Télécharger Imprimer la page

Stryker Power-PRO IT Cot Mode D'emploi page 52

Civière

Publicité

Les langues disponibles

Les langues disponibles

C C l l e e a a n n i i n n g g t t h h e e c c h h a a r r g g e e r r
W W A A R R N N I I N N G G
• Always wear insulated rubber gloves, in addition to personal protective equipment, when cleaning the S S M M R R T T Pak to
reduce the risk of injury.
• Always disconnect the S S M M R R T T charger from the wall outlet before cleaning to avoid the risk of electrical hazards.
• Do not spray fluid directly onto the S S M M R R T T charger.
• Do not power wash the S S M M R R T T charger.
• Do not use solvents, lubricants, or other chemicals to clean the S S M M R R T T charger unless otherwise directed.
• Do not immerse the S S M M R R T T charger in water or allow water to collect on top of the S S M M R R T T charger to avoid the risk of
electric shock.
To clean the S S M M R R T T charger:
1. Disconnect the S S M M R R T T charger from the wall outlet to avoid electrical hazards during cleaning.
2. Wipe surfaces of the S S M M R R T T charger with a soft cloth dampened with a non-abrasive, hospital disinfectant.
3. Wipe with a cloth moistened with clean water to remove any cleaning chemicals or residue.
4. Dry thoroughly before placing the S S M M R R T T charger back into service.
C C l l e e a a n n i i n n g g t t h h e e b b a a t t t t e e r r y y
The S S M M R R T T Pak is designed to be power washable. The preferred method for power washing the S S M M R R T T Pak is with the
standard hospital surgical cart washer or hand held wand unit.
W W A A R R N N I I N N G G
• Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion. Power
washing equipment may aerate contamination.
• Always wear insulated rubber gloves, in addition to personal protective equipment, when cleaning the S S M M R R T T Pak to
reduce the risk of injury.
• Always use only non-conductive materials to wipe the S S M M R R T T Pak.
• Always avoid excessive water exposure to the S S M M R R T T Pak terminals.
• Always refer to the disinfectant's Material Safety Data Sheet (MSDS) to verify the pH range. Disinfectants with pH levels
higher than 10.5 may cause the S S M M R R T T Pak housing material to crack.
• Do not directly handle or make contact with the S S M M R R T T Pak terminals while cleaning to avoid the risk of injury.
• Do not immerse the S S M M R R T T Pak in liquid to reduce the risk of electric shock.
• Do not use solvents, lubricants, or other chemicals to clean the S S M M R R T T Pak unless otherwise directed.
C C A A U U T T I I O O N N
• Do not steam clean or ultrasonically clean the S S M M R R T T Pak.
• Do not exceed 180 °F (82 °C) as the maximum water temperature.
• Do not exceed 240 °F (115 °C) as the maximum air dry temperature (cart washers).
• Do not exceed 1500 psi (103.4 bar) as the maximum water pressure. If you use a hand held wand to wash the product,
keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
To clean the S S M M R R T T Pak:
1. Remove the S S M M R R T T Pak from the P P o o w w e e r r - - P P R R O O cot or S S M M R R T T charger.
2. Inspect the S S M M R R T T Pak housing and terminal area for any cracks or damage.
3. Clean the S S M M R R T T Pak with a hospital grade disinfectant with a pH range of 6.5 to 10.5.
4. Rinse the S S M M R R T T Pak thoroughly with clean water to remove any cleaning chemical or residue. Position the S S M M R R T T Pak
to avoid water from pooling near the terminals.
5. Dry the S S M M R R T T Pak thoroughly before insert the S S M M R R T T Pak into a P P o o w w e e r r - - P P R R O O cot or S S M M R R T T charger.
EN
48
6516-209-005 Rev AA.0

Publicité

loading

Ce manuel est également adapté pour:

6516