Télécharger Imprimer la page

Keysight Technologies N6700C Serie Guide D'utilisation page 96

Publicité

Les langues disponibles

Les langues disponibles

Front Panel Menu Reference
Select Output\Settings\Resistance.
Enter a resistance value. Then press Select.
You can manually select a lower resistance range if you need better output resolution. The following
ranges can be selected:
High resistance range
Medium resistance range
Low resistance range
Front Panel Menu Reference
Select Output\Settings\Resistance.
Select a lower range. Then press Select.
When changing ranges, the load input will be turned off, then back on again.
Set the Output Power
For Keysight N679xA load modules
limit. The maximum programmable output power is either 100 W or 200 W, depending on which load
module is being programmed.
The power selection must previously be selected under Output Mode.
Front Panel Menu Reference
Select Output\Settings\Power.
Enter a Power Limit value. Then press Select.
You can also select a lower power range if you need better output resolution. The following ranges
can be selected:
High power range
Low power range
Keysight N6700C Series User's Guide
To set the output resistance to 60 Ω:
RES 60,(@1)
N6791A
from 30 Ω to 8 kΩ
from 2 Ω to 100 Ω
from 0.08 Ω to 3 Ω
To select the lower range, program a value that
falls within the range:
RES:RANG 1,(@1)
, the following commands program the input power
To set the inout power to 50 W:
POW 50,(@1)
N6791A
from 0.3 W to 100 W
from 0.04 W to 10 W
4 Operating the Power System
SCPI Command
N6792A
from 15 Ω to 8 kΩ
from 2 Ω to 100 Ω
from 0.04 Ω to 3 Ω
SCPI Command
SCPI Command
N6792A
from 0.5 W to 200 W
from 0.1 W to 20 W
95

Publicité

loading

Ce manuel est également adapté pour:

N6701cN6702c