The image displayed on the LeadSync pop-up window may display a different relative Lead orientation
than what was configured in the Lead Placement panel. Between each pair of Leads is a symbol
indicating the correlation of the LeadSync measurement to the displayed Lead offset, as the LeadSync
measurement is dependent on the geometry of the actual Lead placement. There are three symbols
that describe LeadSync measurement results:
Correlation of LeadSync Measurement to the displayed Lead offset is high. See Figure 14.
Correlation of LeadSync Measurement to the displayed Lead offset is low. See Figure 15.
LeadSync is unable to provide meaningful result and will display the Lead configuration
that was manually specified by user. See Figure 16.
Figure 14: LeadSync Screen with Check Mark - A check mark in a green circle indicates that correlation of
LeadSync Measurement to the displayed Lead offset is high. If the displayed LeadSync measurement is applied, the
Leads will be re-aligned with respect to the center of the Lead Group and displayed in the Lead Placement panel.
Configuring the CP and Stimulator
Bionic Navigator™ 3D 4.0 Programming Manual
92395591-02 21 of 1595