You can change the time interval resolution as follows:
Front Panel Menu Reference
Not available
Specify a Window Function
Windowing is a signal conditioning process that reduces the error in average measurements made in
the presence of periodic signals and noise. Two window functions are available: Rectangular and
Hanning. At power-on, the measurement window is Rectangular.
The Rectangular window calculates average measurements without any signal conditioning.
However, in the presence of periodic signals such AC line ripple, a Rectangular window can introduce
errors when calculating average measurements. This can occur when a non-integral number of cycles
of data has been acquired due to the last partial cycle of acquired data.
One way of dealing with AC line ripple is to use a Hanning window. The Hanning window applies a cos4
weighting function to the data when calculating average measurements. This attenuates the AC noise
in the measurement window. The best attenuation is achieved when at least three or more waveform
cycles are in the measurement.
Front Panel Menu Reference
Select Measure\Window.
Select Auto from the voltage or current dropdown
menu. Then press Select.
Trigger the Measurement
The following commands trigger and return measurement data:
Keysight N6700C Series User's Guide
SCPI Command
To set the resolution to 20 or 40 microseconds:
SENS:SWE:TINT:RES RES20
SENS:SWE:TINT:RES RES40
SCPI Command
To select the Hanning window function:
SENS:WIND HANN,(@1)
4 Operating the Power System
135