1. Press the push button (a).
2. Insert the Drill Sleeve into the
Tracker Sleeve (b).
3. Attach the Twist Drill Depth
Stopper on the Twist Drill according
to the drill depth determined by the
user. When using the Elite
Attachment attach the Elite Drill
Depth Stopper accordingly.
4. Tighten the knob of the Drill Depth
Stopper.
5. Perform drilling.
6. Press the push button (a).
7. Remove the Drill Sleeve after
drilling through the device (b).
Stryker
Figure 15: Inserting the Drill Sleeve.
Figure 16: Attaching the Twist Drill Depth Stopper.
Figure 17: Removing the Drill Sleeve.
Intraoperative use
b
a
b
a
15