HEAD SECTION REPLACEMENT

TOOLS REQUIRED:
• (2) 1/2 in. Combination Wrenches
G
PROCEDURE:
1.
Raise the head section to the highest position, lower the rails completely and apply the brakes.
2.
Using two 1/2 in. combination wrenches, remove the four locknuts (C), four shoulder spacers (H), four washers (J)
and four bolts (B) holding the head section to the two coupling bars. Remove the head section.
3.
Reverse the above steps to install the new head section.
4.
Verify proper operation of the head section before returning the stretcher to service.