3 Getting Started
4, 4, Enter
If you program an over-voltage protection level that is lower than the present output
voltage, the over-voltage protection circuit will trip and turn the output off. The front
panel status indicator will show OV.
Exit the Command Menu
There are two ways to exit the command menu
In Case of Trouble
Press the Help key to obtain additional help about any function control menu level. Press the Back key
to exit the Help menu.
The front panel Err indicator comes on if self-test fails or if other operating problems occur with your
instrument. Press the Error key to display the list of errors. Refer to the Service and Maintenance
section in the Operating and Service Guide document for further information.
82
The command menu is now at the function control level. This is the lowest level in
this path. Use the navigation keys to highlight the OVP Level control as shown
below. Enter the desired over-voltage protection level using the numeric keypad.
Then press Enter.
Press Channel at any time to select a different output channel. This can save
time because you can directly access the OVP control of each channel without
having to navigate through the menu levels.
Press Meter to immediately return to the metering screen. This is the quickest
way to return to metering mode.
Press Back to back up one level at a time in the command menu. This method
may be more convenient if there are other menu commands to be given.