
Start Setting (I 0 , V 0 , P 0 , or R 0 )
End Setting (I 1 , V 1 , P 1 , or R 1 )
Delay (T 0 )
Ramp Time (T 1 )
End Time (T 2 )
Keysight N6705C User's Guide
1
Description
The setting before the ramp.
The setting after the ramp.
The delay after the trigger is received but before the ramp starts.
The time that the output ramps up.
The time the output remains at the end setting after the ramp completes.
3 Using the Source/Load Functions
: