Télécharger Imprimer la page

Keysight Truevolt Serie Mode D'emploi page 218

Masquer les pouces Voir aussi pour Truevolt Serie:

Publicité

Les langues disponibles

Les langues disponibles

CALCulate:TRANsform:HISTogram:CLEar[:IMMediate]
Clears the histogram data and restarts histogram ranging if it is enabled
late:TRANsform:HISTogram:RANGe:AUTO
Parameter Typical Return
(none)
(none)
Clear the histogram data:
CALC:TRAN:HIST:CLE
Does not clear measurements in reading memory.
l
To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate].
l
CALCulate:TRANsform:HISTogram:COUNt?
Returns the number of measurements collected since the last time the histogram was cleared.
(none)
Return the number of measurements used to compute the current histogram:
CALC:TRAN:HIST:COUN?
CALCulate:TRANsform:HISTogram:POINts {<value>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:POINts? [{MIN|MAX|DEF}]
Sets the number of bins between the lower and upper range values for the histogram. Two additional bins
always exist: one for measurements below the lower range and one for measurements above the upper
range.
Parameter
{10|20|40|100|200|400|MIN|MAX|DEF}. Default: 100. +100
See
Example
.
You can specify the lower and upper range values using
l
{LOWer|UPPer}. Lower and upper range values are computed automatically if
late:TRANsform:HISTogram:RANGe:AUTO
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
ON).
Parameter
Typical Return
CALCulate:TRANsform:HISTogram:RANGe:
is ON.
SCPI Programming Reference
(CALCu-
Typical
Return
+87
CALCu-
219

Publicité

loading

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a