Reprocessing
6
Reprocessing
6.1 Disassembly for reprocessing
WARNING
- Before disassembly, ALWAYS ensure to remove all components like the Biopsy 
Needle from within the brain.
- During disassembly, ALWAYS hold the Precision Targeting Guide with one hand 
before loosening the central knob of the Precision Targeting Arm with the other 
hand and then move the arm away from the patient.
- Clean and sterilize the instrument before first and every use. Refer to the Guide 
for Cleaning and Steam-Based Sterilization.
1. Remove the sleeve by pressing the Tracker Sleeve push button.
2. Remove the Tracker Sleeve from the Precision Targeting Guide by pressing the 
Precision Targeting Guide push button.
3.   R emove the Navigation Spheres from the Tracker Sleeve.
4. Remove the Precision Targeting Guide from the Precision Targeting Arm by 
pressing the quick connect button.
5. Remove the Precision Targeting Arm from the Arm Connector Component by 
loosening the starburst knob.
6. Remove the Arm Connector Component from the skull clamp.
7. Remove the Drill Depth Stopper from the Twist Drill/Elite attachment by 
unlocking the Drill Depth Stopper knob.
8. Where appropriate, put all the instruments back into the Precision Targeting 
System Insert Tray at their designated location.
6.2 Reprocessing instructions
This section contains only product-specific information. For general reprocessing 
instructions, safety notes, and reprocessing equipment, refer to the G
C
, d
leaninG
isinfeCtion
  Where instructions between this document and the GCDS differ, follow this 
document.
The cleaning group of each instrument according to the GCDS is:
Part
Precision Targeting Guide
Tracker Sleeve Body
Needle Sleeve
18
,
s
-B
s
and
team
ased
(GCds, td6000005750).