Pneumatic Fowler Adjustment - Stryker 1089 Manuel D'entretien

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 39

PNEUMATIC FOWLER ADJUSTMENT

Tools Required:
• (2) 3/8" Combination Wrenches
Procedure:
1 .
Apply the brakes .
2 .
Raise the litter to its highest height .
3 .
Raise the fowler to 70°
4 .
Using (2) 3/8" combination wrenches, loosen the two nuts on the adjustment screw (B) . If the fowler is not holding
its position, turn the adjustment screw up and away from the release pin on the pneumatic cylinder . If the fowler is
not releasing, turn the adjustment screw down, toward the release pin on the pneumatic cylinder . Tighten the two
nuts . Repeat for the other side of the fowler .
5 .
To verify that both sides of the fowler are releasing equally, raise the fowler to the full up position . Using two hands,
pull down on both sides of the fowler, squeeze both release handles and verify that both pneumatic cylinders
release equally . If they do not, repeat step 4 .
6 .
Verify proper operation before returning the unit to service .
A
B
www.stryker.com
Service Information
C
Figure 12
1089-009-006 REV G
Return To Table of Contents
1-29
English

Publicité

Table des Matières
loading