Preparation for use
4.4 Preparing the Precision Targeting System
WARNING
MAINTAIN SYSTEM STABILITY
- ALWAYS ensure proper assembly of the quick connect feature by listening for 
the audible latching of the locking mechanism.
CAUTION
- ALWAYS be careful while attaching the Precision Targeting Guide to the
Precision Targeting Arm and while attaching the Tracker Sleeve to the 
Precision Targeting Guide to avoid a pinch point.
  ALWAYS ensure that the drape is not caught between the starburst
connection of the skull clamp, Arm Connector Component and Precision
Targeting Arm during assembly.
1. Attach the Arm Connector
Component to the skull clamp. 
2. Tighten the starburst knob of the 
Arm Connector Component.
10
Figure 7: Attaching the Arm Connector Component.
Figure 8: Tighten the starburst knob.