Assembly of the Pneumatic Handpiece

The Pneumatic Handpiece is provided non sterile in two pieces, the
handpiece and the air supply tubing that connects it to the control unit.
They must be connected as shown before the handpiece is sterilized
per the recommendations.
The Pneumatic Probes are single use and are packaged sterile. Once
the handpiece has been sterilized and the air supply tubing connected,
attach the Pneumatic Probe in a sterile manner to the front end of the
handpiece with slight axial pressure and a ¼ clockwise turn until it
'clicks' into place.
Once the handpiece is assembled, remove the protective sterilization
cap and connect the plug of the Pneumatic Handpiece into the control
unit. There are no dots to align so just push the connector into the
control unit. The plug must be completely dry before attachment.
4
FB-412_US ed 2008-04 User man pn3 BSC.indd 4
STERILE
Function testing
handpiece.
Verify that probe impacts are generated in the Pneumatic Handpiece
and transmitted to the probe. Hand feel the shock waves by holding
the Pneumatic Probe, however never touch the probe tip for function
testing as this may hurt.
Pneumatic impulses are released by pressing the pneumatic foot pedal
and are controlled by the control unit.
A function test should be performed prior to using the