Reprocessing
- Rinse the devices with water whose quality is defined in the GCDS until the
device is no longer slippery to the touch. Repeat these steps until the device
is visibly clean.
2. Machine cleaning and disinfection of the device:
- Place the instrument in the Precision Targeting System Insert Tray and con-
tinue as instructed in the GCDS Section 4.3 Machine Cleaning and Disinfec-
tion with a Washer-Disinfector.
WARNING
- ALWAYS ensure that the central knob of the Precision Targeting Arm is TIGHT-
ENED during machine-cleaning, disinfection, and when exposed to running
water.
- DO NOT immerse the Precision Targeting Arm in liquids.
3. Sterilization
WARNING
- ALWAYS ensure that the central knob of the Precision Targeting Arm is LOOS-
ENED during sterilization