2 Installation
The following figure shows the connections for 50A power modules.
Setting the Outputs
Outputs connected together in series cannot be grouped.
To program outputs connected in series, first program the current limit of each output to the total
desired current limit point. Then program the voltage of each output so that the sum of both voltages
equals the total desired operating voltage. The simplest way to accomplish this is to program each
output to one half of the total desired operating voltage.
The operating mode of each output is determined by the output's programmed settings,
operating point, and load condition. Because these conditions may change during series
operation, the front panel status indicator will reflect these changes. This is normal.
Momentary status changes are also normal.