4.2 Preparing the Tracker Sleeve
1. Prepare the Tracker Sleeve by 
attaching the 4 Navigation Spheres 
to the tracker post.
4.3 Validating the Tracker Sleeve
  For instructions on how to use the Validation Tip for the validation of the 
Tracker Sleeve in conjunction with the software application, refer to the user 
manual supplied with the software application.
1. Press the push button of the Tracker 
Sleeve (a).
2. Insert the Validation Tip into the 
Tracker Sleeve (b).
→ The Validation Tip needs to be fully 
inserted into the Tracker Sleeve.
3. Place the Validation Tip in the 
Calibration Body's cone. Ensure the 
Validation Tip touches the bottom of 
the cone.
Stryker
Figure 4: Attaching the Navigation Spheres to the tracker post.
a
Figure 5: Inserting the Validation Tip.
Figure 6: Validation Tip in the Calibration Body's cone.
Preparation for use
b
9