Connecting the Stimulator

Connect Left DBS Extension to port 1-L
1. Fully insert the male end of the DBS Extension into the Stimulator until it stops.
a. Insert the header plug to verify no Setscrews obstruct the socket.
b. Wipe the DBS Extension contacts before inserting.
c. Insert the DBS Extensions into the header. When fully inserted, the tip of the
DBS Extension will slide into the back of the port and the retention sleeve on
the DBS Extension will be located under the Setscrew.
Physician Manual
91098825-07 Rev A 30 of 333
Dual DBS Lead Connection
Connect Right DBS Extension to port 2-R
Left
Right
1-L
2-R