Remote Interface Configuration

If you have the security option installed on your instrument, the instrument must be unsecured with the secur-
ity code to perform many of these actions.
The instrument supports remote interface communication over three interfaces: GPIB (optional), USB,
and LAN (optional on 34460A). All three are "live" at power up when the instrument ships from the fact-
ory.
GPIB Interface: Set the instrument's GPIB address and connect to your PC using a GPIB cable.
l
USB Interface: Use the rear-panel USB connector to communicate with your PC. For details, see
l
Settings.
LAN Interface: By default, DHCP is on, which may enable communication over LAN. The acronym
l
DHCP stands for Dynamic Host Configuration Protocol, a protocol for assigning dynamic IP addresses
to networked devices. With dynamic addressing, a device can have a different IP address every time it
connects to the network.

Connectivity Software

The instrument ships with the Keysight Automation-Ready CD. This CD contains Keysight IO Libraries
l
Suite software, which must be installed to enable remote-interface operations. The CD auto-starts
and provides information on installing the software. Also includes Keysight Technologies
USB/LAN/GPIB Connectivity Guide, which contains additional information.

GPIB Configuration

Each device on the GPIB (IEEE-488) interface must have a unique whole number address between 0 and
30. The instrument ships with a default address of 10, and the GPIB address is displayed at power-on.
This setting is non-volatile; it will not be changed by power cycling or
l
Your computer's GPIB interface card address must not conflict with any instrument on the interface
l
bus.
Front Panel: Press [Utility] > I/O Config > GPIB Settings. From this menu, you can set the GPIB
l
address and turn GPIB on or off. After making changes, you must cycle power on the instrument for
the change to take effect.
SCPI:
l
SYSTem:COMMunicate:GPIB:ADDRess <address>
SYSTem:COMMunicate:ENABle {ON|1|OFF|0},GPIB
Keysight Truevolt Series Operating and Service Guide
Introduction to the Instrument
*RST
or SYSTem:PRESet.
USB
27