Front Panel Menu Reference
Select Transient\Mode.
Set the voltage mode to List. Then press Select.
Step 2. Program the list of values for the List function. The order in which the values are entered
determines the order in which the values will be output. To generate the voltage list shown in the
figure, a list may include the following values: 9, 0, 6, 0, 3, 0
Front Panel Menu Reference
Select Transient\List\Config.
Select the List Step number and enter a voltage value.
Then press Select.
Repeat this for each step. Use the
the next step.
Step 3. Determine the time interval, in seconds, that the output remains at each step in the list before
it advances to the next step. To specify the six dwell intervals in the figure, a list may include the
following values: 2, 3, 5, 3, 7, 3
Front Panel Menu Reference
Select Transient\List\Config.
Select the List Step number and enter a voltage value.
Then press Select.
Repeat this for each step. Use the
the next step.
The number of dwell steps must equal the number of voltage steps. If a dwell list has
only one value, that value will be applied to all steps in the list.
Step 4. Determine how the list is paced. To pace the list by dwell time, set the list pacing to Dwell-
paced on the front panel menu. (Set the LIST:STEP command to AUTO.) As each dwell time elapses,
the next step is immediately output.
Front Panel Menu Reference
Select Transient\List\Pace.
Select Dwell paced. Then press Select.
In a trigger-paced list, the list advances one step for each trigger received. To enable trigger-paced
lists, select Trigger-paced on the front panel menu. (Set the LIST:STEP command to ONCE.)
The dwell time associated with each step determines the minimum time that the output remains at the
step. If a trigger is received before the dwell time completes, the trigger is ignored. To ensure that no
triggers are lost in a trigger-paced list, set the dwell time to zero.
Keysight N6700C Series User's Guide
To program output 1, use
VOLT:MODE LIST,(@1)
To program output 1, use
LIST:VOLT 9,0,6,0,3,0,(@1)
ÇÈ
keys to select
To program output 1, use
LIST:DWEL 2,3,5,3,7,3,(@1)
ÇÈ
keys to select
LIST:STEP AUTO,(@1)
4 Operating the Power System
SCPI Command
SCPI Command
SCPI Command
SCPI Command
115