Implantable Pulse Generator

Max Current Amplitude per Electrode vs Impedance

For all Leads other than the 4x8 Surgical Lead
Maximum Amplitude Based on Impedance and Pulse Width
Programmable up to 25.5 mA
25.50
Programmable up to 21.1 mA
20.40
Programmable up to 15.8 mA
15.30
Programmable up to 12.7 mA
10.20
5.10
0.00
300
400
500
For the 4x8 Surgical Lead
Maximum Amplitude Based on Impedance and Pulse Width
Programmable up to 25.5 mA
25.50
20.40
Programmable up to 17.6 mA
15.30
Programmable up to 13.2 mA
Programmable up to 10.6 mA
10.20
5.10
0.00
300
400
500
Implantable Pulse Generator
92177206-07
2 of 53
600
700
800
900
Impedance (Ω)
600
700
800
900
Impedance (Ω)
Pulse Width:
PW = 20 us
20 µsec
PW = 100 us
100 µsec
PW = 200 us
200 µsec
PW = 300 us
300 µsec
PW = 400 us
400 µsec
PW = 600 us
600 µsec
PW = 800 us
800 µsec
PW = 1000 us
1000 µsec
1000
1100
1200
Pulse Width:
PW = 20 us
20 µsec
PW = 100 us
100 µsec
PW = 200 us
200 µsec
PW = 300 us
300 µsec
PW = 400 us
400 µsec
PW = 600 us
600 µsec
PW = 800 us
800 µsec
PW = 1000 us
1000 µsec
1000
1100
1200