Step 2. Install the two front ears and the two rear extender supports on the instrument as shown in
the figure.
Use six M3 x 8mm screws (a) for the front ears and four M3 x 6mm screws (b) for the extender supports.
If the standard extender supports are either too short or too long, use the longer supports (c). Cut the
supports if required (d).
Step 3. Install the two rear ears on the back of the instrument rack as shown in the figure.
Use four plain 10-32 screws to install the rear ears.
Step 4. Slide the instrument into the rack, making sure that the rear extender supports are aligned
inside the rear ears.
Step 5. Attach the front ears to the front of the instrument rack using the four dress 10-32 screws
provided.
Step 6. This is optional. Insert a plain 10-32 screw through the slot of the rear ear and extender
support and attach it using a clip nut.
This will prevent the unit from being slid out of the front of the rack.
Bench Installation
Do not block the air intake and exhaust at the sides, or the exhaust at the rear of the
unit. Refer to the dimension diagram in the
Minimum clearances for bench operation are 2 inches (51 mm) along the sides and back.
Redundant Ground for 400 Hz Operation
At 400 Hz AC input operation, the leakage current of the unit exceeds 3.5 mA. This requires the
installation of a permanent, redundant ground from the instrument chassis to earth ground. This
ensures that ground will always be connected and that any leakage current will be diverted to ground.
Refer to the
Specifications
Refer to Redundant Ground in the Operating and Service Guide document for installation instructions.
Keysight N6700C Series User's Guide
section for power factor statistics at 400 Hz operation.
Specifications
section.
2 Installation
53