• Always mount the S S M M R R T T charger to the mounting bracket option in an enclosed cabinet and out of patient reach during
transport to comply with established crash test standards.
• Always make sure that the mounting bracket option is securely attached to the surface.
• Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion. Power
washing equipment may aerate contamination.
• Always wear insulated rubber gloves, in addition to personal protective equipment, when cleaning the S S M M R R T T Pak to
reduce the risk of injury.
• Always disconnect the S S M M R R T T charger from the wall outlet before cleaning to avoid the risk of electrical hazards.
• Do not spray fluid directly onto the S S M M R R T T charger.
• Do not power wash the S S M M R R T T charger.
• Do not use solvents, lubricants, or other chemicals to clean the S S M M R R T T charger unless otherwise directed.
• Do not immerse the S S M M R R T T charger in water or allow water to collect on top of the S S M M R R T T charger to avoid the risk of
electric shock.
• Always use only non-conductive materials to wipe the S S M M R R T T Pak.
• Always avoid excessive water exposure to the S S M M R R T T Pak terminals.
• Always refer to the disinfectant's Material Safety Data Sheet (MSDS) to verify the pH range. Disinfectants with pH levels
higher than 10.5 may cause the S S M M R R T T Pak housing material to crack.
• Do not directly handle or make contact with the S S M M R R T T Pak terminals while cleaning to avoid the risk of injury.
• Do not immerse the S S M M R R T T Pak in liquid to reduce the risk of electric shock.
• Do not use solvents, lubricants, or other chemicals to clean the S S M M R R T T Pak unless otherwise directed.
• Always relieve pressure before you disconnect hydraulic or other lines. Escaping fluid under pressure can penetrate the
skin and cause serious injury. Tighten all connections before you apply pressure. If an accident occurs, see a doctor
immediately.
• Do not use bare hands to check for hydraulic leaks.
C C A A U U T T I I O O N N
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of
the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case
the user will be required to correct the interference at their expense.
• Always set the cot load height before you place the cot into service.
• Always charge the battery before you place the product into service. An uncharged or depleted battery may cause poor
product performance.
• Always clear any obstacles that may interfere and cause injury to the operator or patient before operating the product.
• Do not load the base storage net above the safe working load of 20 lb (9 kg).
• Always be careful when you retract the base to avoid damaging items that are stored in the base storage net.
• Do not load the head end storage flat above the safe working load of 40 lb (18 kg).
• Always remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours).
• Always place the electrical S S M M R R T T charger power cord where it will not be stepped on, tripped over, or otherwise
subjected to damage or stress.
• Do not touch the S S M M R R T T Pak receptacle terminals with metal objects.
• Always grasp and pull the plug, not the cord, when you disconnect the S S M M R R T T charger to avoid the risk of damage to the
electrical plug and cord.
• Do not steam clean or ultrasonically clean the product.
• Do not exceed 180 °F (82 °C) as the maximum water temperature.
• Do not exceed 1500 psi (103.4 bar) as the maximum water pressure. If you use a hand held wand to wash the product,
keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
• Always allow to air dry.