4 Operating the Power System
Note that you have the ability to change the identity of the mainframe. This functionality is only
intended for compatibility with previous "A" and "B" version mainframes.
Front Panel Menu Reference
Select System\Preferences\IDN
Check the box to identify the instrument as an Agilent
product.
The IDN change and PERSona commands affects the following identification items:
*IDN? command for vendor and model
l
VISA programmatic access APIs for vendor and model
l
LXI instrument web pages
l
LXI XML
l
LXI mDNS announcements
l
Instrument State Storage
The power system has ten storage locations in non-volatile memory to store instrument states. The
locations are numbered 0 through 9. Any state previously stored in the same location will be
overwritten.
Front Panel Menu Reference
Select States\SaveRecall.
In the SaveRecall field, enter a location from 0 to 9. Then
press Select.
Select Save to save the state or Recall to recall a state.
Specifying a power-on state
When shipped from the factory, the power system is configured to automatically recall the reset
(*RST) settings at power-on. However, you can configure the power system to recall the settings you
have stored in memory location 0 (RCL0) at power-on.
Front Panel Menu Reference
Select States\PowerOn.
Select Recall State 0. Then press Select.
Output Groups
Output grouping does not apply to Keysight Models N678xA SMU.
158
SCPI Command
SYST:PERS:MAN "<manufacturer>"
SYST:PERS:MOD "<model number>"
SCPI Command
To save a state in location 1: *SAV 1
To recall a state from location 1: *RCL 1
SCPI Command
OUTP:PON:STAT RCL0
Keysight N6700C Series User's Guide