[SENSe:]{RESistance|FRESistance}:RESolution {<resolution>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:RESolution? [{MIN|MAX|DEF}]
Selects the measurement resolution for all resistance measurements. Specify the resolution in the same
units as the selected measurement function, not in number of digits.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version of
this command or query is identical to using the RESistance version.
< resolution >: See
Resolution Table
is equivalent to 10 PLC. Specify <resolution> in the measurement's units (V, A,
Hz, Ω, and so on).
Configure 2-wire resistance measurements with a 3 Ω resolution. Then make and read one measurement:
CONF:RES 1E6
RES:RES 3
READ?
Typical Response: +6.27531500E+05
You can specify MIN (best resolution) or MAX (worst resolution) instead of <resolution>.
l
To achieve normal mode (line frequency noise) rejection, use a resolution that corresponds to an
l
integration time that is an integral number of power line cycles.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
See Also
[SENSe:]{RESistance|FRESistance}:NPLC
Keysight Truevolt Series Operating and Service Guide
Parameter
or
Range, Resolution and NPLC.
SCPI Programming Reference
Typical Return
The default
+3.00000000E+00