Part
Elite Drill Sleeve
Twist Drill Sleeve
Elite Drill Depth Stopper
Twist Drill Depth Stopper
Precision Targeting Arm
(refer to special reprocessing instructions in this sec-
tion)
Validation Tip
Arm Connector Component
These instruments can be cleaned and sterilized in the Precision Targeting
System Insert Tray (REF 6000-810-000). Refer to the tray IFU for
handling of the tray and placement of the instruments.
WARNING
- ALWAYS open all four locking knob and all four adjustment knobs of the
Precision Targeting Guide such that the lead screws of the linear adjustment
knob and threads of the rotational adjustment knobs are in maximum exposed
state during reprocessing. Disassemble all components with removable parts
during reprocessing.
Reprocessing instructions for the Precision Targeting Arm
Follow the GCds with the following exceptions:
The Precision Targeting Arm can only be cleaned in combination of manual
pre-cleaning and machine washing. Manual pre-cleaning only is not sufficient.
1. Manual pre-cleaning of the device:
- Wet the wipes in the cleaning solution. Dip the wipes into the solution until
saturated. After wetting the wipe, the excess liquid should be squeezed out.
The wipes should be wet but not dripping.
- Remove heavy residues from the devices with the wipes.
- Keep the central knob of Precision Targeting Arm LOOSENED while using
wet wipes and clean around articulated parts at least 3 times.
- TIGHTEN the central knob of the Precision Targeting Arm.
- Remove all residues with the cleaning tools.
Stryker
Reprocessing
Cleaning group
3
3
4
4
4
1
4
19