When a Paddle Lead is dropped into the Lead Placement panel, it is assigned to its own Lead Group, if
an empty Lead Group is available. Additional Leads cannot be added to Lead Groups that are already
occupied by a paddle Lead.
1. Drag and drop the desired Lead onto the Lead Placement panel. The first Lead dropped will
be automatically assigned to Group 1.
2. Subsequent percutaneous Leads dropped onto the Lead Placement panel will be added to
Group 1. When a paddle Lead is dropped, it will be assigned to its own Lead Group if an
empty Lead Group is available.
3. Lead Group assignments can be modified by selecting the desired Lead, then selecting a
different Lead Group radio button on the right of the screen as long as the target lead group
is empty.
4. To add a new Lead to a Lead Group that already has a percutaneous Lead assigned to it, a
new Lead must be added to the target group from the Lead Selection panel on the left.
IMPORTANT: Only Leads that are in the same Lead Group will interact to program a stimulation Area,
except when programming in Manual mode. Refer to "Using Manual Mode".
Configuring the CP and Stimulator
Bionic Navigator™ 3D 4.0 Programming Manual
92395591-02 19 of 1595