Front Panel Menu Reference
Select Transient\Arb\Function.
Select a Voltage, Current, Power, or Resistance Arb Type.
Then select the Pulse Shape. Then press Select.
Step 2. Configure the pulse parameters. The parameters apply to the selected Arb type (Voltage,
Current, Power, or Resistance).
Enter the values for the level before and during the top of the pulse.
Enter the time before the pulse, the time of the pulse, the time after the pulse. A pulse frequency can
only be specified directly using SCPI commands. The pulse frequency can also be specified using the
Time parameters.
Front Panel Menu Reference
Select Transient\Arb\Config\Pulse\Level.
Enter the Start and Top level in volts. Then press Select.
Select Transient\Arb\Config\Pulse\Time.
Enter the Start, Top, and End times in seconds. Then press
Select.
Step 3. Configure the final steps common to all Arbs. Refer to
the end of this section.
Configuring Ramp Arbs
Step 1. Select the Arb Type and Shape. Power and Resistance types apply to
Front Panel Menu Reference
Select Transient\Arb\Function.
Select a Voltage, Current, Power, or Resistance Arb Type.
Then select the Ramp Shape. Then press Select.
Step 2. Configure the ramp parameters. The parameters apply to the selected Arb type (Voltage,
Current, Power, or Resistance).
Keysight N6700C Series User's Guide
SCPI Command
To select a current, voltage, power, or resistance pulse Arb,
use:
ARB:FUNC:TYPE VOLT|CURR|POW|RES,(@1)
ARB:FUNC:SHAP PULS,(@1)
SCPI Command
To configure a voltage pulse Arb, use:
ARB:VOLT:PULS:STAR 0,(@1)
ARB:VOLT:PULS:TOP 10,(@1)
ARB:VOLT:PULS:STAR:TIM 0.25,(@1)
ARB:VOLT:PULS:TOP:TIM 0.5,(@1)
ARB:VOLT:PULS:END:TIM 0.25,(@1)
ARB:VOLT:PULS:FREQ 1,(@1)
Configuring Steps Common to all Arbs
SCPI Command
To select a current, voltage, power, or resistance ramp Arb,
use:
ARB:FUNC:TYPE VOLT|CURR|POW|RES,(@1)
ARB:FUNC:SHAP RAMP,(@1)
4 Operating the Power System
.
at
119