Front Panel Menu Reference
Not available
If a power module has simultaneous measurements (refer to
BOTH voltage and current measurements on the same output channel. Power modules that do not
have simultaneous measurement capability cannot externally log both voltage and current.
Some power modules also have multiple measurement ranges. Selecting a lower measurement
range provides greater measurement accuracy, provided that the measurement does not exceed the
range. To select lower measurement ranges, use:
Front Panel Menu Reference
Not available
Seamless Measurements
Seamless measurement autoranging for both voltage and current measurements is
available on models N678xA SMU and with Option SMR.
Seamless measurement autoranging enables a wide dynamic measurement range with no data lost
when transitioning across ranges. Seamless autoranging does not include the 10 μA range, which
must be selected manually.
Front Panel Menu Reference
Not available
Specify the Integration Period
The integration period can be set from a minimum of 102.4 microseconds to a maximum of 60 seconds.
Front Panel Menu Reference
Not available
Keysight N6700C Series User's Guide
SCPI Command
To enable voltage or current measurements:
SENS:ELOG:FUNC:VOLT ON, (@1)
SENS:ELOG:FUNC:CURR ON, (@1)
To enable min/max measurements:
SENS:ELOG:FUNC:VOLT:MINM ON, (@1)
SENS:ELOG:FUNC:CURR:MINM ON, (@1)
SCPI Command
To set the 5 V measurement range:
SENS:ELOG:VOLT:RANG 5, (@1)
To set the 1 A measurement range:
SENS:ELOG:CURR:RANG 1, (@1)
SCPI Command
To enable seamless elog autoranging:
SENS:ELOG:VOLT:RANG:AUTO ON, (@1)
SENS:ELOG:CURR:RANG:AUTO ON, (@1)
SCPI Command
To set an integration period of 600 microseconds:
SENS:ELOG:PER 0.0006, (@1)
4 Operating the Power System
Model
Differences), you can enable