Scale Membrane Replacement; Accessory Bracket Cover Replacement; (Scale Option) - Stryker CUB Pediatric Crib Manuel De Maintenance

Masquer les pouces Voir aussi pour CUB Pediatric Crib:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

SCALE MEMBRANE REPLACEMENT

TOOLS REQUIRED:
• Small Regular Screwdriver
• Isopropyl Alcohol
PROCEDURE:
1.
Slide a small regular screwdriver under the membrane (A) and lift
to remove it. Use caution when inserting the screwdriver under
the membrane to avoid scratching the plastic cover.
2.
Clean any glue residue using isopropyl alcohol.
3.
Remove the protective film on the backside of the new membrane
and place it on the cover. Ensure proper positioning of the
membrane before permanently adhering it to the cover.
4.
Verify proper operation of the scale controls before returning the
stretcher to service.

ACCESSORY BRACKET COVER REPLACEMENT

(SCALE OPTION)

TOOLS REQUIRED:
• #2 Phillips Screwdriver
• 9/16 in. Combination Wrench
• Diagonal Pliers
• ESD System
PROCEDURE:
1.
Lower the foot rail completely. If it is a fixed rail, see the "Fixed
Endrail Installation" procedure
temporarily lower the rail.
2.
Raise the siderails to the upper position.
3.
Lift the foot section and fold it towards the head end of the stretcher.
4.
Using a #2 Phillips screwdriver, remove the seven screws holding the foot cover plate to the litter frame. Remove
the plate (see Figure 14,
5.
Using a 9/16 in. combination wrench, remove the bolt (J), washer (H), and spacer (G) holding the fixed top socket
(B) to the accessory bracket.
6.
Properly ground yourself (see "Static Discharge Precautions",
7.
Using diagonal pliers, cut the cable ties holding the cable to the lower plate and the accessory bracket.
8.
Lift the accessory bracket cover while slowly pulling on the cables and turn it towards you until the inner side of
the cover is accessible.
9.
Using a #2 Phillips screwdriver, remove the five screws (D) holding the control board (C) to the cover, and remove
the control board.
10. Reverse the above steps to install the new accessory bracket cover.
Note
The new accessory bracket cover includes the membrane and all applicable labels.
11. Verify proper operation of the scale controls before returning the stretcher to service.
Return To Table of Contents
1-42
Service
(see page 1-16)
for reference to
page
1-44).
1900-009-002 REV B
B
D
F
G
H
J
Figure 13
page
1-15).
A
R
C
D
T
S
E
K
L
M
N
P
www.stryker.com