Bionic Navigator™ 3D 4.0 Programming Manual
3. Select the
Stimulator and CP. The Connection Status Indicator displays green when a valid connection
is made.
Note: The Stimulator will remain in pairing mode and available for connection with the CP for 2
minutes. After that time, the Stimulator is no longer available for connection and the Connect button will
be greyed out. Put the Stimulator back in pairing mode, click the Refresh button on the Configuration
Screen, and select the Connect button to create a programming connection between the Stimulator and
CP.
Note: After connecting the desired Stimulator and CP, the Remote Control or the ON/OFF button on the
External Trial Stimulator can be used to modify the Stimulator settings. Any subsequent change to the
Stimulator from the CP will transfer the settings displayed on the CP to the Stimulator.
Caution: If the Remote Control is used to turn off stimulation therapy during programming, set
the Therapy ON/OFF button on the CP to OFF before resuming programming. This will ensure
that the CP is synchronized with the therapy status of the Stimulator.
1
3
Figure 7: Connect Screen
Bionic Navigator™ 3D 4.0 Programming Manual
92395591-02 8 of 1595
button to create a programming connection between the desired
4
2
6
7
5