Front Panel Menu Reference
Select Transient\Arb\Config\Trap\Level.
Enter the Start and Top level in volts. Then press Select.
Select Transient\Arb\Config\Trap\Time.
Enter the Start, Rise, Top, Fall, and End times in seconds..
Then press Select.
Step 3. Configure the final steps common to all Arbs. Refer to
the end of this section.
Configuring Staircase Arbs
Step 1. Select the Arb Type and Shape. Power and Resistance types apply to
Front Panel Menu Reference
Select Transient\Arb\Function.
Select a Voltage, Current, Power, or Resistance Arb Type.
Then select the Staircase Shape. Then press Select.
Step 2. Configure the staircase parameters. The parameters apply to the selected Arb type (Voltage,
Current, Power, or Resistance).
Enter the values for the level before and after the staircase.
Enter the time before the staircase, the stair-time of the ramp, and the time after the staircase.
Then enter the total number of steps.
Front Panel Menu Reference
Select Transient\Arb\Config\Stair\Level.
Enter the Start and End level in volts. Then press Select.
Select Transient\Arb\Config\Stair\Time.
Enter the Start, Stair, and End times in seconds.
Enter the number of steps. Then press Select.
Keysight N6700C Series User's Guide
SCPI Command
To configure a voltage trapezoid Arb, use:
ARB:VOLT:TRAP:STAR 0, (@1)
ARB:VOLT:TRAP:TOP 10,(@1)
ARB:VOLT:TRAP:STAR:TIM 0.25,(@1)
ARB:VOLT:TRAP:RTIM:TIM 0.5,(@1)
ARB:VOLT:TRAP:TOP:TIM 0.5,(@1)
ARB:VOLT:TRAP:FTIM:TIM 0.5,(@1)
ARB:VOLT:TRAP:END:TIM 0.25,(@1)
Configuring Steps Common to all Arbs
SCPI Command
To select a current, voltage, power, or resistance staircase
Arb, use:
ARB:FUNC:TYPE VOLT|CURR|POW|RES,(@1)
ARB:FUNC:SHAP STAIR,(@1)
SCPI Command
To configure a voltage staircase Arb, use:
ARB:VOLT:STA:STAR 0,(@1)
ARB:VOLT:STA:END 10,(@1)
ARB:VOLT:STA:STAR:TIM 0.25,(@1)
ARB:VOLT:STA:TIM 0.5,(@1)
ARB:VOLT:STA:END:TIM 0.25,(@1)
ARB:VOLT:STA:NST 6,(@1)
4 Operating the Power System
.