Digital Multimeters
Truevolt Series
This manual contains information for using, programming, and servicing the Keysight Truevolt Series Digital
Multimeters (DMMs).
OPERATING AND SERVICE GUIDE

Publicité

loading

Sommaire des Matières pour Keysight Truevolt Serie

  • Page 1 Digital Multimeters Truevolt Series This manual contains information for using, programming, and servicing the Keysight Truevolt Series Digital Multimeters (DMMs). OPERATING AND SERVICE GUIDE...
  • Page 2 Notice: This document contains references to Agilent Tech- nologies. Agilent's former Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com.
  • Page 3 Calibration Procedures Introduction to the Instrument Instrument at a Glance Display – Easily display, save and document your measurement results Measurements – Keysight’s Truevolt measurement performance with modern I/O accessibility Programming Language Front Panel at a Glance Front Panel Keys...
  • Page 4 Triggering and Readings Instrument trigger model Trigger delay and multiple samples Storing and clearing readings Probe Hold Math - Introduction Math - Null Math - dB/dBm Scaling Math - Scaling Math - Smoothing Keysight Truevolt Series Operating and Service Guide...
  • Page 5 Frequency and Period Measurement Errors DC Current Temperature Measurements NULL Reading Autozero On/Off Making High–Speed AC Measurements Making High–Speed DC and Resistance Measurements Capacitance Capacitance Measurement Considerations Digitizing Measurements The Sampling Rate Level Triggering Keysight Truevolt Series Operating and Service Guide...
  • Page 6 ABORt FETCh? INITiate[:IMMediate] OUTPut:TRIGger:SLOPe {POSitive|NEGative}OUTPut:TRIGger:SLOPe? R? [<max_readings>] READ? ROUTe:TERMinals? TEST:ALL? UNIT:TEMPerature {C|F|K}UNIT:TEMPerature? CALCulate Subsystem Introduction CALibration Subsystem CONFigure Subsystem DATA Subsystem DISPlay Subsystem FORMat Subsystem HCOPy Subsystem IEEE 488.2 Common Commands LXI Subsystem Keysight Truevolt Series Operating and Service Guide...
  • Page 7 Command Errors (-100…) Execution Errors (-200…) Device-Specific Errors (-300…) Query Errors (-400…) Network Errors (+100...) Instrument Errors (+200…) Miscellaneous Errors (+300... and +500..) Licensing and Self-test Errors (+600...) Calibration Errors (+700...) Miscellaneous Errors (+800...) Keysight Truevolt Series Operating and Service Guide...
  • Page 8 High Current Verification Capacitance Verification (Optional Verification Test) Calibration Adjustment Procedures Calibration Procedures Input Connections Test Considerations Recommended Test Equipment Calibration Adjustment Process Gain Calibration Adjustment Overview Gain and Flatness Calibration Adjustment Overview Keysight Truevolt Series Operating and Service Guide...
  • Page 9 DC Voltage Gain Calibration Ohms Gain Calibration DC Current Gain Calibration DC High Voltage Gain Calibration AC Current 10 A Gain Calibration DC Current 10 A Gain Calibration Capacitance Offset Calibration (Optional) Finishing Calibration Index Keysight Truevolt Series Operating and Service Guide...
  • Page 10 Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this doc- ument that conflict with these terms, the warranty terms in the separate agreement shall control.
  • Page 11 U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Keysight provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data –...
  • Page 12 This product is marked with the ACMA RCM mark for compliance in Australia / New Zealand. A copy of the Manufacturer’s Australia Declaration of Conformity for this instrument can be obtained by contacting your local Keysight Technologies Sales Representative. ISM 1-A This text indicates that the instrument is an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 4).
  • Page 13 Removal of the AC power cord is the disconnect method to remove power from the instrument. Be sure to allow for adequate access to the power cord to permit disconnection from AC power. Use only the Keysight specified power cord for the country of use or one with equivalent ratings.
  • Page 14 The crest factor limit is listed in the product data sheet at www.keysight.com/find/truevolt-doc. Measurement Limits To avoid instrument damage and the risk of electric shock, do not exceed any of the Measurement Limits defined in the following section. Keysight Truevolt Series Operating and Service Guide...
  • Page 15 Measurement Limits are not exceeded. To ensure safe operation of the instrument, do not exceed the Measurement Limits shown on the front and rear panel, and defined as follows: 34460A Front Panel Terminal Area Keysight Truevolt Series Operating and Service Guide...
  • Page 16 The user-replaceable 3 A current-protection fuse is on the rear panel. There are 3 A and 10 A (34461A, 34465A and 34470A) current-protection fuses located inside the unit. Contact your Keysight Sales and Service Center or refer to product service documentation for replacement instructions.
  • Page 17 LO input, HI sense to LO input, and HI sense to LO sense. The 200 Vpk limit on the sense terminals is the Measurement Limit. Operational voltages in resistance meas- urements are much lower – up to ± 12 V in normal operation. Keysight Truevolt Series Operating and Service Guide...
  • Page 18 The Fine Tip, Mini Grabber, and SMT Grabber attachments plug onto the probe end of the Test Leads. Maintenance If any portion of the Test Lead Set is worn or damaged, do not use. Replace with a new Keysight 34139A Test Lead Set.
  • Page 19 Safety and Regulatory Information Declaration of Conformity Declarations of Conformity for this product and for other Keysight products may be downloaded from the Keysight Regulatory Web site: http://regulations.products.keysight.com/DoC/search.htm Keysight Truevolt Series Operating and Service Guide...
  • Page 20 Introduction to the Instrument User Information Front Panel Menu Reference Features and Functions Remote Interface Configuration LAN Configuration Procedure Web Interface Measurements Triggering and Readings Probe Hold Math Display Utility Menu Measurement Tutorial Keysight Truevolt Series Operating and Service Guide...
  • Page 21 Power Supplies Self-Test Procedures Battery Replacement Installing the Optional GPIB Interface User Replaceable Parts Performance Verification Performance Verification Calibration Procedures Calibration Procedures 34460A and 34461A Calibration Procedures 34465A and 34470A Calibration Procedures Firmware Update Keysight Truevolt Series Operating and Service Guide...
  • Page 22 Introduction to the Instrument Introduction to the Instrument The Keysight Technologies 34460A/61A/65A instruments are 6½-digit digital multimeters (DMMs); the 34470A is a 7½-digit DMM. Instrument at a Glance Front Panel at a Glance Rear Panel at a Glance Models and Options...
  • Page 23 Some of the front panel keys have text above them. This indicates that the key has a function that you can access by pressing and releasing [Shift] before pressing the key. For example, if you press and release [Shift] before pressing [Display], you will access the [Utility] function: Keysight Truevolt Series Operating and Service Guide...
  • Page 24 3 A Current Terminal Fuse 3 A Current Terminal (34461A/65A/70A only) Voltmeter Measurement Complete Output External Trigger Input Local Area Network (LAN) Connector USB Interface Connector Instrument Cable Lock Chassis Ground Screw AC Mains Input Keysight Truevolt Series Operating and Service Guide...
  • Page 25 *N/A - Accessory kit for 34460A - Test Leads, USB Standard Standard Standard Cable 3446MEMU *N/A *N/A Option Option 2 MB memory for 34465A and 34470A Truevolt DMMs license *N/A = Not Applicable. Keysight Truevolt Series Operating and Service Guide...
  • Page 26 Connectivity Software The instrument ships with the Keysight Automation-Ready CD. This CD contains Keysight IO Libraries Suite software, which must be installed to enable remote-interface operations. The CD auto-starts and provides information on installing the software. Also includes Keysight Technologies USB/LAN/GPIB Connectivity Guide, which contains additional information.
  • Page 27 If you change this parameter, you must either press the Apply Changes softkey (front panel) or send SYSTem:COMMunicate:LAN:UPDate (remote interface) for the change to take effect. To manually set an IP address, Subnet Mask, or Default Gateway, turn DHCP off, then change the IP setup as described below. Keysight Truevolt Series Operating and Service Guide...
  • Page 28 (for example: 255.255.0.0). SCPI: SYSTem:COMMunicate:LAN:SMASk "<mask>" If you change this parameter, you must either press the Apply Changes softkey (front panel) or send SYSTem:COMMunicate:LAN:UPDate (remote interface) for the change to take effect. Keysight Truevolt Series Operating and Service Guide...
  • Page 29 (remote interface) for the change to take effect. Domain Name A domain name is a registered Internet name that gets translated into an IP address. You cannot set it from the front panel or SCPI. Keysight Truevolt Series Operating and Service Guide...
  • Page 30 Re-selecting the LAN Settings page will display the new settings if a LAN restart took place. Web Interface The instrument includes a built-in Web Interface for remote instrument access and control over LAN via a Web browser. For details, see Interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 31 "nnn.nnn.nnn.nnn" where "nnn" in each case is a byte value in the range 0 through 255. You can enter a new IP address using the front panel arrow keys. Do not enter leading zeros. Keysight Truevolt Series Operating and Service Guide...
  • Page 32 For example, "192.168.020.011" is actually equivalent to decimal "192.168.16.9" because ".020" is inter- preted as "16" expressed in octal, and ".011" as "9". To avoid confusion, use only decimal values from 0 to 255, with no leading zeros. Keysight Truevolt Series Operating and Service Guide...
  • Page 33 SYSTem:IDENtify to set the *IDN? to match the actual model number, update the firmware, and then use SYSTem:IDENtify again to set the *IDN? response to the other model number. Keysight Truevolt Series Operating and Service Guide...
  • Page 34 Introduction to the Instrument Contacting Keysight Technologies You can contact Keysight Technologies for warranty, service, or technical support. In the United States: (800) 829-4444 In Europe: 31 20 547 2111 In Japan: 0120-421-345 www.keysight.com/find/assist to contact Keysight worldwide, or contact your Keysight Technologies representative.
  • Page 35 Use Built-in Help System Rack Mount the Instrument Prepare Instrument for Use Verify that you received the following items. If anything is missing, please contact your nearest Keysight sales office or Keysight authorized reseller. Power cord (for country of destination)
  • Page 36 Use only a 5x20 mm, time-lag, 0.25 A, 250 V certified fuse. The Keysight part number is 2110-0817. AC MAINS Nominal Line Voltage AC Mains Line Voltage...
  • Page 37 Use only a 5x20 mm, time-lag, 0.25 A, 250 V certified fuse. The Keysight part num- ber is 2110-0817. Step 4 Replace the fuse holder assembly by sliding it into the rear panel.
  • Page 38 If you turn off the instrument by disconnecting power (this is not recommended), the instrument turns on as soon as you re-apply power. You will not need to press the power switch. Keysight Truevolt Series Operating and Service Guide...
  • Page 39 Quick Start Adjust the Carrying Handle The handle has three positions, shown below. To adjust the handle position, grasp the sides of the handle, pull outward, and rotate the handle. Keysight Truevolt Series Operating and Service Guide...
  • Page 40 Press and hold any softkey or button, such as [Display]. If the message contains more information than will fit on the display, press the down arrow softkey to scroll down. Press Done to exit Help. Keysight Truevolt Series Operating and Service Guide...
  • Page 41 Press the arrow softkeys or use the front panel arrow keys to highlight the desired topic. Then press Select. You can also press Demos to run interactive demos on how to use the instrument. In this case, the following help topic appears: Keysight Truevolt Series Operating and Service Guide...
  • Page 42 Japanese, Korean, and Russian. To select the local language, press [Utility] > System Setup > User Settings > Help Lang. Then select the desired language. The menu softkey labels and status line messages are not translated. Keysight Truevolt Series Operating and Service Guide...
  • Page 43 You can mount the instrument in a standard 19-inch rack cabinet using one of two optional kits, each of which includes instructions and mounting hardware. You may also mount Another Keysight System II instrument of the same height and width beside the instrument.
  • Page 44 Sliding Support Shelf To install one or two instruments in a sliding support shelf, order shelf 5063-9255 and slide kit 1494- 0015. For a single instrument, also order filler panel 5002-3999. Keysight Truevolt Series Operating and Service Guide...
  • Page 45 Probe Hold Math Menu Display Menu Utility Menu Web Interface Throughout this document, "default" states and values are identified. These are the power-on default states when the instrument is shipped from the factory. Keysight Truevolt Series Operating and Service Guide...
  • Page 46 (readings). The figure below shows the result of digitizing a sine wave. This chapter dis- cusses the various ways to digitize signals. The importance of the sampling rate, and how to use level triggering. Keysight Truevolt Series Operating and Service Guide...
  • Page 47 Samples per trigger is set according to the data log duration (time or samples). Sample timer is put in timer (not immediate) mode and sample time is set according to the data log sample interval. Keysight Truevolt Series Operating and Service Guide...
  • Page 48 On return to Continuous mode, settings are left as done in data log or digitize mode except: Sample source is set to immediate. Pretrigger count is set to 0. Samples per trigger is set to 1. Trigger count is set to infinite. Keysight Truevolt Series Operating and Service Guide...
  • Page 49 Terminals: 3 A or 10 A Range: Auto, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, or 10 A (when terminals set to 10 A) Filter: >3 Hz, >20 Hz, >200 Hz Keysight Truevolt Series Operating and Service Guide...
  • Page 50 Gate Time: 10 ms, 100 ms (default), or 1 s Timeout: 1 s (default) or Auto Configure capacitance measurements: Range: 1 nF, 10 nF, 100 nF, 1 µF, 10 µF, 100 µF, or Auto (default) Keysight Truevolt Series Operating and Service Guide...
  • Page 51 Auto Zero: Off or On (default) (2-wire measurements only; not available for 4- wire measurements) OffstComp: Off (default) or ON. (RTD 2-wire and RTD 4-wire measurements only) Open Check: TCouple measurements only.* Units: °C, °F, or K Start and stop measurements. Keysight Truevolt Series Operating and Service Guide...
  • Page 52 Return to local control of the instrument (when in Remote mode), or indicate that the next front panel key will be "shifted", for example [Probe Hold] instead of [Single]. Keysight Truevolt Series Operating and Service Guide...
  • Page 53 / dBm (34460A/61A only) Configure dB, dBm Scaling (34465A/70A only) Configure scaling: dB, dBm, %, Mx-B Statistics Enable, disable, and clear statistics. Limits Enable or disable high and low limits. [Display] key Keysight Truevolt Series Operating and Service Guide...
  • Page 54 I/O Config Configure LAN (optional on 34460A), USB, and GPIB (optional) inter- faces. Test/Admin Perform self-test, calibration, security, license, and firmware update tasks. System Set user preferences, date and time, and power-on message. Setup Keysight Truevolt Series Operating and Service Guide...
  • Page 55 Features and Functions Measurements The Keysight Truevolt DMMs support many common measurements: DC Voltage AC Voltage DC Current AC Current Resistance Temperature Capacitance Continuity Diode Frequency and Period Data Logging Digitizing Level Triggering Keysight Truevolt Series Operating and Service Guide...
  • Page 56 For the 34460A/61A, press Aperture and choose the number of power-line cycles (PLCs) to use for the measurement. Only 1, 10, and 100 PLC provide normal mode (line frequency noise) rejection. Selecting 100 PLC provides the best noise rejection and resolution, but the slowest measurements: Keysight Truevolt Series Operating and Service Guide...
  • Page 57 It also leads to readings with less noise than the HighZ option, which is included for situations where the 10 MΩ load is significant. Keysight Truevolt Series Operating and Service Guide...
  • Page 58 ±12 VDC. The reference voltage is always autoranged, and the range used for both will be based on the larger result of these two measurements. Configure DCV Ratio measurements as shown: Keysight Truevolt Series Operating and Service Guide...
  • Page 59 ACV function with the signal already connected. If the DC bias remains constant, subsequent measurements can be made to full accuracy without additional settling delays. Step 1: Configure the test leads as shown. Step 2: Press [ACV] on the front panel. Keysight Truevolt Series Operating and Service Guide...
  • Page 60 If measurement speed is not an issue, choosing a lower frequency filter may result in quieter meas- urements, depending on the signal that you are measuring. For accurate displayed statistics of AC measurements in Front Panel mode, the default manual trigger delay ([Acquire] > Delay Man) must be used. Keysight Truevolt Series Operating and Service Guide...
  • Page 61 (PLCs) to use for the measurement . 1, 10, and 100 PLC provide normal mode (line frequency noise) rejection. Selecting 100 PLC provides the best noise rejection and resolution, but the slowest measurements: Keysight Truevolt Series Operating and Service Guide...
  • Page 62 (Off), the DMM measures the offset once and subtracts the offset from all subsequent measurements. The DMM takes a new offset measurement each time you change the function, range, or integration time. ( There is no autozero setting for 4-wire measurements.) Keysight Truevolt Series Operating and Service Guide...
  • Page 63 3 A terminals and the 10 A input terminals. When you change this to 10 A, the measurement range automatically becomes 10 A. When making measurements using the 10A terminals, the presence of a signal on the 3A ter- minals can cause significant errors. Keysight Truevolt Series Operating and Service Guide...
  • Page 64 If measurement speed is not an issue, choosing a lower frequency filter may result in quieter meas- urements, depending on the signal that you are measuring. For accurate displayed statistics of AC measurements in Front Panel mode, the default manual trigger delay ([Acquire] > Delay Man) must be used. Keysight Truevolt Series Operating and Service Guide...
  • Page 65 (PLCs) to use for the measurement . 1, 10, and 100 PLC provide normal mode (line frequency noise) rejection. Selecting 100 PLC provides the best noise rejection and resolution, but the slowest measurements: Keysight Truevolt Series Operating and Service Guide...
  • Page 66 DUT. With low power On, pressing Range displays the lower current sourced for each range: Low-power resistance measurements apply to the 100Ω through 100kΩ ranges only. The 1 MΩ through 1 GΩ ranges source the same current regardless of the low-power setting. Keysight Truevolt Series Operating and Service Guide...
  • Page 67 Under the same conditions, the 34401A returns the measurement’s absolute value so as to prevent the confusion associated with negative readings. The Keysight Truevolt Series DMMs will return negative val- ues. This allows the most accurate results after a NULL operation.
  • Page 68 Step 2: Press [Temp] on the front panel. The following menu appears. Step 3: Press Probe and choose the probe type. If you choose to use an RTD, the menu will have a softkey to specify the RTD's resistance at 0 degrees Celsius (R Keysight Truevolt Series Operating and Service Guide...
  • Page 69 Only 1, 10, and 100 PLC provide normal mode (line frequency noise) rejection. Selecting 100 PLC provides the best noise rejection and resolution, but the slowest measurements: Step 6: Use the Units softkey to display temperature in degrees Celsius, degrees Fahrenheit, or Kelvin. Keysight Truevolt Series Operating and Service Guide...
  • Page 70 R, or T thermocouples. This section describes general temperature measurement configuration information. For a detailed tutorial on temperature measurements, refer to Keysight Application Note 290 Practical Tem- perature Measurements available at www.keysight.com. Step 1: Configure the test leads as shown.
  • Page 71 Low Power: Disables (Off) or enables (On) low power measurements. The low power mode sources less test current per measurement range than is normally sourced for standard resistance measurements, to reduce power dissipation and self-heating in the probe. Press Done to return to the main temperature menu. Keysight Truevolt Series Operating and Service Guide...
  • Page 72 ( There is no autozero setting for 4-wire measurements.) Thermistor Requirements The DMM converts the measured thermistor resistance to temperature using the Steinhart-Hart ther- mistor equation: 1⁄T = A + B (Ln(R)) + C (Ln(R)) Where: Keysight Truevolt Series Operating and Service Guide...
  • Page 73 For example, if the measured internal temperature is +20.68 °C and the specified offset is +5 °C, the artificial cold junction mathematics uses the sum of the two values. Keysight recommends that the off- set be left at 0 unless metrology has indicated otherwise.
  • Page 74 If an open connection is detected (greater than 5 kΩ on the 10 kΩ range), the instrument reports an overload condition. For more information about making temperature measurements refer to Keysight Application Note 290 Practical Temperature Measurements available at www.keysight.com.
  • Page 75 If you apply a DC voltage or a short to the input terminals in capacitance measurement mode, the instrument reports an overload. Keysight Truevolt Series Operating and Service Guide...
  • Page 76 Continuity measurements behave as follows: ≤ 10 Ω displays measured resistance and beeps (if beeper enabled) 10 Ω to 1.2 kΩ displays measured resistance without beeping > 1.2 kΩ displays OPEN with no beep Keysight Truevolt Series Operating and Service Guide...
  • Page 77 0 to 5 V voltage displayed on the front panel, and instrument beeps when signal transitions into the 0.3 to 0.8 V threshold (if beeping is enabled) > 5 V the front panel displays OPEN, and SCPI returns 9.9E37 Keysight Truevolt Series Operating and Service Guide...
  • Page 78 For example, when measuring a signal between 20 and 200 Hz, use the 20 Hz fil- ter. If measurement speed is not an issue, choosing a lower frequency filter may result in quieter meas- urements, depending on the signal that you are measuring. Keysight Truevolt Series Operating and Service Guide...
  • Page 79 DUT failure may result in no signal; in this case, the failure can be detected sooner, speeding up test throughput. For accurate displayed statistics of AC measurements in Front Panel mode, the default manual trigger delay ([Acquire] > Delay Man) must be used. Keysight Truevolt Series Operating and Service Guide...
  • Page 80 *When accessed from remote, the instrument will continue data logging or digitizing to completion, and you can retrieve the readings from remote. Data Logging Overview This section is a summary of the steps involved in setting up data logging. Detailed steps are described below in Detailed Data Logging Steps. Keysight Truevolt Series Operating and Service Guide...
  • Page 81 Step 3: Press Sample Interval and specify the time interval between samples (readings). Note: You may see this message when configuring data logging: Sample interval is limited by meas- urement settings. The measurement time is determined by the measurement function, NPLC, Aperture, Keysight Truevolt Series Operating and Service Guide...
  • Page 82 (see Step 7 below). The number of read- ings you can store in memory depends on the amount of instrument memory. With the MEM option, the limit is 2,000,000 readings. Without the MEM option, the limit is 50,000 readings. Keysight Truevolt Series Operating and Service Guide...
  • Page 83 Separator - Specifies the character (Comma, Tab, or Semicolon) to use for separating the information on each row. When finished configuring reading storage, press Done > Done to return to the main data logging menu. Keysight Truevolt Series Operating and Service Guide...
  • Page 84 Options as described above in Step 6. Trend Chart for Data Logging The Trend Chart is particularly useful for viewing data logging measurements. See Trend Chart (Digitize and Data Log Modes) for details. Keysight Truevolt Series Operating and Service Guide...
  • Page 85 Interface. The Web User Interface monitor does not set the instrument to remote. *When accessed from remote, the instrument will continue data logging or digitizing to completion, and you can retrieve the readings from remote. Keysight Truevolt Series Operating and Service Guide...
  • Page 86 To digitize DC current, press DCI and configure the test leads as shown. On the 34461A/65A/70A, you can also configure the measurement using the 10 A terminal, which is recommended when measuring current above 1 A: Keysight Truevolt Series Operating and Service Guide...
  • Page 87 (Hz), or press the Sample Rate softkey to specify a sample interval (time between samples): Step 4: Press the Duration softkey to specify the length of time to digitize: or, press Duration again to specify the total number of samples to digitize: Keysight Truevolt Series Operating and Service Guide...
  • Page 88 The Pretrigger Count is limited to one less than the total number of readings specified to be taken by the Duration setting. Keysight Truevolt Series Operating and Service Guide...
  • Page 89 X and Y cursors (shown below), tracking curs- ors, pan control, and zoom allow you to analyze digitized data. See Trend Chart (Digitize and Data Log Modes) for details. Keysight Truevolt Series Operating and Service Guide...
  • Page 90 For lowest latency and highest sensitivity, use a fixed range when using level trigger. However, trigger level accuracy is reduced when the hardware detector is used. To increase trigger level accuracy and reduce sensitivity (avoid false triggers due to noise), use autor- ange: Keysight Truevolt Series Operating and Service Guide...
  • Page 91 Frequency and period considerations Trigger level accuracy is increased, latency is increased, and sensitivity is decreased as gate time is increased. Fixed voltage range eliminates uncertainties (due to range change) in trigger latency. Keysight Truevolt Series Operating and Service Guide...
  • Page 92 2-Wire, 4-Wire Resistance Pre-Math , Peak, Pre-Math Frequency , Frequency, Pre-Math Frequency Period Period, ACV, Pre-Math Period Frequency Frequency, ACV, Pre-Math Temperature Sensor Sensor, Pre-Math Ratio Input/Ref Input/Ref, Pre-Math Capacitance Pre-Math Cont None Diode None Keysight Truevolt Series Operating and Service Guide...
  • Page 93 After making one or more primary measurements for approximately 4 seconds, the DMM makes one sec- ondary measurement. For the Peak secondary measurement, the Clear Peaks softkey allows you to clear the accumulated his- tory of the peak-to-peak function (resets the record of peak values). Keysight Truevolt Series Operating and Service Guide...
  • Page 94 Auto - the instrument continuously takes measurements, automatically issuing a new trigger as soon as a measurement is completed. Single - the instrument issues one trigger each time the front panel [Single] key is pressed. Keysight Truevolt Series Operating and Service Guide...
  • Page 95 /Stop] stops and resumes measurements, and pressing [Single] switches the instrument to single trigger mode. In the Single and Ext modes, pressing [Run/Stop] stops readings if they are in progress, or switches the mode to Auto if readings are stopped. Keysight Truevolt Series Operating and Service Guide...
  • Page 96 Samples/Trig softkey) is satisfied. From the front panel, the instrument prevents you from specifying a sample timer that is shorter than the time required to make measurements based on the present function, range, and integration time. Keysight Truevolt Series Operating and Service Guide...
  • Page 97 Separator - Specifies the character (Comma, Tab, or Semicolon) to use for separating the information on each row. When finished configuring reading storage, press Done > Save Readings to save the readings in memory to a file. Keysight Truevolt Series Operating and Service Guide...
  • Page 98 Changing samples per trigger or trigger delay Changing display modes Changing the VM Comp output polarity Changing the digit mask Changing the histogram, bar meter, or trend chart scale Changing user preferences Executing self-test Keysight Truevolt Series Operating and Service Guide...
  • Page 99 Probe Hold is a front panel only function. Readings recorded in Probe Hold mode cannot be accessed from remote. You can, however, take a screen capture of the probe hold display, see Utility Menu - Man- age Files for details. Keysight Truevolt Series Operating and Service Guide...
  • Page 100 The [Math] key is the shifted [Null]: These math functions are available for the 34460A/61A: Null dB/dBm Scaling Statistics Limits These math functions are available for the 34465A/70A: Null Smoothing Filter Scaling Statistics Limits Keysight Truevolt Series Operating and Service Guide...
  • Page 101 To turn off the null feature, press [Null] again or press [Math] and set the first softkey to Off. See Also Math - Introduction dB/dBm Scaling Statistics Limits Keysight Truevolt Series Operating and Service Guide...
  • Page 102 = reading in dBm – reference value in dBm The relative value must be from -200 to +200 dBm (default 0). You can either measure this value by press- ing Measure Ref Value, or you can enter a specified value. Keysight Truevolt Series Operating and Service Guide...
  • Page 103 The reference resistance value (Ref R) may be 50, 75, 93, 110, 124, 125, 135, 150, 250, 300, 500, 600 (default), 800, 900, 1000, 1200, or 8000 Ω. Press Ref R and use the front panel up and down arrows to make your selection. See Also Math - Introduction Null Statistics Limits Keysight Truevolt Series Operating and Service Guide...
  • Page 104 The relative value must be from -200 to +200 dBm (default 0). You can either measure this value by press- ing Measure Ref Value, or you can use the dB Ref Value softkey to specify a value. dB scaling applies to ACV and DCV measurements only. Keysight Truevolt Series Operating and Service Guide...
  • Page 105 The User Units softkey allows you to specify the custom alphanumeric label of up to four characters (for example, RPM or °C) for Mx-B scaled measurements. Mx-B scaling applies to all measurement functions except Ratio, Continuity, and Diode. Keysight Truevolt Series Operating and Service Guide...
  • Page 106 Features and Functions See Also Math - Introduction Null Statistics Limits Keysight Truevolt Series Operating and Service Guide...
  • Page 107 A single reading, that is beyond the allowable variation from the readings in the filter, will reset the fil- ter and set the filter icon to solid yellow ( If all readings are within the allowable variation from the readings in the filter, the filter icon will be solid green ( Keysight Truevolt Series Operating and Service Guide...
  • Page 108 The average and standard deviation are not shown if dB or dBm scaling is in use. Remarks The Span value is the Max minus the Min. Press Clear Readings to clear reading memory and re-start statistics calculation. See Also Math - Introduction Null dB/dBm Scaling Limits Keysight Truevolt Series Operating and Service Guide...
  • Page 109 The Beeper softkey enables or disables beeping when limits are violated (this also enables or disables the beeper for the other functions that use the beeper - probe hold, diode, continuity, and errors. Clear Condi- tion resets the limit borders to green, as described below. Keysight Truevolt Series Operating and Service Guide...
  • Page 110 Because the limit is 10 V, the 9.994 VDC value is shown with the stand- ard background. In contrast, the 11.083 VDC reading is highlighted in red to indicate that it is outside the limits. Keysight Truevolt Series Operating and Service Guide...
  • Page 111 FAIL indicates that a limit has been exceeded. Number The bright red color (shown below) indicates that the displayed measurement exceeds the limits. The Number display also indicates how many times the limits have been exceeded. Keysight Truevolt Series Operating and Service Guide...
  • Page 112 Features and Functions See Also Math - Introduction Null dB/dBm Scaling Statistics Keysight Truevolt Series Operating and Service Guide...
  • Page 113 For the Number and Bar Meter displays, many primary measurement functions allow you to display a sec- ondary measurement. See Secondary Measurements for details. Selecting the Display Pressing the key, followed by the Display softkey allows you to select the display type: Keysight Truevolt Series Operating and Service Guide...
  • Page 114 100 ms, min 20 µs. See Digit- ize Mode details. Click a link below for more information on any of the display types: Number Bar Meter Trend Chart (34461A/65A/70A only) Histogram Secondary Measurements Keysight Truevolt Series Operating and Service Guide...
  • Page 115 Features and Functions Number By default, the instrument displays readings as a number: Keysight Truevolt Series Operating and Service Guide...
  • Page 116 ACV, Peak, or Pre-Math as the secondary measurement function. With ACV selec- ted as secondary, the display shows the DCV measurement near to top of the display, and the ACV meas- urement near the bottom of the display: Keysight Truevolt Series Operating and Service Guide...
  • Page 117 Features and Functions Refer to Secondary Measurements for more information on secondary measurements available for each measurement function. Keysight Truevolt Series Operating and Service Guide...
  • Page 118 In contrast, this image shows 4½ digits. The Auto softkey specifies that the number of digits displayed is based on other function-specific set- tings, such as the measurement aperture, set in NPLCs. Measurements are always rounded, never trun- cated. Keysight Truevolt Series Operating and Service Guide...
  • Page 119 Features and Functions See Also Display - Introduction Bar Meter Trend Chart (34461A/65A/70A only) Histogram Secondary Measurements Keysight Truevolt Series Operating and Service Guide...
  • Page 120 Features and Functions Bar Meter The bar meter (below) adds a moving bar below the standard Number display. Keysight Truevolt Series Operating and Service Guide...
  • Page 121 ACV, Peak, or Pre-Math as the secondary measurement function. With ACV selec- ted as secondary, the display shows the DCV measurement as a number near to top of the display, DCV in the bar meter, and the ACV measurement above the bar meter: Keysight Truevolt Series Operating and Service Guide...
  • Page 122 Features and Functions Refer to Secondary Measurements for more information on secondary measurements available for each measurement function. See Also Display - Introduction Number Trend Chart (34461A/65A/70A only) Histogram Secondary Measurements Keysight Truevolt Series Operating and Service Guide...
  • Page 123 Each pixel column on the trend chart display represents 1/400 of the Time Window value; the Time Window setting determines the amount of time represented by each pixel column as follows: Keysight Truevolt Series Operating and Service Guide...
  • Page 124 When the reading rate is slower than the time per pixel column, some pixel columns may represent no readings. In this situation, the trend chart continues with a horizontal line across the pixel: Scaling The Vertical Scale softkey specifies how the current vertical scale is determined. Keysight Truevolt Series Operating and Service Guide...
  • Page 125 Span of 5 V. If you have enabled limits, the (Limits) softkey also appears. This sets the vertical scale to match the lim- its. See Also Display - Introduction Number Bar Meter Histogram Keysight Truevolt Series Operating and Service Guide...
  • Page 126 The display is 400 pixels wide. You can select a zoom percentage of 1%, 2%, 5%, 10%, 20%, 50%, or 100%. For example, the graphic above shows 5% zoom and the graphic below shows the same signal at 100% zoom: Keysight Truevolt Series Operating and Service Guide...
  • Page 127 Tip: Zoom to 100% to pan one reading at a time. After selecting a reading, you can then decrease the zoom, if necessary, to view the surrounding signal. Cursors - Display and control X1, X2, Y1, Y2, and tracking cursors (shown as lines) on the trend chart. Keysight Truevolt Series Operating and Service Guide...
  • Page 128 Place Yn on Screen - Press Y1 or Y2 followed by this softkey to set the cursor to a value that puts it into view on the chart. X and Y – Displays the X1, X2, Y1 and Y2 cursors. In this mode, these softkeys are available: Keysight Truevolt Series Operating and Service Guide...
  • Page 129 X1 - Use the arrow keys to control the X1 cursor. X2 - Use the arrow keys to control the X2 cursor. ∆X Lock - When ON, keeps the X cursors a fixed distance apart as either cursor is moved. Keysight Truevolt Series Operating and Service Guide...
  • Page 130 When data logging to files, the trend chart behaves in a manner similar to that of the continuous meas- urement mode. That is, the number of readings shown per pixel column depends on the reading rate and the selected Time Window. Keysight Truevolt Series Operating and Service Guide...
  • Page 131 Zoom, Pan, and Cursors are not available when data logging to files. Refer to Trend Chart (Continuous Measurement Mode) for more information. See Also Display - Introduction Trend Chart (Continuous Measurement Mode) Number Bar Meter Histogram Keysight Truevolt Series Operating and Service Guide...
  • Page 132 = 20 bins, 501 to 1000 readings = 40 bins, 1001 to 5000 readings = 100 bins, 5001 to 10000 readings = 200 bins, >10000 readings = 400 bins. If the NPLC setting is < 1 PLC, or Aperture Time (34465A/70A only) is < 20 ms, the maximum number of bins is 100. Keysight Truevolt Series Operating and Service Guide...
  • Page 133 0 to 100% on the vertical scale, regardless of the his- togram’s scale. The final button on the histogram menu, Clear Readings, clears reading memory and starts a new his- togram. Keysight Truevolt Series Operating and Service Guide...
  • Page 134 2.250 00 k to 2.500 00k - Range of frequency measurement values for this bin. #29 - Number of samples in this bin. 11.5% - Percentage of the total number of samples in this bin. Keysight Truevolt Series Operating and Service Guide...
  • Page 135 Showing statistics (Shift > Math > Statistics) is particularly useful for the histogram display. For example, in the graphic below, the thick blue line is the average, each thin blue line represents one standard devi- ation from the average. Keysight Truevolt Series Operating and Service Guide...
  • Page 136 Features and Functions See Also Display - Introduction Number Bar Meter Trend Chart (34461A/65A/70A only) Keysight Truevolt Series Operating and Service Guide...
  • Page 137 Manage files, including creating screen shots Configure LAN (optional on 34460A), GPIB (optional on all models), and USB Perform instrument self-tests and administrative functions Configure the instrument's date and time and various user settings Keysight Truevolt Series Operating and Service Guide...
  • Page 138 Step 4: (Optional step) Press the Edit Cal String softkey and enter the information for the cal string mes- sage. Common messages include the last calibration date, calibration due date, or calibration department contact information. When finished, press the Done softkey Step 5: Press the Perform Autocal softkey: Keysight Truevolt Series Operating and Service Guide...
  • Page 139 Press Type to select whether you want to store a state or preferences file, and press Store State (shown above) or Store Pref (the softkey's label if you are storing preferences) to store the specified file. Action allows you to either store a file or to create a new folder. Keysight Truevolt Series Operating and Service Guide...
  • Page 140 (Last), a user-selected state file (User Defined), or the factory default state (Factory Defaults). Set to Defaults Set to Defaults loads the instrument's factory default state or preference settings. Keysight Truevolt Series Operating and Service Guide...
  • Page 141 Use the front panel arrows and [Select] key to navigate through the list, and press Select or Cancel to exit the browse window. The left and right arrows contract or expand a folder to hide or show its files. Keysight Truevolt Series Operating and Service Guide...
  • Page 142 Use the front panel arrows to point to a letter, and Previous Char and Next Char to move the cursor in the area where the name is entered. In the image below, there is no Next Char softkey because the cursor is at the end. Press [Done] or [Cancel] to finish. Keysight Truevolt Series Operating and Service Guide...
  • Page 143 (DHCP or Manual) selected. For example, press Host Name or Service mDNS to modify the instrument's host name or mDNS Service Name, shown below. Keysight Truevolt Series Operating and Service Guide...
  • Page 144 The instrument Telnet port is 5024. Open SCPI sessions on the Telnet connection by entering: telnet IP address 5024 Refer to the Keysight IO Libraries help for information on the VXI-11, Sockets, and HiSLIP protocols. USB Settings USB Settings configures the front panel USB (storage) and rear-panel USB (connectivity) connectors.
  • Page 145 To use Easy File Access at the same time that you are remotely programming the instrument with SCPI over the USB interface (USB SCPI), you must have Keysight IO Libraries Suite 16.3 or later installed on your PC. You may download the latest version at www.keysight.com/find/iosuite.
  • Page 146 Self-Test verifies proper instrument operation. See Self-Test Procedures for details, and always safely dis- connect inputs to the DMM terminals before running the Full Test. Calibrate Calibrate accesses the instrument calibration procedure. See Calibration for details. Keysight Truevolt Series Operating and Service Guide...
  • Page 147 Install License enables licensed instrument features. For information on obtaining licenses, go to www.keysight.com/find/truevolt. After receiving a license file from Keysight, use the following procedure to install the license: 1. Save the license file to a USB drive and connect the USB drive to the DMM's front panel USB connector.
  • Page 148 Diode – the forward bias voltage is between 0.3 and 0.8 V Continuity – a short is measured (less than or equal to 10 Ω) Error - an error is generated from the front panel or remote interface Keysight Truevolt Series Operating and Service Guide...
  • Page 149 *IDN? to return a particular manufacturer and model number. If you have an older Agilent 34460A or 34461A and upgrade to the new (Keysight) firmware, your instru- ment continues to respond with the manufacturer name “Agilent”, not “Keysight” until you set the front panel SCPI ID to 34460A or 34461A, send the...
  • Page 150 Use the front panel arrows and [Select] key to select the letters. Then press Done to exit and save the message. The message will appear as shown below when you turn the instrument on or press [Help] > About. Keysight Truevolt Series Operating and Service Guide...
  • Page 151 Features and Functions Web Interface The Keysight Truevolt Series DMMs include a built-in Web Interface for monitoring and controlling the instrument via a Web browser. To begin, connect your instrument to the LAN and enter the instrument's IP address into the address bar at the top of your PC's Web browser, or select the instrument in Con- nection Expert and click the Instrument Web Interface...
  • Page 152 In Allow Full Control mode, this window configures the instrument and takes measurements. Simply select the DMM settings and click Apply. The six buttons along the top of the screen are described below. Keysight Truevolt Series Operating and Service Guide...
  • Page 153 Graphics Display radio button and click Get Display Image to update the graphical display: View Data For both the Observe Only and Allow Full Control modes, this window keeps a running display of Read- ings in Memory and has these active controls: Keysight Truevolt Series Operating and Service Guide...
  • Page 154 SCPI configuration, and error queue. Check the desired boxes on the left side of the screen and then click Generate Report. The Change in SCPI Configuration box produces a list of all of the SCPI Keysight Truevolt Series Operating and Service Guide...
  • Page 155 Use the Enter SCPI command or query field to enter SCPI commands and click Send,to execute the com- mand, Read to read back a response, or Send & Read to execute the command and read back the response. Keysight Truevolt Series Operating and Service Guide...
  • Page 156 (typically 15 to 20 seconds). Configuration Page The Configuration page allows you to view the instrument's network configuration parameters, which con- figure the instrument to operate in your environment. To modify a parameter, click Modify Configuration. Keysight Truevolt Series Operating and Service Guide...
  • Page 157 Features and Functions Help Help is available for each of the tabs described above. For example: Keysight Truevolt Series Operating and Service Guide...
  • Page 158 Measurement Tutorial Measurement Tutorial Follow the suggestions in the sections below to achieve the greatest accuracy with Keysight Truevolt Ser- ies multimeters. Metrology DC Measurement Considerations Noise Rejection Resistance Measurement Considerations True RMS AC Measurements Capacitance Digitizing Data Logging Other Primary Measurement Functions...
  • Page 159 Measurement Tutorial Measurement Considerations To achieve the greatest accuracy with the Keysight Truevolt Series multimeters, you must eliminate potential measurement errors. This chapter describes common errors and suggests ways to avoid them. Keysight Truevolt Series Operating and Service Guide...
  • Page 160 Measurement Tutorial Metrology The 34470A 7½-digit DMM is well-suited for high-accuracy metrology applications. Keysight has a num- ber of application notes pertaining to maximizing measurement accuracy, such as: Practical Tips for Optimizing Measurement Accuracy with your Digital Multimeter -This Web page has a number of links to application notes addressing DMM measurement errors and how to make temperature measurements.
  • Page 161 To reduce the effects of loading errors and to minimize noise pickup, set the multimeter's input resistance to >10 GΩ (high-impedance) for the 100 mVDC, 1 VDC, and 10 VDC ranges. The input resistance is main- tained at 10 MΩ for the 100 VDC and 1000 VDC ranges. Keysight Truevolt Series Operating and Service Guide...
  • Page 162 Loose or vibrating test leads will also induce error voltages. Tie down test leads securely when operating near magnetic fields. Whenever possible, use magnetic shielding mater- ials or increased distance from magnetic sources. Keysight Truevolt Series Operating and Service Guide...
  • Page 163 The best way to eliminate ground loops is to isolate the multimeter from earth by not grounding the input terminals. If the multimeter must be earth–referenced, connect it and the DUT to one common ground point. Also connect the multimeter and DUT to the same electrical outlet whenever possible. Keysight Truevolt Series Operating and Service Guide...
  • Page 164 Short the test lead ends together and read the displayed test lead resistance. Press Null. The multimeter will store the test lead resistance as the 2-wire ohms null value, and subtract that value from subsequent measurements. See also "Null Measurements." Keysight Truevolt Series Operating and Service Guide...
  • Page 165 "dirty" surface films. Nylon and PVC are relatively poor insulators (10 Ω) when compared to PTFE insulators (10 Ω). Leakage from nylon or PVC insulators can easily contribute a 0.1% error when meas- uring a 1 MΩ resistance in humid conditions. Keysight Truevolt Series Operating and Service Guide...
  • Page 166 The Keysight Truevolt Series uses the same techniques to measure true rms voltage and true rms current. The effective AC voltage bandwidth is 300 kHz, while the effective AC current bandwidth is 10 kHz.
  • Page 167 Reducing the prf increases the density of lines in the Fourier spectrum, and increases the portion of the input signal’s spectral energy within the multimeter’s bandwidth, which improves accuracy. In summary, error in rms measurements arise when there is significant input signal energy at frequencies above the multimeter’s bandwidth. Keysight Truevolt Series Operating and Service Guide...
  • Page 168 The first zero crossing of a simple pulse occurs at = 1/t This gives an immediate impression of the high-frequency content by identifying where this crossing occurs as a function of crest factor: f =(CF )(prf). Keysight Truevolt Series Operating and Service Guide...
  • Page 169 Factor of 3) and duration 111 µs. The prf can be calculated to be 1000 Hz, as follows: Thus, from the table above, this AC waveform can be measured with 0.18 percent additional error. Keysight Truevolt Series Operating and Service Guide...
  • Page 170 AC or DC voltages above 15 volts peak on the un-needed sense terminals are likely to cause measurement errors. If unexpected errors are occurring, signals on the un-needed ter- minals is an area to check. Keysight Truevolt Series Operating and Service Guide...
  • Page 171 = 2.362e-4, C = 9.285e-8. Using an incorrect type of thermistor can result in errors greater than 20 °C for a temperature being measured of 100 °C. For a detailed tutorial on temperature measurements, refer to Keysight Application Note 290 Practical Temperature Measurements available at www.keysight.com.
  • Page 172 0.2 seconds. This settling time only affects measurement accuracy when DC offset levels vary from sample to sample. If maximum measurement speed is desired, you may want to add an Keysight Truevolt Series Operating and Service Guide...
  • Page 173 Set the integration time (NPLC or aperture) to minimum Select a fixed range (autorange off) Disable autozero Disable offset compensation (resistance measurements) Refer to the particular measurement type in Measurements for more information on the above functions. Keysight Truevolt Series Operating and Service Guide...
  • Page 174 (Dt). This measurement is repeated at two different times during the exponential rise that occurs. An algorithm takes the data from these four points, and by linearizing that exponential rise over these “short apertures”, accurately calculates the capacitance value. Keysight Truevolt Series Operating and Service Guide...
  • Page 175 A high quality film capacitor typically shows the least of this and an electrolytic capacitor the most, with ceramic capacitors typically in between. Keysight Truevolt Series Operating and Service Guide...
  • Page 176 This limits the bandwidth of the input signal so that aliasing cannot occur. Since the multimeter has a variable sample rate for DCV digitizing, and to preserve the upper bandwidth Keysight Truevolt Series Operating and Service Guide...
  • Page 177 NPLC setting reduced to achieve digitize rate/interval, you can press Shift > Help > 1 View the last mes- sage displayed > Select to view more information. The DIG option increases the maximum reading rates from 5k readings/s (standard) to 50k readings/s (maximum). Keysight Truevolt Series Operating and Service Guide...
  • Page 178 After acquisition is complete – trend chart pan, zoom, and cursors can be used to review data. Statistics and histogram data are calculated after the digitizing is completed. Keysight Truevolt Series Operating and Service Guide...
  • Page 179 The Web User Interface monitor does not set the instrument to remote. *When accessed from remote, the instrument will continue data logging or digitizing to completion, and you can retrieve the readings from remote. Keysight Truevolt Series Operating and Service Guide...
  • Page 180 (particularly NPLC settings for DC and res- istance measurements). In this situation, press the measurement function key (for example, DCV) and decrease the aperture setting (either NPLC or Time). Keysight Truevolt Series Operating and Service Guide...
  • Page 181 _00001, the third filename with _00002, and so on. When data logging to files, the maximum number of readings you can data log to files is 100 hours x 1000rdgs/sec =360,000,000. Keysight Truevolt Series Operating and Service Guide...
  • Page 182 You can specify a comma, tab, or semicolon delimiter to separate the values. When Metadata is off, only the reading data is saved: Keysight Truevolt Series Operating and Service Guide...
  • Page 183 Measurement Tutorial Keysight Truevolt Series Operating and Service Guide...
  • Page 184 That is, the number of readings shown per pixel column depends on the reading rate and the selected Time Window. Keysight Truevolt Series Operating and Service Guide...
  • Page 185 The Web User Interface monitor does not set the instrument to remote. *When accessed from remote, the instrument will continue data logging or digitizing to completion, and you can retrieve the readings from remote. Keysight Truevolt Series Operating and Service Guide...
  • Page 186 For lowest latency and highest sensitivity, use a fixed range when using level trigger. However, trigger level accuracy is reduced when the hardware detector is used. To increase trigger level accuracy and reduce sensitivity (avoid false triggers due to noise), use autor- ange: Keysight Truevolt Series Operating and Service Guide...
  • Page 187 Frequency and period considerations Trigger level accuracy is increased, latency is increased, and sensitivity is decreased as gate time is increased. Fixed voltage range eliminates uncertainties (due to range change) in trigger latency. Keysight Truevolt Series Operating and Service Guide...
  • Page 188 The table below shows the multimeter's approximate input resistance at various frequencies. Input Frequency Input Resistance (kΩ) 100 Hz 1 kHz 10 kHz 100 kHz 15.7 For low frequencies, the loading error is: At high frequencies, the additional loading error is: Keysight Truevolt Series Operating and Service Guide...
  • Page 189 AC current is larger due to the multimeter's series inductance and your measurement connections. The burden voltage increases as the input frequency increases. Some circuits may oscillate when performing current measurements due to the multimeter's series inductance and your meas- urement connections. Keysight Truevolt Series Operating and Service Guide...
  • Page 190 Typically, the DMM exhibits about 0.06% additional error for a 100 V, 100 kHz reverse input. You can use the grounding techniques described for DC common mode problems to minimize AC common mode voltages. Keysight Truevolt Series Operating and Service Guide...
  • Page 191 Unnecessary signals applied to the Hi and Lo Sense terminals can also cause errors. AC or DC voltages above 15 volts peak on the sense terminals are likely to cause measurement errors. Keysight Truevolt Series Operating and Service Guide...
  • Page 192 Another way to put it is that for continuous and data log modes, sample timer/interval is always con- trolled by the other measurement settings. In digitize mode, the sample timer/interval, NPLC, and aper- ture have equal control; changing any one setting can change the other settings. Keysight Truevolt Series Operating and Service Guide...
  • Page 193 SCPI Programming Reference SCPI Programming Reference This section contains information to help you program the Keysight Truevolt Series digital multimeters over a remote interface using the SCPI programming language. Introduction to the SCPI Language Commands by Subsystem Command Quick Reference...
  • Page 194 Optional parameters are enclosed in square brackets ( [ ] ). The brackets are not sent with the command string. If you do not specify a value for an optional parameter, the instrument uses a default value. Keysight Truevolt Series Operating and Service Guide...
  • Page 195 The following example sets the trigger count to 10 measurements: TRIG:COUN 10 You can then query the count value by sending: TRIG:COUN? You can also query the minimum or maximum count allowed as follows: TRIG:COUN? MIN TRIG:COUN? MAX Keysight Truevolt Series Operating and Service Guide...
  • Page 196 (or "m"). For your convenience, the instrument interprets "mV" (or "MV") as millivolts, but "MHZ" (or "mhz") as megahertz. Likewise "MΩ" (or "mΩ") is interpreted as megΩ. You can use the prefix "MA" for mega. For example, "MAV" is interpreted as megavolts. Keysight Truevolt Series Operating and Service Guide...
  • Page 197 For example, The following example displays the message "WAITING..." on the instrument's front panel (the quotes are not displayed). DISP:TEXT "WAITING..." You can also display the same message using The following example with single quotes. DISP:TEXT 'WAITING...' Keysight Truevolt Series Operating and Service Guide...
  • Page 198 ABORt CALCulate:CLEar[:IMMediate] CALCulate:AVERage Subsystem CALCulate:LIMit Subsystem CALCulate:SCALe Subsystem CALCulate:SMOothing Subsystem CALCulate:TCHart[:STATe] CALCulate:TRANsform:HISTogram Subsystem CALibration Subsystem CONFigure Subsystem DATA Subsystem DISPlay Subsystem FETCh? FORMat Subsystem HCOPy Subsystem IEEE 488.2 Common Commands INITiate[:IMMediate] LXI Subsystem Keysight Truevolt Series Operating and Service Guide...
  • Page 199 [SENSe:]CURRent Subsystem [SENSe:]DATA2 Subsystem [SENSe:]{FREQuency|PERiod} Subsystem [SENSe:]{RESistance|FRESistance} Subsystem [SENSe:]TEMPerature Subsystem [SENSe:]VOLTage Subsystem STATus Subsystem SYSTem Subsystem - General Purpose Commands SYSTem Subsystem - I/O Configuration SYSTem:LICense Subsystem SYSTem:LOCK Subsystem TEST:ALL? TRIGger Subsystem UNIT:TEMPerature Keysight Truevolt Series Operating and Service Guide...
  • Page 200 Typical Return (none) (none) Abort a measurement in progress: ABORt Use this to abort a measurement when the instrument is waiting for a trigger, or for aborting a long measurement or series of measurements. Keysight Truevolt Series Operating and Service Guide...
  • Page 201 Data Register's condition register (see Status System Introduction). The instrument clears all measurements from reading memory when the measurement configuration changes, or when any of these commands are executed: INITiate, MEASure:<function>?, READ?, *RST, SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 202 Data Register's condition register (see Status System Introduction). To retrieve the measurements from the reading memory, use FETCh?. Use DATA:REMove? read and erase all or part of the available measurements. ABORt to return to idle. Keysight Truevolt Series Operating and Service Guide...
  • Page 203 CONF:VOLT:DC 10 SAMP:COUN 2 OUTP:TRIG:SLOP POS INIT On the 34460A, this requires option 34460A-LAN or option 3446LANU. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 204 14) is set in the Questionable Data Register's condition register (see Status System Introduction). The instrument clears all measurements from reading memory when the measurement configuration changes, or when any of these commands are executed: INITiate, MEASure:<function>?, READ?, *RST, SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 205 Data Register's condition register (see Status System Introduction). The instrument clears all measurements from reading memory when the measurement configuration changes, or when any of these commands are executed: INITiate, MEASure:<function>?, READ?, *RST, SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 206 Parameter Typical Return (none) FRON or REAR Return the state of the Front/Rear switch: ROUT:TERM? On the 34460A, this query always returns FRON. Keysight Truevolt Series Operating and Service Guide...
  • Page 207 If one or more tests fail, the query returns +1 and stores an error in the error queue. For a complete listing of the error messages related to self-test failures, see Self-test Error Messages. Following the test, the instrument returns to the instrument state that existed before the self-test. Keysight Truevolt Series Operating and Service Guide...
  • Page 208 The command also accepts CEL or FAR, but the query returns C or F. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). See Also CONFigure:TEMPerature MEASure:TEMPerature? [SENSe:]FUNCtion[:ON] Keysight Truevolt Series Operating and Service Guide...
  • Page 209 It also optionally performs the following math operations, as shown in the diagram below: Smoothing Scaling Statistics Limit checking Histogram Only calculations enabled by CALC:SCAL:STAT, CALC:TRAN:HIST:STAT, CALC:LIM:STAT, and CALC:AVER:STAT are performed. Parts of the CALCulate Subsystem CALCulate:CLEar[:IMMediate] AVERage Subsystem Keysight Truevolt Series Operating and Service Guide...
  • Page 210 SCPI Programming Reference HISTogram Subsystem LIMit Subsystem SCALing Subsystem SMOothing Subsystem Keysight Truevolt Series Operating and Service Guide...
  • Page 211 Clears all limits, histogram data, statistics, and measurements. Parameter Typical Return (none) (none) Clear all limits, histogram data, statistics, and measurements: CALC:CLE:IMM Items are cleared synchronously; histogram, statistics, and limit data all restart at the same time that measurements restart. Keysight Truevolt Series Operating and Service Guide...
  • Page 212 Failed) of the Questionable Status Register; measurements below 3.2 V set bit 11 (Lower Limit Failed). *CLS STAT:PRES CONF:VOLT 10,.001 SAMP:COUN 100 CALC:LIM:LOW 3.2 CALC:LIM:UPP 3.4 CALC:LIM:STAT ON INIT *WAI STAT:QUES? Typical Response: +4096 (at least one measurement was above the upper limit) Command Summary CALCulate:LIMit:CLEar[:IMMediate] CALCulate:LIMit:{LOWer|UPPer}[:DATA] CALCulate:LIMit[:STATe] Keysight Truevolt Series Operating and Service Guide...
  • Page 213 Questionable Data Register when the measurement function changes, or when any of the following commands is executed: CALCulate:LIMit:STATe ON INITiate MEASure:<function>? READ? CALCulate:LIMit:CLEar *RST SYSTem:PRESet To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate]. Keysight Truevolt Series Operating and Service Guide...
  • Page 214 12 ("Upper Limit Failed"). See STATus Subsystem Introduction for further information. CONFigure resets both limits to 0. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 215 Questionable Data Register when the measurement function changes, or when any of the following commands is executed: CALCulate:LIMit:STATe ON INITiate MEASure:<function>? READ? CALCulate:LIMit:CLEar *RST SYSTem:PRESet The instrument turns this setting OFF when the measurement function is changed or after *RST SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 216 CALCulate:TRANsform:HISTogram:CLEar[:IMMediate] CALCulate:TRANsform:HISTogram:POINts CALCulate:TRANsform:HISTogram:RANGe:AUTO CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer} CALCulate:TRANsform:HISTogram[:STATe] INITiate[:IMMediate] MEASure:<function>? READ? *RST SYSTem:PRESet Command Summary CALCulate:TRANsform:HISTogram:ALL? CALCulate:TRANsform:HISTogram:CLEar[:IMMediate] CALCulate:TRANsform:HISTogram:COUNt? CALCulate:TRANsform:HISTogram:DATA? CALCulate:TRANsform:HISTogram:POINts CALCulate:TRANsform:HISTogram:RANGe:AUTO CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer} CALCulate:TRANsform:HISTogram[:STATe] Keysight Truevolt Series Operating and Service Guide...
  • Page 217 The number of measurements greater than the upper range value Range values are real numbers returned in the form +1.00000000E+00. The number of measurements and bin data are signed, positive integers returned in the form +100. Keysight Truevolt Series Operating and Service Guide...
  • Page 218 You can specify the lower and upper range values using CALCulate:TRANsform:HISTogram:RANGe: {LOWer|UPPer}. Lower and upper range values are computed automatically if CALCu- late:TRANsform:HISTogram:RANGe:AUTO is ON. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 219 If no histogram data exists, 9.91E37 (Not a Number) is returned. Lower and upper range values are computed automatically if CALCu- late:TRANsform:HISTogram:RANGe:AUTO is ON. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 220 SCPI Programming Reference CALCulate:TRANsform:HISTogram[:STATe] {ON|1|OFF|0} CALCulate:TRANsform:HISTogram[:STATe]? Enables or disables histogram computation. Parameter Typical Return {ON|1|OFF|0}. Default: OFF. 0 (OFF) or 1 (ON) Example The instrument turns this setting OFF when the measurement function is changed or after *RST SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 221 The dB relative value parameter is relative to the dBm reference resistance set with CALCu- late:SCALe:DBM:REFerence. The instrument sets the reference value to 0.0 with automatic reference selection enabled after a Factory Reset (*RST), an Instrument Preset (SYSTem:PRESet), or a change in measurement function. Keysight Truevolt Series Operating and Service Guide...
  • Page 222 (CALCulate:SCALe:DBM:REFerence), relative to 1 milliwatt. (dBm = 10 × log 10 (measurement 2 / reference resistance / 1 mW)). dBm scaling applies to ACV and DCV measurements only. Keysight Truevolt Series Operating and Service Guide...
  • Page 223 Enable Mx-B scaling with a gain of 100 and an offset of 5. CALC:SCAL:FUNC SCAL CALC:SCAL:GAIN 100 CALC:SCAL:OFFS 5 CALC:SCAL:STAT ON Applies only to the 34465A and 34470A. The instrument sets the gain value to 0.0 after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 224 Specifying a reference value disables automatic reference selection (CALCu- late:SCALe:REFerence:AUTO OFF). The instrument sets the reference value to 0.0 with automatic reference selection enabled after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet), or a change in measurement function. Keysight Truevolt Series Operating and Service Guide...
  • Page 225 You must execute CALCulate:SCALe:FUNCtion before executing CALCulate:SCALe[:STATe], oth- erwise an incompatibility error will be generated. Scaling is set to OFF when you change measurement functions (for example, changing from DCV to ACV). You must re-enable scaling after changing measurement functions. Keysight Truevolt Series Operating and Service Guide...
  • Page 226 (comma, ampersand, …), and some symbols, including the degree symbol ( ° ). If you set the measurement label to °C, °F, or K, this has no effect the temperature units set using UNIT:TEMPerature. Keysight Truevolt Series Operating and Service Guide...
  • Page 227 Applies only to the 34465A and 34470A. CALCulate:SCALe:UNIT specifies the unit string to be displayed. The instrument disables displaying the unit string after a Factory Reset (*RST) or an Instrument Pre- set (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 228 SCPI Programming Reference CALCulate:AVERage Subsystem This subsystem calculates measurement statistics. Command Summary CALCulate:AVERage:ALL? CALCulate:AVERage:AVERage? CALCulate:AVERage:CLEar[:IMMediate] CALCulate:AVERage:COUNt? CALCulate:AVERage:MAXimum? CALCulate:AVERage:MINimum? CALCulate:AVERage:PTPeak? CALCulate:AVERage:SDEViation? CALCulate:AVERage[:STATe] Keysight Truevolt Series Operating and Service Guide...
  • Page 229 The other six queries listed above return individual values. Parameter Typical Return (none) (see below) Return the statistics of 100 frequency measurements: CONF:FREQ 1.0E3,.001 SAMP:COUN 100 CALC:AVER:STAT ON INIT *WAI CALC:AVER:ALL? Typical Response: +1.00520000E+03,+1.00512000E+03,+1.00527000E+03,+4.13500000E-01 Keysight Truevolt Series Operating and Service Guide...
  • Page 230 CALCulate:AVERage:STATe ON CALCulate:AVERage:CLEar INITiate MEASure:<function>? READ? *RST SYSTem:PRESet All values are returned in the form +1.23450000E+01. When dB or dBm scaling is used, the CALC:AVER:AVER and CALC:AVER:SDEV queries return +9.91000000E+37 (not a number). Keysight Truevolt Series Operating and Service Guide...
  • Page 231 CALCulate:AVERage[:STATe] {ON|1|OFF|0} CALCulate:AVERage[:STATe]? Enables or disables statistics computation. Parameter Typical Return {ON|1|OFF|0}. Default: OFF. 0 (OFF) or 1 (ON) Return the statistics of 100 frequency measurements: CONF:FREQ 1.0E3,.001 SAMP:COUN 100 CALC:AVER:STAT ON INIT *WAI CALC:AVER:ALL? Typical Response: +1.00520000E+03,+1.00512000E+03,+1.00527000E+03,+4.13500000E-01 Keysight Truevolt Series Operating and Service Guide...
  • Page 232 Statistics are cleared when the measurement function changes or when any of these commands is executed: CALCulate:AVERage:STATe ON CALCulate:AVERage:CLEar INITiate MEASure:<function>? READ? *RST SYSTem:PRESet The instrument turns this setting OFF when the measurement function is changed or after *RST SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 233 10, 50, or 100 measurements. Equal weighting is applied to all measurements used in the average. The instrument sets the response time to FAST after a Factory Reset (*RST ) or after an Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 234 Applies only to the 34465A and 34470A. CALCulate subsystem and the individual calculate functions are disabled for timestamp meas- urements. The instrument disables the smoothing (moving average) filter after a Factory Reset (*RST) or an Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 235 Trend chart data must be disabled (OFF) to make measurements at the highest speed (50 kHz). This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 236 You must unsecure the instrument to calibrate it or to change the calibration code or security code. Parameter Typical Return (none) 0 (successful) or 1 (unsuccessful) Calibrate the ADC: CAL:ADC? This is the first step in a re-calibration sequence, and must be done before doing the offset cal- ibration. Keysight Truevolt Series Operating and Service Guide...
  • Page 237 You can perform this query regardless of whether the instrument is secured. This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. See Also SYSTem:SECurity:COUNt? Keysight Truevolt Series Operating and Service Guide...
  • Page 238 When shipped from the factory, the instrument is secured, with the security code (calibration pass- word) set to AT3446XA. This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 239 You must unsecure the instrument to calibrate it or to change the calibration code or security code. Parameter Typical Return (none) (none) Store calibration constants into non-volatile memory: CAL:STOR Keysight Truevolt Series Operating and Service Guide...
  • Page 240 Return the calibration time: CAL:TIME? You can perform this query regardless of whether the instrument is secured. This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 241 You must unsecure the instrument to calibrate it or to change the calibration code or security code. Parameter Typical Return Numeric, default 0.0 +2.37000000E-02 Specify calibration value 0.0237: CAL:VAL 2.37E-2 This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 242 AUTO (including voltage range for frequency and period measurements) Samples per Trigger 1 sample Trigger Count 1 trigger Trigger Delay Automatic delay Trigger Source Immediate Trigger Slope NEGative Math Functions Disabled. Other parameters are unchanged. Per-function Null State Disabled Keysight Truevolt Series Operating and Service Guide...
  • Page 243 The following example configures the instrument for 2-wire resistance measurements, triggers the instru- ment to make one measurement using INITiate, and stores the measurement in reading memory. The 10 kΩ range is selected with 100 Ω resolution. CONF:RES 10000,100 INIT FETC? Typical Response: +5.95850000E+03 Keysight Truevolt Series Operating and Service Guide...
  • Page 244 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. READ? INITiate to start the measurement. Keysight Truevolt Series Operating and Service Guide...
  • Page 245 From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above 1.2 kΩ, the instrument displays "OPEN" with no beep. The FETCh?, READ?, and MEASure:CONTinuity? queries return the measured resistance, regardless of its value. READ? INITiate to start the measurement. Keysight Truevolt Series Operating and Service Guide...
  • Page 246 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. To control the measurement rate for AC measurements, change the trigger delay or AC filter band- Keysight Truevolt Series Operating and Service Guide...
  • Page 247 5.05 V, the front panel shows "OPEN", and the value returned from SCPI is 9.9E37. The FETCh?, READ?, and MEASure:DIODe? queries return the measured voltage, regardless of its value. READ? INITiate to start the measurement. Keysight Truevolt Series Operating and Service Guide...
  • Page 248 Overload on the front panel and returns "9.9E37" from the remote interface. Autor- anging can be enabled for the input voltage. READ? INITiate to start the measurement. See Also CONFigure? MEASure:{FREQuency|PERiod}? Keysight Truevolt Series Operating and Service Guide...
  • Page 249 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. READ? INITiate to start the measurement. See Also CONFigure? MEASure:{RESistance|FRESistance}? Keysight Truevolt Series Operating and Service Guide...
  • Page 250 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. READ? INITiate to start the measurement. Keysight Truevolt Series Operating and Service Guide...
  • Page 251 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. To control the measurement rate for AC measurements, change the trigger delay or AC filter band- Keysight Truevolt Series Operating and Service Guide...
  • Page 252 SCPI Programming Reference width. READ? INITiate to start the measurement. See Also CONFigure? MEASure[:VOLTage]:{AC|DC}? [SENSe:]CURRent:AC:BANDwidth Keysight Truevolt Series Operating and Service Guide...
  • Page 253 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. READ? INITiate to start the measurement. See Also CONFigure? Keysight Truevolt Series Operating and Service Guide...
  • Page 254 SCPI Programming Reference MEASure[:VOLTage][:DC]:RATio? Keysight Truevolt Series Operating and Service Guide...
  • Page 255 You can store up to 1,000 measurements in the reading memory of the 34460A, 10,000 meas- urements on the 34461A, 50,000 measurements on the 34465A/70A (without the MEM option), or 2,000,000 measurements on the 34465A/70A (with the MEM option). Keysight Truevolt Series Operating and Service Guide...
  • Page 256 The Standard Operation Condition register shows the current condition. See STATus Subsystem Intro- duction for further information. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 257 No error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Ques- tionable Data Register's condition register (see Status System Introduction). Keysight Truevolt Series Operating and Service Guide...
  • Page 258 OFF. While a message is displayed, information relating to the current instrument operation is not sent to the front panel display. The display text is unaffected by *RST. It is cleared at power-on. Keysight Truevolt Series Operating and Service Guide...
  • Page 259 Display histogram: DISP:VIEW HIST The display is not turned on if the display state is OFF. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 260 Specifies the data format to be either ASCII or REAL. Affects the data format of the MEASure?, READ?, FETCh?, DATA:REMove? commands only. Parameter Typical Return <length> ASCii: "9" or REAL: "64" ASC,9 Specify the 64-bit REAL data format: FORM:DATA REAL,64 or FORM:DATA REAL Keysight Truevolt Series Operating and Service Guide...
  • Page 261 If REAL is specified, numeric data is transferred as REAL binary data in IEEE 488.2 definite-length block format. The <length> parameter is optional, but "64" is the default and only allowed value. The byte order is controlled by FORMat:BORDer. Keysight Truevolt Series Operating and Service Guide...
  • Page 262 8-bit extended ASCII codes. Example The image format (PNG or BMP) is specified by HCOPy:SDUMp:DATA:FORMat. HCOPy:SDUMp:DATA:FORMat {PNG|BMP} HCOPy:SDUMp:DATA:FORMat? Specifies the format for images returned by HCOPy:SDUMp:DATA?. Parameter Typical Return {PNG|BMP}. Default: PNG. PNG or BMP Example Keysight Truevolt Series Operating and Service Guide...
  • Page 263 IEEE 488.2 Common Commands These commands and queries are associated with the IEEE 488.2 standard. Command Summary *CAL? *CLS *ESE *ESR? *IDN? *LRN? *OPC *OPC? *OPT? *PSC *RCL *RST *SAV *SRE *STB? *TRG *TST? *WAI Keysight Truevolt Series Operating and Service Guide...
  • Page 264 One or more bits are set in the Status Byte Register and may generate a Request for Service *SRE (RQS). Bits must be enabled using Standard Operation One or more bits are set in the Standard Operation Register (bits must be enabled, see Summary STATus:OPERation:ENABle Keysight Truevolt Series Operating and Service Guide...
  • Page 265 Following the autocalibration, the instrument returns to the instrument state that existed before the autocalibration. *CLS Clear Status Command. Clears the event registers in all register groups. Also clears the error queue. Parameter Typical Return (none) (none) Clear event register bits and error queue: *CLS Keysight Truevolt Series Operating and Service Guide...
  • Page 266 The <enable> setting is nonvolatile; it does not change after a Factory Reset (*RST) , Instrument Pre- set (SYSTem:PRESet), Status Preset (STATus:PRESet) or Clear Status (*CLS). Keysight Truevolt Series Operating and Service Guide...
  • Page 267 To set the enable register mask, write a decimal value to the register using *ESE. Once a bit is set, it remains set until cleared by reading the event register or by sending *CLS (clear status). Keysight Truevolt Series Operating and Service Guide...
  • Page 268 - Main firmware revision pp.pp - Front panel firmware revision mm.mm - Measurement board firmware revision gg.gg - Measurement board FPGA revision bb - Measurement board revision fp - Front panel board revision Keysight Truevolt Series Operating and Service Guide...
  • Page 269 - Main firmware revision pp.pp - Front panel firmware revision mm.mm - Measurement board firmware revision gg.gg - Measurement board FPGA revision bb - Measurement board revision fp - Front panel board revision Keysight Truevolt Series Operating and Service Guide...
  • Page 270 Other commands may be executed before Operation Complete bit is set. The difference between *OPC and *OPC? is that*OPC sets a status bit when the operation completes, and *OPC? outputs "1" when the operation completes. Keysight Truevolt Series Operating and Service Guide...
  • Page 271 A 0 is returned for options that are not installed. For example, a 34461A with only GPIB installed returns GPB,0. A 34460A with only GPB and SEC installed returns GPB,0,SEC. Options vary by model, as shown here. Keysight Truevolt Series Operating and Service Guide...
  • Page 272 It is overwritten with the power-down state of the instrument when power is cycled. The instrument state enable settings are stored in nonvolatile memory. They do not change when power has been off, after a Factory Reset (*RST) or after an Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 273 (power-on status clear) determines whether Status Byte enable register is cleared at power on. For example, *PSC 0 preserves the enable register contents through power cycles. The Status Byte enable register is not cleared by *RST. Keysight Truevolt Series Operating and Service Guide...
  • Page 274 Typical Response:+1.00520000E+01, … (5 measurements) After setting the trigger source, you must send INITiate to place the DMM in the "wait-for-trigger" state. The *TRG command is not accepted unless the DMM is in the "wait-for-trigger" state. Keysight Truevolt Series Operating and Service Guide...
  • Page 275 Parameter Typical Return (none) (none) Wait until all pending operations complete: *WAI Configuration changes and measurements are considered pending operations. Therefore, *WAI causes these changes to occur and complete. Keysight Truevolt Series Operating and Service Guide...
  • Page 276 This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). This parameter is set to its default value when the instrument is shipped from the factory and after SYSTem:SECurity:IMMediate. Keysight Truevolt Series Operating and Service Guide...
  • Page 277 Typical Return Quoted string of up to 63 characters. "Keysight 34460A Digital Multimeter - 98765" Default: "Keysight <model number> Digital Multimeter - <serial>" where <serial> is the last 5 digits of the instrument's serial number Set the mDNS service name: LXI:MDNS:SNAM:DES "LAB1-DMM"...
  • Page 278 If the LAN interface or specific LAN services (VXI-11, sockets, and so on) have been disabled by SYSTem:COMMunicate:ENABle, you must separately re-enable the interface or services and cycle instrument power for the LAN to be operational. Keysight Truevolt Series Operating and Service Guide...
  • Page 279 AUTO (including voltage range for frequency and period measurements) Samples per Trigger 1 sample Trigger Count 1 trigger Trigger Delay Automatic delay Trigger Source Immediate Trigger Slope NEGative Math Functions Disabled. Other parameters are unchanged. Per-function Null State Disabled Keysight Truevolt Series Operating and Service Guide...
  • Page 280 The following example configures the instrument for 2-wire resistance measurements, triggers the instru- ment to take a measurement, and reads the measurement. The 1 kΩ range is selected with 0.1 Ω res- olution. MEAS:RES? 1000,0.1 Typical Response: +3.27150000E+02 Keysight Truevolt Series Operating and Service Guide...
  • Page 281 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 282 From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above 1.2 kΩ, the instrument displays "OPEN" with no beep. The FETCh?, READ?, and MEASure:CONTinuity? queries return the measured resistance, regardless of its value. Keysight Truevolt Series Operating and Service Guide...
  • Page 283 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 284 0.3 and 0.8 V (unless beeper is disabled). If the signal exceeds 5.05 V, the front panel shows "OPEN", and the value returned from SCPI is 9.9E37. The FETCh?, READ?, and MEASure:DIODe? queries return the measured voltage, regardless of its value. Keysight Truevolt Series Operating and Service Guide...
  • Page 285 0.1 ppm × <range> (MINimum) The input signal for frequency or period measurements has an AC voltage component. By default, this command uses autorange to select the voltage range. With no signal applied, 0 is returned. Keysight Truevolt Series Operating and Service Guide...
  • Page 286 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 287 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 288 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 289 If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 290 The MMEMory subsystem's general purpose file management commands are listed below. The MMEMory subsystem also includes two other types of commands: STATe and PREFerence File Management Data Transfer Commands and Queries MMEMory:CATalog[:ALL]? MMEMory:CDIRectory MMEMory:COPY MMEMory:DELete MMEMory:MDIRectory MMEMory:MOVE MMEMory:RDIRectory Keysight Truevolt Series Operating and Service Guide...
  • Page 291 NOT begin with \ or /. Folder and file names cannot contain the following characters: \ / : * ? " < > | The combination of folder and file name cannot exceed 240 characters. Keysight Truevolt Series Operating and Service Guide...
  • Page 292 <file_size> is the file size in bytes. If no files exist, only <mem_used>,<mem_free> is returned. Because the instrument uses a small amount of space in the flash file system for internal use, the <mem_used> value is never zero. Keysight Truevolt Series Operating and Service Guide...
  • Page 293 If the destination file exists, it will be overwritten, unless marked as hidden or system. To copy a file to a file of the same name in a different folder, specify just the <drive> and/or <path> for <destination>. Keysight Truevolt Series Operating and Service Guide...
  • Page 294 To rename a file, specify the same folder for <file1> and <file2>. To move a file to a file of the same name in a different folder, you can specify just the <drive>:<path> for <file2>. Keysight Truevolt Series Operating and Service Guide...
  • Page 295 Trend chart and histogram display settings Data Log and Digitize mode settings Temperature probe settings Thermocouple reference junction offset adjust- ment The MMEMory subsystem also includes two other types of commands: General Purpose File Management Data Transfer Keysight Truevolt Series Operating and Service Guide...
  • Page 296 NOT begin with \ or /. Folder and file names cannot contain the following characters: \ / : * ? " < > | The combination of folder and file name cannot exceed 240 characters. Keysight Truevolt Series Operating and Service Guide...
  • Page 297 Load nonvolatile I/O settings and user preferences from the specified file: MMEM:LOAD:PREF "INT:\MyPreferences" See the State File/Preferences File table for a listing of the information contained in each file. The specified folder must exist and cannot be marked as hidden or system. Keysight Truevolt Series Operating and Service Guide...
  • Page 298 When shipped from the factory, the instrument is configured to automatically recall the power-down state file when power is restored. This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 299 The specified folder must exist and cannot be marked as hidden or system. The specified file cannot be empty or marked as hidden or system. Use this query before sending *RCL MMEMory:LOAD:STATe to determine whether a state has been stored in the file. Keysight Truevolt Series Operating and Service Guide...
  • Page 300 These commands transfer files into and out of the instrument's mass memory. The MMEMory subsystem also includes two other types of commands: General Purpose File Management STATe and PREFerence File Management Command Summary MMEMory:DOWNload:DATA MMEMory:DOWNload:FNAMe MMEMory:FORMat:READing:CSEParator MMEMory:FORMat:READing:INFormation MMEMory:FORMat:READing:RLIMit MMEMory:STORe:DATA MMEMory:UPLoad? Keysight Truevolt Series Operating and Service Guide...
  • Page 301 NOT begin with \ or /. Folder and file names cannot contain the following characters: \ / : * ? " < > | The combination of folder and file name cannot exceed 240 characters. Keysight Truevolt Series Operating and Service Guide...
  • Page 302 Specify tabs as the column separator: MMEMory:FORMat:READing:CSEParator TAB Applies only to the 34465A and 34470A. Affects the format for the .csv file created by the MMEMory:STORe:DATA command, or by the data logging/digitizing features in the 34465A/34470A. Keysight Truevolt Series Operating and Service Guide...
  • Page 303 ON - Readings are stored in files named dat00001.csv, dat00002.csv, dat00003.csv ... OFF - Readings are is stored in a single file named dat00001.csv. The file size is limited to 2 4.294967296 GBytes (the maximum allowed by the FAT32 file system). Keysight Truevolt Series Operating and Service Guide...
  • Page 304 Uploads the contents of a file from the instrument to the host computer. Parameter Typical Return Any valid file name IEEE 488.2 definite-length block Upload the state file "Myfile.sta" in the root directory of the internal flash file system to the host computer: MMEM:UPL? "INT:\Myfile.sta" Keysight Truevolt Series Operating and Service Guide...
  • Page 305 No error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Ques- tionable Data Register's condition register (see Status System Introduction). Keysight Truevolt Series Operating and Service Guide...
  • Page 306 In this example, if the trigger occurs after 5 pretrigger samples have been taken (instead of the specified 20,000), only 30,000 post-trigger samples are taken. This is because memory was reserved for 20,000 pretrigger samples. Keysight Truevolt Series Operating and Service Guide...
  • Page 307 IMMediate - The first sample starts one trigger delay time after the trigger, and then the trigger delay time is inserted between the first and second samples, the second and third, and so on: Keysight Truevolt Series Operating and Service Guide...
  • Page 308 "wait-for-trigger" state. The instrument sets the sample source to "IMM" after a Factory Reset ( *RST command) or an instru- ment Preset (SYSTem:PRESet command). See Also TRIGger:COUNt TRIGger:DELay TRIGger:SOURce SAMPle:COUNt SAMPle:TIMer Keysight Truevolt Series Operating and Service Guide...
  • Page 309 Basically, MIN is automatically determined so that the sample interval is always greater than the sampling time (see above illus- tration). Execute SAMPle:TIMer? MIN to determine the recommended interval for the current meas- urement configuration. Keysight Truevolt Series Operating and Service Guide...
  • Page 310 However, the value is rounded to the nearest step. For dc measurements, the step size is 1 µs. For ac measurements, it is ac bandwidth dependent. The instrument sets the sample timer to 1 second after a reset ( *RST command) or an instrument pre- (SYSTem:PRESet command). See Also TRIGger:COUNt Keysight Truevolt Series Operating and Service Guide...
  • Page 311 SCPI Programming Reference TRIGger:DELay TRIGger:SOURce SAMPle:COUNt SAMPle:SOURce Keysight Truevolt Series Operating and Service Guide...
  • Page 312 Changing the measurement function disables scaling, limit testing, histogram, statistics, and trend chart data collection (CALC:SCAL:STAT, CALC:LIM:STAT, CALC:TRAN:HIST:STAT, CALC:AVER:STAT, CALC:TCH:STAT set to OFF). This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 313 Enabling the scaling function also enables automatic null value selection ([SENSe:]CAPa- citance:NULL:VALue:AUTO). To set a fixed null value, use: [SENSe:]CAPacitance:NULL:VALue. The instrument disables the null function after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 314 To use the null value, turn on the null state with the [Sense:]CAPacitance:NULL[:STATe] command. This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 315 Configure capacitance measurements, turn autorange on, turn null function on, use the AUTO null func- tion. Make two measurements and send them to the instrument's output buffer: CONF:CAP CAP:RANG:AUTO ON CAP:NULL ON CAP:NULL:VAL:AUTO ON SAMP:COUN 2 READ? Typical Response: +3.01963841E-10,+3.01566277E-10 Keysight Truevolt Series Operating and Service Guide...
  • Page 316 If the input signal is greater than can be measured on the specified fixed range, the instrument dis- plays Overload on front panel and returns "9.9E37" from the remote interface. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 317 "CALCulate:DATA" - The measurement value before any math operations are done (including NULL). READ? INITiate to start the measurement. Use [SENSe:]DATA2? to retrieve the secondary measurement. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 318 SCPI Programming Reference [SENSe:]CURRent Subsystem This subsystem configures AC and DC current measurements. Command Summary [SENSe:]CURRent:AC:BANDwidth [SENSe:]CURRent:{AC|DC}:NULL[:STATe] [SENSe:]CURRent:{AC|DC}:NULL:VALue [SENSe:]CURRent:{AC|DC}:NULL:VALue:AUTO [SENSe:]CURRent:{AC|DC}:RANGe [SENSe:]CURRent:{AC|DC}:RANGe:AUTO [SENSe:]CURRent:{AC|DC}:TERMinals [SENSe:]CURRent:AC:SECondary [SENSe:]CURRent[:DC]:APERture [SENSe:]CURRent[:DC]:APERture:ENABled [SENSe:]CURRent[:DC]:NPLC [SENSe:]CURRent[:DC]:RESolution [SENSe:]CURRent[:DC]:SECondary [SENSe:]CURRent[:DC]:ZERO:AUTO [SENSe:]CURRent:SWITch:MODE Keysight Truevolt Series Operating and Service Guide...
  • Page 319 20 Hz - 300 kHz (Medium) 0.25 s/measurement 200 Hz - 300 kHz (Fast) 0.025 s/measurement This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 320 Enabling the scaling function also enables automatic null value selection ([SENSe:]CURRent: {AC|DC}:NULL:VALue:AUTO ON). To set a fixed null value, use: [SENSe:]CURRent:{AC|DC}:NULL:VALue. The instrument disables the null function after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 321 OFF). To use the null value, the null state must be on ([SENSe:]CURRent:{AC|DC}:NULL:STATe ON). This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 322 The instrument enables automatic null value selection when the null function is enabled ([SENSe:]CURRent:{AC|DC}:NULL:STATe ON). This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 323 For example, if you have a measurement with 1 µA resolution on the 100 mA range, chan- ging to the 1 A range results in a resolution of 10 µA. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 324 10 A range, but it does not affect the values of [SENSe:]CURRent: {AC|DC}:RANGe or [SENSe:]CURRent:{AC|DC}:RANGe:AUTO. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 325 1 µA resolution on the 1 A range, changing to the 10 A terminals results in a resolution of 10 µA. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 326 (Without the DIG option) 200 µs to 1 s (2 µs resolution), default: 100 ms. (With the +1.00000000E- DIG option) 20 µs to 1 s (2 µs resolution), default: 100 ms. Enable aperture time mode, set the aperture time to 300 ms: CURR:APER:ENAB ON CURR:DC:APER 300E-03 Keysight Truevolt Series Operating and Service Guide...
  • Page 327 This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 328 Setting the integration time also sets the measurement resolution. Resolution Table shows the rela- tionship between integration time and resolution. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). See Also [SENSe:]CURRent[:DC]:RESolution Keysight Truevolt Series Operating and Service Guide...
  • Page 329 To achieve normal mode (line frequency noise) rejection, use a resolution that corresponds to an integration time that is an integral number of power line cycles. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). See Also [SENSe:]CURRent[:DC]:NPLC Keysight Truevolt Series Operating and Service Guide...
  • Page 330 AC and DC current measurements from remote, program each function separately, see CONFigure:CURRent:{AC|DC}. "PTPeak" - The peak-to-peak, maximum, and minimum values of the input signal. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 331 MEASure:CURRent:DC?. Autozero is automatically turned OFF when you select an integration time less than 1 PLC with these commands. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 332 The switch mode applies to the 3A current terminals only. The 10A terminals are always effectively in FAST mode. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 333 DATA2? to retrieve the secondary meas- urement. The instrument clears all measurements from reading memory when the measurement configuration changes, or when any of these commands are executed: INITiate, MEASure:<function>?, READ?, *RST, SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 334 DCV reading, Min Peak, Max Peak, Peak-to-Peak) Later, clear peak-to-peak secondary measurement data. Make and retrieve measurements. DATA2:CLE READ?;DATA2? Typical Response: +1.29177114E-03;+6.07221096E-04,+3.43677658E-03,+2.82955548E-03 (Data returned in this order: DCV reading, Min Peak, Max Peak, Peak-to-Peak) Keysight Truevolt Series Operating and Service Guide...
  • Page 335 SCPI Programming Reference [SENSe:]{FREQuency|PERiod} Subsystem This subsystem configures frequency and period measurements. Command Summary [SENSe:]{FREQuency|PERiod}:APERture [SENSe:]{FREQuency|PERiod}:NULL[:STATe] [SENSe:]{FREQuency|PERiod}:NULL:VALue [SENSe:]{FREQuency|PERiod}:NULL:VALue:AUTO [SENSe:]{FREQuency|PERiod}:RANGe:LOWer [SENSe:]{FREQuency|PERiod}:TIMeout:AUTO [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe:AUTO [SENSe:]FREQuency:SECondary [SENSe:]PERiod:SECondary Keysight Truevolt Series Operating and Service Guide...
  • Page 336 1 ppm × <range> 100 ms 1 ppm × <range> (MINimum) 0.1 ppm × <range> (MINimum) This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 337 Enabling the scaling function also enables automatic null value selection ([SENSe:]{FREQuency|PERi- od}:NULL:VALue:AUTO ON). To set a fixed null value, use: [SENSe:]{FREQuency|PERiod}:NULL:VALue. The instrument disables the null function after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 338 Configure frequency measurements, using the null function to subtract 1 kHz from them. Make and read two measurements: CONF:FREQ FREQ:NULL:STAT ON;VAL 1 kHz SAMP:COUN 2 READ? Typical Response: +1.04530000E+00,+1.04570000E+00 Make a second set of measurements using automatic null value selection: FREQ:NULL:VAL:AUTO ON READ? Typical Response: +0.00000000E+00,+0.01420000E+00 Keysight Truevolt Series Operating and Service Guide...
  • Page 339 20 Hz - 300 kHz (Medium) 0.25 s/measurement 200 Hz - 300 kHz (Fast) 0.025 s/measurement This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 340 This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. Keysight Truevolt Series Operating and Service Guide...
  • Page 341 Overload on the front panel and returns "9.9E37" from the remote interface. Autor- anging can be enabled for the input voltage. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 342 Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of range. Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 343 “FREQuency” - The frequency measurement of the input signal. "VOLTage:AC" - The AC voltage measurement of the input signal. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 344 SCPI Programming Reference [SENSe:]{RESistance|FRESistance} Subsystem This subsystem configures two- and four-wire resistance measurements. Command Summary [SENSe:]{RESistance|FRESistance}:APERture [SENSe:]{RESistance|FRESistance}:APERture:ENABled [SENSe:]{RESistance|FRESistance}:NPLC [SENSe:]{RESistance|FRESistance}:NULL[:STATe] [SENSe:]{RESistance|FRESistance}:NULL:VALue [SENSe:]{RESistance|FRESistance}:NULL:VALue:AUTO [SENSe:]{RESistance|FRESistance}:OCOMpensated [SENSe:]{RESistance|FRESistance}:POWer:LIMit[:STATe] [SENSe:]{RESistance|FRESistance}:RANGe [SENSe:]{RESistance|FRESistance}:RANGe:AUTO [SENSe:]{RESistance|FRESistance}:RESolution [SENSe:]{FRESistance|RESistance}:SECondary [SENSe:]RESistance:ZERO:AUTO Keysight Truevolt Series Operating and Service Guide...
  • Page 345 [SENSe:]<function>:APERture:ENABled command to 1 (ON); send- ing [SENSe:]<function>:NPLC sets it to 0 (OFF). This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 346 This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 347 Setting the integration time also sets the measurement resolution. Resolution Table shows the rela- tionship between integration time and resolution. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 348 Enabling the scaling function also enables automatic null value selection (SENSe: {RESistance|FRESistance}:NULL:VALue:AUTO ON). To set a fixed null value, use: [SENSe:]{RESistance|FRESistance}:NULL:VALue. The instrument disables the null function after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 349 OFF). To use the null value, the null state must be on ([SENSe:]{RESistance|FRESistance}:NULL:STATe ON). This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 350 The instrument enables automatic null value selection when the null function is enabled ([SENSe:] {RESistance|FRESistance}:NULL:STATe ON). This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 351 Applies only to resistance measurements on the 100Ω, 1 kΩ, 10 kΩ and 100 kΩ ranges. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 352 Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging. If the input signal is greater than can be measured on the specified manual range, the instrument dis- plays the word Overload on front panel and returns "9.9E37" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 353 With autoranging enabled, the instrument selects the range based on the input signal. Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 354 To achieve normal mode (line frequency noise) rejection, use a resolution that corresponds to an integration time that is an integral number of power line cycles. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). See Also [SENSe:]{RESistance|FRESistance}:NPLC Keysight Truevolt Series Operating and Service Guide...
  • Page 355 (including NULL). READ? INITiate to start the measurement. Use [SENSe:]DATA2? to retrieve the secondary measurement. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 356 MEASure:{RESistance|FRESistance}?. Autozero is automatically turned OFF when you select an integration time less than 1 PLC with these commands. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 357 SCPI Programming Reference [SENSe:]TEMPerature Subsystem This subsystem configures temperature measurements. Command Summary [SENSe:]TEMPerature:APERture [SENSe:]TEMPerature:APERture:ENABled [SENSe:]TEMPerature:NPLC [SENSe:]TEMPerature:NULL[:STATe] [SENSe:]TEMPerature:NULL:VALue [SENSe:]TEMPerature:NULL:VALue:AUTO [SENSe:]TEMPerature:SECondary [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:OCOMpensated [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe] [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence] [SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:POWer:LIMit[:STATe] [SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:TYPE [SENSe:]TEMPerature:TRANsducer:TCouple:CHECk [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:OFFSet:ADJust [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:TYPE [SENSe:]TEMPerature:TRANsducer:TCouple:TYPE [SENSe:]TEMPerature:TRANsducer:TYPE [SENSe:]TEMPerature:ZERO:AUTO Keysight Truevolt Series Operating and Service Guide...
  • Page 358 This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 359 Setting the integration time also sets the measurement resolution. Resolution Table shows the rela- tionship between integration time and resolution. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 360 OFF). To use the null value, the null state must be on ([SENSe:]TEMPerature:NULL:STATe ON). This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 361 The instrument enables automatic null value selection when the null function is enabled ([SENSe:]TEMPerature:NULL:STATe ON). This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 362 (thermocouple measurements apply to the 34465A and 34470A only). READ? INITiate to start the measurement. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 363 Typical Response: +1.05451008E+01,+1.07391062E+01 Applies only to the 34465A and 34470A. Affects both 2-wire and 4-wire resistance measurements. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 364 100.1 Ω. Make and read two meas- urements: CONF:TEMP FRTD TEMP:TRAN:FRTD:RES 100.1 SAMP:COUN 2 READ? Typical Response: +1.04530000E+02,+1.04570000E+02 This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 365 TEMP:TRAN:FRTD:RES 100.1 SAMP:COUN 2 READ? Typical Response: +2.54530000E+01,+2.54570000E+01 TCouple applies only to the 34465A and 34470A. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 366 Configure K-type thermocouple measurements. Select fixed reference junction and set reference junction temperature to +20 °C. Enable thermocouple check. Make and read two measurements: CONF:TEMP TC,K TEMP:TRAN:TC:RJUN:TYPE FIX TEMP:TRAN:TC:RJUN 20.0 TEMP:TRAN:TC:CHEC ON SAMP:COUN 2 READ? Typical Response: +3.31785468E+00,+3.61554515E+00 Keysight Truevolt Series Operating and Service Guide...
  • Page 367 This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 368 Celsius regardless of the temperature units currently selected (see UNIT:TEMPerature command). You can read the internal reference junction temperature using the SYSTem:TEMPerature command. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 369 FTHermistor command form makes no difference, as both commands affect the same parameter. Parameter Typical Return 5000 (only available choice) +5000 Configure 4-wire thermistor measurements. Make and read two measurements: CONF:TEMP FTH TEMP:TRAN:FTH:TYPE 5000 SAMP:COUN 2 READ? Typical Response: +1.04530000E+02,+1.04570000E+02 Keysight Truevolt Series Operating and Service Guide...
  • Page 370 MEASure:TEMPerature?. Autozero is automatically turned OFF when you select an integration time less than 1 PLC with these commands. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 371 SCPI Programming Reference [SENSe:]VOLTage Subsystem This subsystem configures AC voltage, DC voltage, and ratio measurements. Command Summary [SENSe:]VOLTage:AC:BANDwidth [SENSe:]VOLTage:{AC|DC}:NULL[:STATe] [SENSe:]VOLTage:{AC|DC}:NULL:VALue [SENSe:]VOLTage:{AC|DC}:NULL:VALue:AUTO [SENSe:]VOLTage:{AC|DC}:RANGe [SENSe:]VOLTage:{AC|DC}:RANGe:AUTO [SENSe:]VOLTage:AC:SECondary [SENSe:]VOLTage[:DC]:APERture [SENSe:]VOLTage[:DC]:APERture:ENABled [SENSe:]VOLTage[:DC]:IMPedance:AUTO [SENSe:]VOLTage[:DC]:NPLC [SENSe:]VOLTage[:DC]:RATio:SECondary [SENSe:]VOLTage[:DC]:RESolution [SENSe:]VOLTage[:DC]:SECondary [SENSe:]VOLTage[:DC]:ZERO:AUTO Keysight Truevolt Series Operating and Service Guide...
  • Page 372 20 Hz - 300 kHz (Medium) 0.6250 s/measurement 200 Hz - 300 kHz (Fast) 0.0250 s/measurement This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 373 To set a fixed null value, use: [SENSe:]VOLTage:{AC|DC}:NULL:VALue. The null function is not available for DC ratio measurements. The instrument disables the null function after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 374 To use the null value, the null state must be on ([SENSe:]VOLTage:{AC|DC}:NULL:STATe ON). The null function is not available for DC ratio measurements. This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 375 The instrument enables automatic null value selection when the null function is enabled ([SENSe:]VOLTage:{AC|DC}:NULL:STATe ON). The null function is not available for DC ratio measurements. This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function. Keysight Truevolt Series Operating and Service Guide...
  • Page 376 Overload on front panel and returns "9.9E37" from the remote interface. The instrument is set to the default range, with autoranging enabled ([SENSe:]VOLTage: {AC|DC}:RANGe:AUTO ON), after a Factory Reset (*RST) or an Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 377 With autoranging enabled, the instrument selects the range based on the input signal. Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 378 (Without the DIG option) 200 µs to 1 s (2 µs resolution), default: 100 ms. (With the +1.00000000E- DIG option) 20 µs to 1 s (2 µs resolution), default: 100 ms. Enable aperture time mode, set the aperture time to 300 ms: VOLT:APER:ENAB ON VOLT:DC:APER 300E-03 Keysight Truevolt Series Operating and Service Guide...
  • Page 379 This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 380 The 100 V and 1000 V ranges remain at a 10 MΩ input impedance. CONFigure MEASure? commands automatically select "AUTO OFF". This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 381 Setting the integration time also sets the measurement resolution. Resolution Table shows the rela- tionship between integration time and resolution. See [SENSe:]VOLTage[:DC]:RESolution. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 382 To achieve normal mode (line frequency noise) rejection, use a resolution that corresponds to an integration time that is an integral number of power line cycles. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 383 AC and DC voltage measurements from remote, program each function separately, see CONFigure[:VOLTage]:{AC|DC}. "PTPeak" - The peak-to-peak, maximum, and minimum values of the input signal. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 384 MEASure:VOLTage:DC:RATio?. Autozero is automatically turned OFF when you select an integration time less than 1 PLC with these commands. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 385 SCPI Programming Reference STATus Subsystem Keysight Truevolt Series Operating and Service Guide...
  • Page 386 The most recent measurement failed the upper limit test. Not Used 8192 (Reserved for future use) Memory Overflow 16384 Reading memory is full. One or more (oldest) measurements have been lost. Not Used 32768 (Reserved for future use) Keysight Truevolt Series Operating and Service Guide...
  • Page 387 Set if any remote interface has an error in its error queue; cleared otherwise. Not Used 16384 (Reserved for future use) Not Used 32768 "0" is returned. Command Summary STATus:OPERation:CONDition? STATus:OPERation:ENABle STATus:OPERation[:EVENt]? STATus:PRESet STATus:QUEStionable:CONDition? STATus:QUEStionable:ENABle STATus:QUEStionable[:EVENt]? Keysight Truevolt Series Operating and Service Guide...
  • Page 388 The <enable> setting is nonvolatile; it does not change after a Factory Reset (*RST) , Instrument Pre- set (SYSTem:PRESet), Status Preset (STATus:PRESet) or Clear Status (*CLS). *PSC command controls whether the enable register is cleared at power on. Keysight Truevolt Series Operating and Service Guide...
  • Page 389 (none) +4096 Read the condition register (bit 12 is set): STAT:QUES:COND? The condition register bits reflect the current condition. If a condition goes away, the corresponding bit is cleared in the condition register. Keysight Truevolt Series Operating and Service Guide...
  • Page 390 Typical Return (none) +1024 Read the event register (bit 10 is set): STAT:QUES? Once a bit is set, it remains set until cleared by reading the event register or by sending *CLS (clear status). Keysight Truevolt Series Operating and Service Guide...
  • Page 391 The SYSTem subsystem includes general commands (listed below), license management commands, and remote interface configuration commands. Command Summary SYSTem:ACALibration:DATE? SYSTem:ACALibration:TEMPerature? SYSTem:ACALibration:TIME? SYSTem:BEEPer[:IMMediate] SYSTem:BEEPer:STATe SYSTem:CLICk:STATe SYSTem:DATE SYSTem:ERRor[:NEXT]? SYSTem:HELP? SYSTem:IDENtify SYSTem:LABel SYSTem:LFRequency? SYSTem:LOCal SYSTem:PRESet SYSTem:SECurity:COUNt? SYSTem:SECurity:IMMediate SYSTem:TEMPerature? SYSTem:TIME SYSTem:UPTime? SYSTem:VERSion? SYSTem:WMESsage Keysight Truevolt Series Operating and Service Guide...
  • Page 392 (34465A/70A only) You can perform this query regardless of whether the instrument is secured. This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. *CAL? to perform autocal. Keysight Truevolt Series Operating and Service Guide...
  • Page 393 (none) (none) Issue a single beep: SYST:BEEP Programmed beeps may be useful for program development and troubleshooting. You can issue a beep with this command regardless of the current beeper state (SYSTem:BEEPer- :STATe). Keysight Truevolt Series Operating and Service Guide...
  • Page 394 Does not affect the beeper for diode and continuity tests, errors and so on. This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 395 The date and time are not affected by *RST or SYSTem:PRESet. The real-time clock does not adjust itself for time zone changes or daylight savings time. Be sure to set the date and time when you first receive your instrument. Keysight Truevolt Series Operating and Service Guide...
  • Page 396 Errors have the following format (the error string may contain up to 255 characters). <error code>,<error string> Where: <error code> = a positive or negative integer <error string> = a quoted ASCII string up to 255 characters See Also *SRE Keysight Truevolt Series Operating and Service Guide...
  • Page 397 Syntax for each command is found in the Command Quick Reference. The list may include commands or queries not described in this document. These are deprecated com- mands included for compatibility with older instruments. Keysight Truevolt Series Operating and Service Guide...
  • Page 398 "Keysight Technologies,34465A, ..." "Keysight Technologies,34470A, ..." If you have an older Agilent 34460A or 34461A and upgrade to the new (Keysight) firmware, your instrument continues to respond with the manufacturer name “Agilent”, not “Keysight” until you set the front panel SCPI ID to 34460A or 34461A, send the...
  • Page 399 The line frequency value is not affected by *RST or SYSTem:PRESet. SYSTem:LOCal Sets the instrument state to local, clears the Remote display annunciator, and enables front-panel oper- ation. Parameter Typical Return (none) (none) Set the instrument state to local: SYST:LOC Keysight Truevolt Series Operating and Service Guide...
  • Page 400 The act of securing or unsecuring the instrument does not increment the counter. Be sure to read the security count upon receiving your instrument. See Also CALibration:COUNt? Keysight Truevolt Series Operating and Service Guide...
  • Page 401 MMEMory Subsystem - STATe and PREFerence Files). Increments the security count. SYSTem:TEMPerature? Returns the instrument's internal temperature in °C. Parameter Typical Return (none) +2.85000000E+01 Return the instrument's internal temperature: SYST:TEMP? The value returned is not affected by UNIT:TEMPerature. Keysight Truevolt Series Operating and Service Guide...
  • Page 402 Returns the version of SCPI (Standard Commands for Programmable Instruments) that the instrument complies with. There is no front panel equivalent. See Introduction to the Introduction to the SCPI Lan- guage for details. Parameter Typical Return (none) 1994.0 Return the SCPI version: SYST:VERS? Keysight Truevolt Series Operating and Service Guide...
  • Page 403 This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). This parameter is set to its default value when the instrument is shipped from the factory and after SYSTem:SECurity:IMMediate. Keysight Truevolt Series Operating and Service Guide...
  • Page 404 I/O locking commands, license management commands, and remote interface configuration commands (listed below). Command Summary SYSTem:COMMunicate:ENABle SYSTem:COMMunicate:GPIB:ADDRess SYSTem:COMMunicate:LAN:CONTrol? SYSTem:COMMunicate:LAN:DHCP SYSTem:COMMunicate:LAN:DNS[{1|2}] SYSTem:COMMunicate:LAN:DOMain? SYSTem:COMMunicate:LAN:GATeway SYSTem:COMMunicate:LAN:HOSTname SYSTem:COMMunicate:LAN:IPADdress SYSTem:COMMunicate:LAN:MAC? SYSTem:COMMunicate:LAN:SMASk SYSTem:COMMunicate:LAN:TELNet:PROMpt SYSTem:COMMunicate:LAN:TELNet:WMESsage SYSTem:COMMunicate:LAN:UPDate SYSTem:COMMunicate:LAN:WINS[{1|2}] SYSTem:USB:HOST:ENABle Keysight Truevolt Series Operating and Service Guide...
  • Page 405 For HiSLIP, LAN, SOCKets, TELNet, VXI11, or WEB on the 34460A, you must have option 34460A-LAN or option 3446LANU. The interface enable settings are stored in nonvolatile memory. They do not change when power has been off, after a Factory Reset (*RST) or after an Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 406 This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). This parameter is set to its default value when the instrument is shipped from the factory and after SYSTem:SECurity:IMMediate. Keysight Truevolt Series Operating and Service Guide...
  • Page 407 If a DHCP LAN address is not assigned by a DHCP server, then an Auto-IP address is obtained after approximately 2 minutes. An Auto-IP address has the form 169.254.nnn.nnn. This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 408 Dynamic DNS service at power-on. A null string ("") indicates that no domain name is assigned. On the 34460A, this requires option 34460A-LAN or option 3446LANU. Keysight Truevolt Series Operating and Service Guide...
  • Page 409 However, if the DHCP server fails to assign a valid address, the currently configured default gateway is used. This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 410 This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). This parameter is set to its default value when the instrument is shipped from the factory and after SYSTem:SECurity:IMMediate. Keysight Truevolt Series Operating and Service Guide...
  • Page 411 The MAC address is also known as the link-layer address, the Ethernet (station) address, LANIC ID or Hardware Address. This is an unchangeable 48-bit address assigned by the manufacturer to each unique Internet device. Keysight Truevolt Series Operating and Service Guide...
  • Page 412 STATic: returns static address from non-volatile memory. This address is used if DHCP is disabled or unavailable. This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 413 To exit a Telnet session, press <Ctrl-D>. This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). SYSTem:SECurity:IMMediate sets this parameter to its default value. Keysight Truevolt Series Operating and Service Guide...
  • Page 414 Typical Return Quoted string of up to 63 characters. "Welcome to the Telnet Ses- sion" Default: "Welcome to Keysight's <instrument model number> Digital Multimeter" Define a welcome message: SYST:COMM:LAN:TELN:WMES "Welcome to the Telnet Session" On the 34460A, this requires option 34460A-LAN or option 3446LANU.
  • Page 415 The following example configures the instrument to use statically assigned LAN settings. SYST:COMM:LAN:DHCP OFF SYST:COMM:LAN:DNS "198.105.232.4" SYST:COMM:LAN:DNS2 "198.105.232.5" SYST:COMM:LAN:GAT "198.105.232.1" SYST:COMM:LAN:HOST "LAB1-DMM" SYST:COMM:LAN:IPAD "198.105.232.101" SYST:COMM:LAN:SMAS "255.255.255.0" SYST:COMM:LAN:WINS "198.105.232.4" SYST:COMM:LAN:WINS "198.105.232.5" SYST:COMM:LAN:UPD The following example configures the instrument back to use DHCP. SYST:COMM:LAN:DHCP ON SYST:COMM:LAN:UPD Keysight Truevolt Series Operating and Service Guide...
  • Page 416 CURRent: returns address currently being used by the instrument. STATic: returns static address from non-volatile memory. This address is used if DHCP is disabled or unavailable. Keysight Truevolt Series Operating and Service Guide...
  • Page 417 If the front panel USB host port is disabled, it cannot be used to store or load measurements, state files, preferences or other files or to perform firmware update. This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru- ment Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 418 State = locked, Count = 1 (FROM USB) SYST:LOCK:REL State = unlocked, Count = 0 For each successful lock request, a lock release is required. Two requests require two releases. Command Summary SYSTem:LOCK:NAME? SYSTem:LOCK:OWNer? SYSTem:LOCK:RELease SYSTem:LOCK:REQuest? Keysight Truevolt Series Operating and Service Guide...
  • Page 419 Interface Locking Example When a lock is active, Bit 10 in the Standard Operation Register will be set (STATus:OPERa- tion:CONDition?). When the lock is released on all I/O interfaces, this bit will be cleared. Keysight Truevolt Series Operating and Service Guide...
  • Page 420 Granting a lock sets Bit 10 in the Standard Operation Register (STATus:OPERation:CONDition?). In addition, the entire instrument front panel, including the Local key, is locked out while a lock is in place ("Front panel locked." is displayed). Keysight Truevolt Series Operating and Service Guide...
  • Page 421 If no licensed options are found, the query returns "". Parameter Typical Return (none) "LAN","SEC" Return currently licensed options: SYST:LIC:CAT? This query returns a string with the following format: 34460A: "GPB","LAN","SEC" 34461A: "GPB","SEC" 34465A/70A: "GPB","SEC","DIG","MEM" Options vary by model, as shown here. Keysight Truevolt Series Operating and Service Guide...
  • Page 422 "DIG", "MEM", or "SEC" (34465A/70A) Return description for option 34460A-LAN or option 3446LANU: SYST:LIC:DESC? "LAN" Valid option names are double quoted strings representing the installed licensed options. Installed options can be easily identified using SYSTem:LICense:CATalog?. Keysight Truevolt Series Operating and Service Guide...
  • Page 423 Returns a definite-length block containing multi-line ASCII text, including carriage returns and line feeds. SYSTem:LICense:ERRor:COUNt? Returns the number of license errors generated by SYSTem:LICense:INSTall. Parameter Typical Return (none) Return number of license errors: SYST:LIC:ERR:COUN? Keysight Truevolt Series Operating and Service Guide...
  • Page 424 "\" or "/" and start at the root folder of <drive>. If <drive> is omitted, <path> is relative to the folder specified by MMEMory:CDIRectory. Relative paths must NOT begin with \ or /. Keysight Truevolt Series Operating and Service Guide...
  • Page 425 SCPI Programming Reference The <folder> parameter cannot exceed 240 characters. The specified folder must exist and cannot be marked as hidden or system. Keysight Truevolt Series Operating and Service Guide...
  • Page 426 Recommendation: All triggered measurements should be made using an appropriate fixed manual range. That is, turn autorange off ([SENSe:]<func- tion>:RANGe:AUTO OFF) or set a fixed range using the [SENSe:]<function>:RANGe, CONFigure, or MEASure command. Command Summary TRIGger:COUNt TRIGger:DELay TRIGger:DELay:AUTO TRIGger:LEVel TRIGger:SLOPe TRIGger:SOURce Keysight Truevolt Series Operating and Service Guide...
  • Page 427 (front panel), the instrument ignores the trigger count setting; however, when you return to remote, the trigger count returns to the previous value you selected. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 428 If you have configured the instrument for more than one measurement per trigger (SAMPle:COUNt >1), the delay is inserted after the trigger and between consecutive measurements. The instrument selects automatic trigger delay after a Factory Reset (*RST) or an Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 429 Supported only for ac and dc voltage, ac and dc current, frequency, period, 2-wire resistance, 4-wire resistance, RTD and thermistor temperature measurements. Resistance and temperature meas- urements must have offset compensation and low power disabled. Keysight Truevolt Series Operating and Service Guide...
  • Page 430 0 to +3.6 A ange 10 A terminals 0 to +12 A FREQuency *N/A 3 Hz to 300 kHz PERiod *N/A 3.33333333 µs to 0.333333333 s *N/A = Not Applicable. See Also TRIGger:DELay TRIGger:SOURce Keysight Truevolt Series Operating and Service Guide...
  • Page 431 Typical Response: +1.00520000E+01, … (50 measurements) On the 34460A, this requires option 34460A-LAN or option 3446LANU. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 432 SCPI Programming Reference TRIGger:SOURce {IMMediate|EXTernal|BUS|INTernal} TRIGger:SOURce? Keysight Truevolt Series Operating and Service Guide...
  • Page 433 That is, turn autorange off ([SENSe:]<function>:RANGe:AUTO OFF) or set a fixed range using the [SENSe:]<function>:RANGe, CONFigure, or MEASure command. This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 434 MEASure:CURRent:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]] MEASure:DIODe? MEASure:{FREQuency|PERiod}? [{<range>|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]] MEASure:{RESistance|FRESistance}? [{<range>|AUTO|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]] MEASure:TEMPerature? [{FRTD|RTD|FTHermistor|THERmistor|TCouple|DEFault} [, {<type>|DEFault} [,1 , {<resolution>|MIN|MAX|DEF}]]]] MEASure[:VOLTage]:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]] MEASure[:VOLTage][:DC]:RATio? [{<range>|AUTO|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]] Measurement Configuration Commands [SENSe:]FUNCtion[:ON] "<function>" [SENSe:]FUNCtion[:ON]? Keysight Truevolt Series Operating and Service Guide...
  • Page 435 [SENSe:]VOLTage:{AC|DC}:RANGe:AUTO {OFF|ON|ONCE} [SENSe:]VOLTage:{AC|DC}:RANGe:AUTO? [SENSe:]VOLTage:AC:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"VOLTage[:DC]"} [SENSe:]VOLTage:AC:SECondary? [SENSe:]VOLTage[:DC]:APERture {<seconds>|MIN|MAX|DEF} [SENSe:]VOLTage[:DC]:APERture? [{MIN|MAX|DEF}] [SENSe:]VOLTage[:DC]:APERture:ENABled {OFF|ON} [SENSe:]VOLTage[:DC]:APERture:ENABled? [SENSe:]VOLTage[:DC]:IMPedance:AUTO {OFF|ON} [SENSe:]VOLTage[:DC]:IMPedance:AUTO? [SENSe:]VOLTage[:DC]:NPLC {<PLCs>|MIN|MAX|DEF} [SENSe:]VOLTage[:DC]:NPLC? [{MIN|MAX|DEF}] [SENSe:]VOLTage[:DC]:RATio:SECondary {"OFF"|"CALCulate:DATA"|"SENSe:DATA"} [SENSe:]VOLTage[:DC]:RATio:SECondary? [SENSe:]VOLTage[:DC]:RESolution {<resolution>|MIN|MAX|DEF} [SENSe:]VOLTage[:DC]:RESolution? [{MIN|MAX|DEF}] [SENSe:]VOLTage[:DC]:SECondary {"OFF"|"CALCulate:DATA"|"VOLTage:AC"|"PTPeak"} [SENSe:]VOLTage[:DC]:SECondary? [SENSe:]VOLTage[:DC]:ZERO:AUTO {OFF|ON|ONCE} [SENSe:]VOLTage[:DC]:ZERO:AUTO? Keysight Truevolt Series Operating and Service Guide...
  • Page 436 [SENSe:]{RESistance|FRESistance}:NULL[:STATe] {OFF|ON} [SENSe:]{RESistance|FRESistance}:NULL[:STATe]? [SENSe:]{RESistance|FRESistance}:NULL:VALue {<value>|MIN|MAX|DEF} [SENSe:]{RESistance|FRESistance}:NULL:VALue? [{MIN|MAX|DEF}] [SENSe:]{RESistance|FRESistance}:NULL:VALue:AUTO {OFF|ON} [SENSe:]{RESistance|FRESistance}:NULL:VALue:AUTO? [SENSe:]{RESistance|FRESistance}:OCOMpensated {OFF|ON} [SENSe:]{RESistance|FRESistance}:OCOMpensated? [SENSe:]{RESistance|FRESistance}:POWer:LIMit[:STATe] {OFF|ON} [SENSe:]{RESistance|FRESistance}:POWer:LIMit[:STATe]? [SENSe:]{RESistance|FRESistance}:RANGe {<range>|MIN|MAX|DEF} [SENSe:]{RESistance|FRESistance}:RANGe? [{MIN|MAX|DEF}] [SENSe:]{RESistance|FRESistance}:RANGe:AUTO {OFF|ON|ONCE} [SENSe:]{RESistance|FRESistance}:RANGe:AUTO? [SENSe:]{RESistance|FRESistance}:RESolution {<resolution>|MIN|MAX|DEF} [SENSe:]{RESistance|FRESistance}:RESolution? [{MIN|MAX|DEF}] [SENSe:]{FRESistance|RESistance}:SECondary {"OFF"|"CALCulate:DATA"} [SENSe:]{FRESistance|RESistance}:SECondary? [SENSe:]RESistance:ZERO:AUTO {OFF|ON|ONCE} [SENSe:]RESistance:ZERO:AUTO? Keysight Truevolt Series Operating and Service Guide...
  • Page 437 [SENSe:]CURRent:{AC|DC}:RANGe:AUTO {OFF|ON|ONCE} [SENSe:]CURRent:{AC|DC}:RANGe:AUTO? [SENSe:]CURRent:{AC|DC}:TERMinals {3|10} [SENSe:]CURRent:{AC|DC}:TERMinals? [SENSe:]CURRent:AC:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"CURRent[:DC]"} [SENSe:]CURRent:AC:SECondary? [SENSe:]CURRent[:DC]:APERture {<seconds>|MIN|MAX|DEF} [SENSe:]CURRent[:DC]:APERture? [{MIN|MAX|DEF}] [SENSe:]CURRent[:DC]:APERture:ENABled {OFF|ON} [SENSe:]CURRent[:DC]:APERture:ENABled? [SENSe:]CURRent[:DC]:NPLC {<PLCs>|MIN|MAX|DEF} [SENSe:]CURRent[:DC]:NPLC? [{MIN|MAX|DEF}] [SENSe:]CURRent[:DC]:RESolution {<resolution>|MIN|MAX|DEF} [SENSe:]CURRent[:DC]:RESolution? [{MIN|MAX|DEF}] [SENSe:]CURRent[:DC]:SECondary {"OFF"|"CALCulate:DATA"|"CURRent:AC"|"PTPeak"} [SENSe:]CURRent[:DC]:SECondary? [SENSe:]CURRent[:DC]:ZERO:AUTO {OFF|ON|ONCE} [SENSe:]CURRent[:DC]:ZERO:AUTO? [SENSe:]CURRent:SWITch:MODE {FAST|CONTinuous} [SENSe:]CURRent:SWITch:MODE? Keysight Truevolt Series Operating and Service Guide...
  • Page 438 [SENSe:]TEMPerature:NPLC {<PLCs>|MIN|MAX|DEF} [SENSe:]TEMPerature:NPLC? [{MIN|MAX|DEF}] [SENSe:]TEMPerature:NULL[:STATe] {OFF|ON} [SENSe:]TEMPerature:NULL[:STATe]? [SENSe:]TEMPerature:NULL:VALue {<value>|MIN|MAX|DEF} [SENSe:]TEMPerature:NULL:VALue? [{MIN|MAX|DEF}] [SENSe:]TEMPerature:NULL:VALue:AUTO {OFF|ON} [SENSe:]TEMPerature:NULL:VALue:AUTO? [SENSe:]TEMPerature:SECondary {"OFF"|"CALCulate:DATA"|"SENSe:DATA"} [SENSe:]TEMPerature:SECondary? [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:OCOMpensated {OFF|ON} [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:OCOMpensated? [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe] {OFF|ON} [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe]? [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence] {<reference>|MIN|MAX|DEF} [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence]? [{MIN|MAX|DEF}] [SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:POWer:LIMit[:STATe] {OFF|ON} [SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:POWer:LIMit[:STATe]? [SENSe:]TEMPerature:TRANsducer:TCouple:CHECk {OFF|ON} [SENSe:]TEMPerature:TRANsducer:TCouple:CHECk? Keysight Truevolt Series Operating and Service Guide...
  • Page 439 [SENSe:]{FREQuency|PERiod}:NULL[:STATe]? [SENSe:]{FREQuency|PERiod}:NULL:VALue {<value>|MIN|MAX|DEF} [SENSe:]{FREQuency|PERiod}:NULL:VALue? [{MIN|MAX|DEF}] [SENSe:]{FREQuency|PERiod}:NULL:VALue:AUTO {OFF|ON} [SENSe:]{FREQuency|PERiod}:NULL:VALue:AUTO? [SENSe:]{FREQuency|PERiod}:RANGe:LOWer {<freq>|MIN|MAX|DEF} [SENSe:]{FREQuency|PERiod}:RANGe:LOWer? [SENSe:]{FREQuency|PERiod}:TIMeout:AUTO {OFF|ON} [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe {<range>|MIN|MAX|DEF} [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe? [{MIN|MAX|DEF}] [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe:AUTO {OFF|ON|ONCE} [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe:AUTO? [SENSe:]FREQuency:SECondary {"OFF"|"CALCulate:DATA"|"PERiod"|"VOLTage:AC"} [SENSe:]FREQuency:SECondary? [SENSe:]PERiod:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"VOLTage:AC"} [SENSe:]PERiod:SECondary? Continuity and Diode Configuration Commands CONFigure:CONTinuity CONFigure:DIODe Keysight Truevolt Series Operating and Service Guide...
  • Page 440 [SENSe:]CAPacitance:SECondary {"OFF"|"CALCulate:DATA"} [SENSe:]CAPacitance:SECondary? [SENSe:]CURRent:AC:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"CURRent[:DC]"} [SENSe:]CURRent:AC:SECondary? [SENSe:]CURRent[:DC]:SECondary {"OFF"|"CALCulate:DATA"|"CURRent:AC"|"PTPeak"} [SENSe:]CURRent[:DC]:SECondary? [SENSe:]DATA2? [SENSe:]DATA2:CLEar[:IMMediate] [SENSe:]{FRESistance|RESistance}:SECondary {"OFF"|"CALCulate:DATA"} [SENSe:]{FRESistance|RESistance}:SECondary? [SENSe:]FREQuency:SECondary {"OFF"|"CALCulate:DATA"|"PERiod"|"VOLTage:AC"} [SENSe:]PERiod:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"VOLTage:AC"} [SENSe:]TEMPerature:SECondary {"OFF"|"CALCulate:DATA"|"SENSe:DATA"} [SENSe:]TEMPerature:SECondary? [SENSe:]VOLTage:AC:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"VOLTage[:DC]"} [SENSe:]VOLTage:AC:SECondary? [SENSe:]VOLTage[:DC]:RATio:SECondary {"OFF"|"CALCulate:DATA"|"SENSe:DATA"} [SENSe:]VOLTage[:DC]:RATio:SECondary? [SENSe:]VOLTage[:DC]:SECondary {"OFF"|"CALCulate:DATA"|"VOLTage:AC"|"PTPeak"} [SENSe:]VOLTage[:DC]:SECondary? Miscellaneous Configuration Commands ROUTe:TERMinals? Keysight Truevolt Series Operating and Service Guide...
  • Page 441 OUTPut:TRIGger:SLOPe {POSitive|NEGative} OUTPut:TRIGger:SLOPe? READ? SAMPle:COUNt {<count>|MIN|MAX|DEF} SAMPle:COUNt? [{MIN|MAX|DEF}] SAMPle:COUNt:PRETrigger {<count>|MIN|MAX|DEF} SAMPle:COUNt:PRETrigger? [{MIN|MAX|DEF}] *TRG TRIGger:COUNt {<count>|MIN|MAX|DEF|INFinity} TRIGger:COUNt? [{MIN|MAX|DEF}] TRIGger:DELay {<seconds>|MIN|MAX|DEF} TRIGger:DELay? [{MIN|MAX|DEF}] TRIGger:DELay:AUTO {OFF|ON} TRIGger:DELay:AUTO? TRIGger:LEVel {<level>|MIN|MAX|DEF} TRIGger:LEVel? [{MIN|MAX|DEF}] TRIGger:SLOPe {POSitive|NEGative} TRIGger:SLOPe? TRIGger:SOURce {IMMediate|EXTernal|BUS|INTernal} TRIGger:SOURce? Keysight Truevolt Series Operating and Service Guide...
  • Page 442 Calculation (Math) Commands Overall CALCulate:CLEar[:IMMediate] Histogram CALCulate:TRANsform:HISTogram:ALL? CALCulate:TRANsform:HISTogram:CLEar[:IMMediate] CALCulate:TRANsform:HISTogram:COUNt? CALCulate:TRANsform:HISTogram:DATA? CALCulate:TRANsform:HISTogram:POINts {<value>|MIN|MAX|DEF} CALCulate:TRANsform:HISTogram:POINts? [{MIN|MAX|DEF}] CALCulate:TRANsform:HISTogram:RANGe:AUTO {OFF|ON} CALCulate:TRANsform:HISTogram:RANGe:AUTO? CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer} {<value>|MIN|MAX|DEF} CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer}? [{MIN|MAX|DEF}] CALCulate:TRANsform:HISTogram[:STATe] {OFF|ON} CALCulate:TRANsform:HISTogram[:STATe]? Limit Testing CALCulate:LIMit:CLEar[:IMMediate] CALCulate:LIMit:{LOWer|UPPer}[:DATA] {<value>|MIN|MAX|DEF} CALCulate:LIMit:{LOWer|UPPer}[:DATA]? [{MIN|MAX|DEF}] CALCulate:LIMit[:STATe] {OFF|ON} CALCulate:LIMit[:STATe]? Keysight Truevolt Series Operating and Service Guide...
  • Page 443 CALCulate:SCALe:DB:REFerence? [{MIN|MAX|DEF}] CALCulate:SCALe:DBM:REFerence {<reference>|MIN|MAX|DEF} CALCulate:SCALe:DBM:REFerence? [{MIN|MAX|DEF}] CALCulate:SCALe:FUNCtion {DB|DBM|PCT|SCALe} CALCulate:SCALe:FUNCtion? CALCulate:SCALe:GAIN {<gain>|MIN|MAX|DEF} CALCulate:SCALe:GAIN? [{MIN|MAX|DEF}] CALCulate:SCALe:OFFSet {<offset>|MIN|MAX|DEF} CALCulate:SCALe:OFFSet? [{MIN|MAX|DEF}] CALCulate:SCALe:REFerence:AUTO {OFF|ON} CALCulate:SCALe:REFerence:AUTO? CALCulate:SCALe:REFerence {<reference>|MIN|MAX|DEF} CALCulate:SCALe:REFerence? [{MIN|MAX|DEF}] CALCulate:SCALe[:STATe] {OFF|ON} CALCulate:SCALe[:STATe]? CALCulate:SCALe:UNIT <quoted_string> CALCulate:SCALe:UNIT? CALCulate:SCALe:UNIT:STATe {OFF|ON} CALCulate:SCALe:UNIT:STATe? Keysight Truevolt Series Operating and Service Guide...
  • Page 444 CALCulate:AVERage:MAXimum? CALCulate:AVERage:MINimum? CALCulate:AVERage:PTPeak? CALCulate:AVERage:SDEViation? CALCulate:AVERage[:STATe] {OFF|ON} CALCulate:AVERage[:STATe]? Smoothing CALCulate:SMOothing:RESPonse {SLOW|MEDium|FAST} CALCulate:SMOothing:RESPonse? CALCulate:SMOothing[:STATe] {OFF|ON} CALCulate:SMOothing[:STATe]? Trend Chart CALCulate:TCHart[:STATe] {OFF|ON} CALCulate:TCHart[:STATe]? Reading Memory Commands DATA:LAST? DATA:POINts:EVENt:THReshold <count> DATA:POINts:EVENt:THReshold? DATA:POINts? DATA:REMove? <num_readings> [,WAIT] FETCh? R? [<max_readings>] Keysight Truevolt Series Operating and Service Guide...
  • Page 445 CALibration:TIME? CALibration:VALue <value> CALibration:VALue? SYSTem:ACALibration:DATE? SYSTem:ACALibration:TEMPerature? SYSTem:ACALibration:TIME? State Storage and Preferences Commands *LRN? MMEMory:LOAD:PREFerences <file> MMEMory:LOAD:STATe <file> MMEMory:STORe:PREFerences <file> MMEMory:STORe:STATe <file> MMEMory:STATe:RECall:AUTO {OFF|ON} MMEMory:STATe:RECall:AUTO? MMEMory:STATe:RECall:SELect <file> MMEMory:STATe:RECall:SELect? MMEMory:STATe:VALid? <file> *RCL {0|1|2|3|4} *SAV {0|1|2|3|4} Keysight Truevolt Series Operating and Service Guide...
  • Page 446 SCPI Programming Reference General Purpose File Management Commands MMEMory:CATalog[:ALL]? [<folder>[<filespec>]] MMEMory:CDIRectory <folder> MMEMory:CDIRectory? MMEMory:COPY <file1>, <file2> MMEMory:DELete {<file>|<filespec>} MMEMory:MDIRectory <folder> MMEMory:MOVE <file1>, <file2> MMEMory:RDIRectory <folder> Keysight Truevolt Series Operating and Service Guide...
  • Page 447 MMEMory:STORe:DATA RDG_STORE, <file> MMEMory:UPLoad? <file> IEEE-488 Commands *CAL? *CLS *ESE <enable_value> *ESE? *ESR? *IDN? *LRN? *OPC *OPC? *OPT? *PSC {0|1} *PSC? *RCL {0|1|2|3|4} *RST *SAV {0|1|2|3|4} *SRE <enable_value> *SRE? *STB? *TRG *TST? *WAI Keysight Truevolt Series Operating and Service Guide...
  • Page 448 HCOPy:SDUMp:DATA:FORMat {PNG|BMP} HCOPy:SDUMp:DATA:FORMat? HCOPy:SDUMp:DATA? *IDN? LXI:IDENtify[:STATe] {OFF|ON} LXI:IDENtify[:STATe]? LXI:MDNS:ENABle {OFF|ON} LXI:MDNS:ENABle? LXI:MDNS:HNAMe[:RESolved]? LXI:MDNS:SNAMe:DESired "<name>" LXI:MDNS:SNAMe:DESired? LXI:MDNS:SNAMe[:RESolved]? LXI:RESet LXI:RESTart *RST SYSTem:ACALibration:DATE? SYSTem:ACALibration:TEMPerature? SYSTem:ACALibration:TIME? SYSTem:BEEPer[:IMMediate] SYSTem:BEEPer:STATe {OFF|ON} SYSTem:BEEPer:STATe? SYSTem:CLICk:STATe {OFF|ON} SYSTem:CLICk:STATe? SYSTem:DATE <year>, <month>, <day> SYSTem:DATE? Keysight Truevolt Series Operating and Service Guide...
  • Page 449 SYSTem:SECurity:IMMediate SYSTem:TEMPerature? SYSTem:TIME <hour>, <minute>, <second> SYSTem:TIME? SYSTem:UPTime? SYSTem:USB:HOST:ENABle {OFF|ON} SYSTem:USB:HOST:ENABle? SYSTem:VERSion? SYSTem:WMESsage "<string>" SYSTem:WMESsage? TEST:ALL? *TST? *Parameters vary by DMM model number. SeeSYSTem:IDENtify for details. Interface Locking Commands SYSTem:LOCK:NAME? SYSTem:LOCK:OWNer? SYSTem:LOCK:RELease SYSTem:LOCK:REQuest? Keysight Truevolt Series Operating and Service Guide...
  • Page 450 SYSTem:COMMunicate:LAN:DNS[{1|2}]? [{CURRent|STATic}] SYSTem:COMMunicate:LAN:DOMain? SYSTem:COMMunicate:LAN:GATeway "<address>" SYSTem:COMMunicate:LAN:GATeway? [{CURRent|STATic}] SYSTem:COMMunicate:LAN:HOSTname "<name>" SYSTem:COMMunicate:LAN:HOSTname? [{CURRent|STATic}] SYSTem:COMMunicate:LAN:IPADdress "<address>" SYSTem:COMMunicate:LAN:IPADdress? [{CURRent|STATic}] SYSTem:COMMunicate:LAN:MAC? SYSTem:COMMunicate:LAN:SMASk "<mask>" SYSTem:COMMunicate:LAN:SMASk? [{CURRent|STATic}] SYSTem:COMMunicate:LAN:TELNet:PROMpt "<string>" SYSTem:COMMunicate:LAN:TELNet:PROMpt? SYSTem:COMMunicate:LAN:TELNet:WMESsage "<string>" SYSTem:COMMunicate:LAN:TELNet:WMESsage? SYSTem:COMMunicate:LAN:UPDate SYSTem:COMMunicate:LAN:WINS[{1|2}] "<address>" SYSTem:COMMunicate:LAN:WINS[{1|2}]? [{CURRent|STATic}] SYSTem:USB:HOST:ENABle {OFF|ON} SYSTem:USB:HOST:ENABle? Keysight Truevolt Series Operating and Service Guide...
  • Page 451 SCPI Programming Reference Status System Commands *CLS *ESE <enable_value> *ESE? *ESR? *PSC {0|1} *PSC? *SRE <enable_value> *SRE? STATus:OPERation:CONDition? STATus:OPERation:ENABle <enable_value> STATus:OPERation:ENABle? STATus:OPERation[:EVENt]? STATus:PRESet STATus:QUEStionable:CONDition? STATus:QUEStionable:ENABle <enable_value> STATus:QUEStionable:ENABle? STATus:QUEStionable[:EVENt]? *STB? Keysight Truevolt Series Operating and Service Guide...
  • Page 452 1E-2 1E-1 10 k 3E-2 1E-1 3E-1 3E-3 1E-2 3E-2 1E-1 100 k 3E-1 3E-2 1E-1 3E-1 3E-1 10 M 100 M 10 k 30 k 10 k ResFactor x Range = Resolution. Keysight Truevolt Series Operating and Service Guide...
  • Page 453 1E-3 3E-3 5E-3 1E-2 3E-2 10E-2 30E-2 100 k 1E-3 3E-3 1E-2 3E-2 5E-2 1E-1 3E-1 10E-1 30E-1 1E-2 3E-2 1E-1 3E-1 5E-1 10 M 1E-1 3E-1 100 M 10 k 30 k Keysight Truevolt Series Operating and Service Guide...
  • Page 454 SCPI Programming Reference ResFactor x Range = Resolution., 34465A/70A with DIG option only. Keysight Truevolt Series Operating and Service Guide...
  • Page 455 (DEFault) (DEFault) (DEFault) Range Range 100 PLC 3 ppm × Range 0.3 ppm × Range 0.03 ppm x 0.01 ppm x 1.67 s (MAXimum) (MINimum) (MINimum) Range Range 34465A/70A with DIG option only. Keysight Truevolt Series Operating and Service Guide...
  • Page 456 20 µs 40 µs 100 µs 0.3 ms 1 ms 3 ms Aperture (50 Hz Power) 20 ms 0.2 s Range Delay 1 ms 1.5 ms 34465A/70A with DIG option only. 34465A/70A only. Keysight Truevolt Series Operating and Service Guide...
  • Page 457 1 MΩ 10 ms 15 ms 10 MΩ 100 MΩ 100 ms 1 GΩ 34465A/70A with DIG option only. 34465A/70A only. AC Voltage Default Delays Filter (Hz) Range 2.5 s 0.625 s 0.025 s Keysight Truevolt Series Operating and Service Guide...
  • Page 458 SCPI Programming Reference AC Current Default Delays Filter (Hz) Range 1.66 s 0.25 s 0.025 s Frequency and Period The default delay for frequency and period is 1 s, regardless of aperture (gate time) setting. Keysight Truevolt Series Operating and Service Guide...
  • Page 459 The rear-panel VM Comp (voltmeter complete) output provides a 3.3 V pulse after each measurement completion to implement a standard hardware handshake sequence between measurement and switch- ing devices. The signal shown is for a negative pulse. Keysight Truevolt Series Operating and Service Guide...
  • Page 460 (Clear Status) command and when power is cycled. All errors are cleared when you read the error queue. The error queue is not cleared by a Factory Reset (*RST) or an Instrument Preset (SYSTem:PRESet). Keysight Truevolt Series Operating and Service Guide...
  • Page 461 "arrow" keys to view the error messages. Press Back to exit and clear the error queue. SCPI: SYSTem:ERRor? Read and clear one error from the queue Errors have the following format (the error string may contain up to 80 characters): -113,"Undefined header" Keysight Truevolt Series Operating and Service Guide...
  • Page 462 -130,"Suffix error" -131,"Invalid suffix" -134,"Suffix too long" -138,"Suffix not allowed" -140,"Character data error" -141,"Invalid character data" -144,"Character data too long" -148,"Character data not allowed" -150,"String data error" -151,"Invalid string data" -158,"String data not allowed" Keysight Truevolt Series Operating and Service Guide...
  • Page 463 SCPI Programming Reference -160,"Block data error" -161,"Invalid block data" -168,"Block data not allowed" Keysight Truevolt Series Operating and Service Guide...
  • Page 464 -221,"Settings conflict; limit test not allowed in current measurement function; CALC:LIM:STAT OFF set" -221,"Settings conflict; lower limit > upper limit; CALC:LIM:UPP set to CALC:LIM:LOW value" -221,"Settings conflict; offset compensation not allowed with low power ohms; low power disabled" Keysight Truevolt Series Operating and Service Guide...
  • Page 465 -240,"Hardware error; cannot communicate with measurement hardware" -240,"Hardware error; CPU board initialization failed" -240,"Hardware error; GPIB interface failed" -240,"Hardware error; measurement FPGA FIFO overflow" -240,"Hardware error; measurement hardware initialization failed" -241,"Hardware missing" -250,"Mass storage error: file read/write error" Keysight Truevolt Series Operating and Service Guide...
  • Page 466 -257,"File name error; path is a folder name" -257,"File name error; path name missing" -257,"File name error; path too long" -257,"File name error; relative path not allowed" -257,"File name error; unknown file extension" -292,"Referenced name does not exist" Keysight Truevolt Series Operating and Service Guide...
  • Page 467 -315,"Configuration memory lost; due to firmware revision change" -315,"Configuration memory lost; memory corruption detected" -350,"Queue overflow" Query Errors (-400…) -400,"Query error" -410,"Query INTERRUPTED" -420,"Query UNTERMINATED" -430,"Query DEADLOCKED" -440,"Query UNTERMINATED after indefinite response" Network Errors (+100...) +100,"Network Error" +110,"LXI mDNS Error" Keysight Truevolt Series Operating and Service Guide...
  • Page 468 +514,"Not allowed; Instrument locked by another I/O session" +532,"Not able to achieve requested resolution" +540,"Cannot use overload as math reference" +541,"Cannot use zero as math reference for PCT scaling function" +542,"Measured dB reference too small" Keysight Truevolt Series Operating and Service Guide...
  • Page 469 +628,"Self-test failed; Fine ADC off-scale low" +629,"Self-test failed; Fine ADC off-scale high" +630,"Self-test failed; Fine ADC range" +631,"Self-test failed; Fine ADC bits stuck" +632,"Self-test failed; Fine ADC slope" +633,"Self-test failed; Fine ADC linearity" +634,"Self-test failed; low DC zero" Keysight Truevolt Series Operating and Service Guide...
  • Page 470 +663,"Self-test failed; autocal 1 MOhm to 10 MOhm transfer" +664,"Self-test failed; autocal 10 MOhm internal resistor" +665,"Self-test aborted; high ac voltage present on input terminals" +666,"Self-test aborted; high dc voltage present on input terminals" Keysight Truevolt Series Operating and Service Guide...
  • Page 471 +713,"Calibration error; ADC calibration failed; fine merge for setup 'DCV'" +713,"Calibration error; ADC calibration failed; fine offset for setup 'ACI'" +713,"Calibration error; ADC calibration failed; fine offset for setup 'ACV'" +713,"Calibration error; ADC calibration failed; fine offset for setup 'DCV'" Keysight Truevolt Series Operating and Service Guide...
  • Page 472 +735,"Calibration error; 1 kHz AC flatness frequency point must be last frequency point in sequence" +736,"Calibration error; DCI 100uA range self-cal failed" +740,"Calibration data lost: secure state" +740,"Calibration information lost; count, security state, security code, string" Keysight Truevolt Series Operating and Service Guide...
  • Page 473 +810,"State has not been stored" +820,"Model and serial numbers not restored" +821,"Controller and measurement board model numbers do not match" +822,"Controller and measurement board serial numbers do not match" +823,"Incorrect firmware loaded for this model number" Keysight Truevolt Series Operating and Service Guide...
  • Page 474 Offset Compensation (2- and 4- wire RTD) Low Power Measurements(2- and 4- wire RTD or thermistor) Off Thermocouple (34465A/70A only): Check Reference Junction Temperature, Type, Adjust 0° C, Internal, 0° C Thermocouple Type * For all DC measurement functions. Keysight Truevolt Series Operating and Service Guide...
  • Page 475 System-Related Operations Factory Setting • Beeper Mode • On • Thousands Separator • On Display State Reading Memory Cleared Error Queue See note • Stored States • No Change • Calibration State • Secured Keysight Truevolt Series Operating and Service Guide...
  • Page 476 2 LAN setting changes require a LAN restart. From SCPI you must cycle power. 3 Where nnnnn represents the last 5 digits of the instrument serial number. The instrument uses LAN port 5024 for SCPI Telnet sessions and port 5025 for SCPI Socket sessions. Keysight Truevolt Series Operating and Service Guide...
  • Page 477 Installing the Optional GPIB Interface Types of Service Available If your instrument fails during the warranty period, Keysight Technologies will repair or replace it under the terms of your warranty. After your warranty expires, Keysight offers repair services at competitive prices.
  • Page 478 Do not use detergent or solvents. Do not attempt to clean internally. If needed, contact a Keysight Technologies Sales and Service office to arrange for proper cleaning to ensure that safety features and performance are maintained. Electrostatic Discharge (ESD) Precautions Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling.
  • Page 479 The following measurements use the frame of the instrument as the low voltage reference. For the fol- lowing measurements, the instrument should be turned on, and the light below the power switch should be green. Location Detail C1103 pin 1 15 V ± Keysight Truevolt Series Operating and Service Guide...
  • Page 480 J1102 pin 21: 3.3 V ± 5% Low reference: large metal shield R1020 (either side of component): 10 V ± 20% Low reference: large metal shield E1006 (either side of component): 3.3 V ± 5% Keysight Truevolt Series Operating and Service Guide...
  • Page 481 1.2 V ± 5% Low reference: large metal shield U1004 pin 3: 5 V ± 5% Low reference: large metal shield U301: pin 2: +16.8 V ± 5% pin 3: -16.4 V ± 5% Keysight Truevolt Series Operating and Service Guide...
  • Page 482 Keysight Truevolt Series Operating and Service Guide...
  • Page 483 The voltage between the red and orange wires (below)should measure 7.5 to 11 VAC. The voltage between the brown and white wires (below) should measure 11 to 15 VAC. 5. If the voltages are correct, replace the main board; otherwise replace the transformer. Keysight Truevolt Series Operating and Service Guide...
  • Page 484 7. If the self test generates an error, record the error code and message and contact Keysight support if necessary.
  • Page 485 Service and Repair 10. Also verify that the voltage is 3.3 V ± 5% at resistor R121, referenced to the instrument chassis. If this voltage is incorrect, replace the front panel board. Keysight Truevolt Series Operating and Service Guide...
  • Page 486 Front Panel Execution 1. Press [Shift] > [Utility] > Test/Admin > Self Test. 2. View any failures by pressing [Shift] > [Help] > View remote command error queue. Keysight Truevolt Series Operating and Service Guide...
  • Page 487 Front Panel Shield 5041-5225 Front/Rear Push Rod (34461A, 34465A, 34470A) 5041-5232 Fan (34461A, 34465A, 34470A) 5041-5261 Fan Shield (34465A, 34470A) 5041-5257 Cable Assembly, Thermistor (34465A, 34470A) 1420-0356 Battery (in front panel) 34470-66303 Reference Assembly Keysight Truevolt Series Operating and Service Guide...
  • Page 488 Do not turn the screw that fastens the shield in place, as this will affect the instru- ment's calibration. Tools Required The following tools are required. T20 Torx driver (most disassembly) Small, flat-blade screw driver (for prying) Keysight Truevolt Series Operating and Service Guide...
  • Page 489 At this point, the instrument it sufficiently disassembled for power supply troubleshooting purposes. You could also change the battery or override calibration security at this point. If you need to disassemble the front panel, continue with the steps shown below. Keysight Truevolt Series Operating and Service Guide...
  • Page 490 9. Slide the white plastic backing plate to the right and lift up on the left side of the backing plate to remove it. 10. Remove the front panel PCBA by holding the connector housing shown below and lifting it out. Keysight Truevolt Series Operating and Service Guide...
  • Page 491 Service and Repair 11. Remove the ESD shield by prying it out gently from the plastic housing to which it attaches. This concludes the disassembly procedure. To re-assemble the instrument, reverse the procedure. Keysight Truevolt Series Operating and Service Guide...
  • Page 492 Always disconnect all inputs, cords, and cables before disassembling the instru- ment. Tools Required T20 Torx driver (most disassembly) Small, flat-blade screw driver (for prying) Keysight Truevolt Series Operating and Service Guide...
  • Page 493 6. Position the new battery in the battery holder, + side up. Push down gently to seat the battery in the holder. 7. Re-assemble the instrument completely before connecting any cables or cords. This concludes the battery replacement procedure. Keysight Truevolt Series Operating and Service Guide...
  • Page 494 Do not turn the screw that fastens the shield in place, as this will affect the instru- ment's calibration. Tools Required T20 Torx driver (most disassembly) Small, flat-blade screw driver (for prying) Keysight Truevolt Series Operating and Service Guide...
  • Page 495 DMM in continuity measurement mode, and short the HI Input terminal to the 10 A current ter- minal. If the measurement displays OPEN, replace the internal 10 A current path fuse as described below. Keysight Truevolt Series Operating and Service Guide...
  • Page 496 6. Position the new fuse in the fuse holder. Push down gently to seat the fuse in the holder. 7. Re-assemble the instrument completely before connecting any cables or cords. This concludes the fuse replacement procedure. Keysight Truevolt Series Operating and Service Guide...
  • Page 497 GPIB option. The instrument should never be connected to power or inputs on the measurement terminals without either the GPIB module or the cover plate securely covering the rear-panel opening. Locate the connector, just inside the unit, for the cable attached to the GPIB module Keysight Truevolt Series Operating and Service Guide...
  • Page 498 Put the module into the unit and slide the GPIB board to the right so that it is flush against the instru- ment's sheet metal. Use the screw that was removed above to secure the GPIB plate in place. This concludes the GPIB installation procedure. Keysight Truevolt Series Operating and Service Guide...
  • Page 499 If the firmware revision ends with 02 or greater, use Procedure A. If the firmware revision ends with 01, use Procedure If you encounter problems, contact Keysight Technologies for technical support. In the United States: (800) 829-4444 In Europe: 31 20 547 2111 In Japan: 0120-421-345 www.keysight.com/find/assist to contact Keysight worldwide, or contact your Keysight Technologies representative.
  • Page 500 Press [Shift] > [Utility] > Test / Admin > Security > New Code. b. Enter AT3446XA as the new passcode. c. Press Done. d. Verify the new code and press Yes to confirm the change. Keysight Truevolt Series Operating and Service Guide...
  • Page 501 Service and Repair 5. Remove power, disassemble the instrument, and remove the short. Then reassemble the instrument to complete the security code (calibration password) reset procedure. Keysight Truevolt Series Operating and Service Guide...
  • Page 502 3. Reassemble the instrument, apply power and turn it on; the instrument will generate an error and beep (this is normal). Wait 30 to 40 seconds for the instrument to power up completely. Keysight Truevolt Series Operating and Service Guide...
  • Page 503 Zero Offset Verification DC Volts and DC Current Gain Verification Frequency Accuracy Verification AC Voltage and AC Current Verification High Current Verification Capacitance Verification (Optional Verification Test) Keysight Truevolt Series Operating and Service Guide...
  • Page 504 Perform only the performance verification tests indicated with the letter Q. If the instrument fails the quick performance check, calibration or repair is required. The Quick Performance Check can be limited by the factors described in the following section (Per- formance Verification Tests). Keysight Truevolt Series Operating and Service Guide...
  • Page 505 An alternate method is to use the Keysight 3458A 8½-digit Digital Multimeter to measure less accurate, yet stable, sources. The output value measured from the source can be entered into the instrument as the target calibration value.
  • Page 506 ± 100 mΩ Short 4-wire Ohms 100 kΩ ± 500 mΩ ± 1 Ω 1 MΩ ± 10 Ω 10 MΩ ± 100 Ω [1] Select 10 PLC. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 507 ± 100 mΩ Short 4-wire Ohms 100 kΩ ± 500 mΩ ± 1 Ω 1 MΩ ± 10 Ω 10 MΩ ± 100 Ω [1] Select 10 PLC. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 508 10 MΩ ± 100 Ω Short 2-wire Ohms 100 MΩ 1 kΩ [1] Select 10 PLC. [2] These values are typical and these ranges are calibrated internally. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 509 10 MΩ ± 100 Ω Short 2-wire Ohms 100 MΩ 1 kΩ [1] Select 10 PLC. [2] These values are typical and these ranges are calibrated internally. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 510 ± 0.13 µA ± 0.36 µA ± 0.56 µA ± 0.66 µA 100 µA 100 µA ± 0.03 µA ± 0.065 µA ± 0.075 µA ± 0.085 µA [1] Select 10 PLC. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 511 ± 45 mV ± 55 mV ± 70 mV 1000 V -500 V ± 16 mV ± 27.5 mV ± 32.5 mV ± 40 mV [1] Select 10 PLC. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 512 ± 5 µA ± 7 µA ± 8 µA 1 mA 1 mA ± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA [1] Select 10 PLC. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 513 ± 3.1 kΩ 10 MΩ 10 MΩ 2-wire Ohms 100 MΩ 100 MΩ ± 101 kΩ ± 201 kΩ ± 301 kΩ ± 401 kΩ [1] Select 10 PLC. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 514 ± 3.1 kΩ 10 MΩ 10 MΩ 2-wire Ohms 100 MΩ 100 MΩ ± 101 kΩ ± 201 kΩ ± 301 kΩ ± 401 kΩ [1] Select 10 PLC. : Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 515 2. Select the Frequency function, default range, 3 Hz filter, and 1 second aperture. 3. Connect the Keysight 33500B Series Function/Arbitrary Waveform Generator to the input terminals. For the 34461A/65A/70A, use the front panel input terminals and select the front input terminals with the Front/Rear switch.
  • Page 516 ± 675 mV ± 825 mV ± 900 mV ± 975 mV 210 V 50 kHz 750 V 200 Hz ± 573 mV ± 669 mV ± 690 mV ± 711 mV 70 V 300 kHz ± 6.6 V Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 517 ± 450 mV ± 600 mV ± 675 mV ± 750 mV 210 V 50 kHz 750 V 200 Hz ± 510 mV ± 606 mV ± 627 mV ± 648 mV 70 V 300 kHz ± 6.6 V Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 518 14 µA 5 kHz 1 kHz 1 mA 1 mA 200 Hz 1.4 µA 5 kHz 1 kHz 100 µA 100 µA 200 Hz 0.14 µA 5 kHz Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 519 ± 1.1 V 210 V 50 kHz 750 V 200 Hz ± 330 mV ± 351 mV ± 372 mV ± 393 mV 70 V 300 kHz ± 1.5 V Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 520 ± 1.1 µA ± 1.3 µA ± 1.4 µA 5 kHz 1 kHz 100 µA 100 µA 200 Hz ± 0.110 µA ± 0.130 µA ± 0.140 µA 5 kHz Quick performance verification test point. Keysight Truevolt Series Operating and Service Guide...
  • Page 521 ± 24 mA (34465A/70A) 10 A 5 kHz 10 A ± 29 mA ± 29 mA (34461A) If you are NOT using the Fluke 5725 for calibration, reduce the current to 2 A. Keysight Truevolt Series Operating and Service Guide...
  • Page 522 5. Compare measurement results to the test limits shown in the table. (Be certain to allow time for source settling.) Input Error from Capacitance Range Nominal 1 μF 1 μF ± 0.005 μF 1 μF 10 μF ± 0.014 μF Keysight Truevolt Series Operating and Service Guide...
  • Page 523 Calibration Adjustment Procedures Calibration Adjustment Procedures You will need a test input cable and connectors set and a low thermal input short, Keysight 34172B, to cal- ibrate the instrument (see Input Connections). Two input shorts are recommended for the 34461A/65A/70A.
  • Page 524 The instrument should be calibrated on a regular interval determined by the accuracy requirements of your application. A 1-year interval is adequate for most applications. Accuracy specifications are war- ranted only if calibration is made at regular calibration intervals. Keysight Technologies never recom- mends calibration adjustment intervals beyond two years.
  • Page 525 SCPI programming. Input Connections Test connections to the instrument are best accomplished using a Keysight Technologies 34172B cal- ibration short for low–thermal offset measurements and a 34171B DMM connector configured to inter- face to the calibrator output. Shielded, twisted–pair, PTFE interconnect cables of minimum length are recommended between the calibrator and the multimeter to reduce settling and noise errors.
  • Page 526 An alternate method is to use the Keysight 3458A 8½-digit Digital Multimeter to measure less accurate, yet stable, sources. The output value measured from the source can be entered into the instrument as the target calibration value.
  • Page 527 Procedures). 5. Secure the instrument against calibration (see Calibration Security). 6. Note the security code and calibration count in the instrument’s maintenance records. 7. Perform the performance verification tests to verify the calibration. Keysight Truevolt Series Operating and Service Guide...
  • Page 528 100 mV to 10 V 0.9 to 1.1 x Full Scale AC Volts (rms) 100 V 40 V to 110 V 750 V 400 to 750 V Valid frequencies are nominal frequencies ±1%. Keysight Truevolt Series Operating and Service Guide...
  • Page 529 Most measurement functions and ranges have gain calibration procedures, and calibration for each func- tion should be performed in the order shown. Keysight Truevolt Series Operating and Service Guide...
  • Page 530 To store calibration values in non-volatile memory, use CALibration:STORe. Front Panel: [Utility] > Test/Admin > Calibrate > Save Cal Values You must have performed a successful calibration immediately before pressing Save Cal Values. Keysight Truevolt Series Operating and Service Guide...
  • Page 531 SYSTem:PRESet. Security code rules: Unquoted string up to 12 characters Must start with letter (A-Z) May contain letters, numbers (0-9) and underscores Front Panel: [Utility] > Test / Admin > Security SCPI: CALibration:SECure:STATe Keysight Truevolt Series Operating and Service Guide...
  • Page 532 To read the calibration message, press Utility > Test/Admin > Calibrate or send CALibration:STRing? from the remote interface. To change the calibration message, unsecure the instrument and press Utility > Test/Admin > Calibrate > Edit Cal String or send CALibration:STRing "<string>" from the remote interface. Keysight Truevolt Series Operating and Service Guide...
  • Page 533 To read the calibration count, press [Utility] > Test/Admin > Calibrate on the front panel or send CALibra- tion:COUNt? from the remote interface. You can read the calibration count regardless of whether the instrument is secured. Keysight Truevolt Series Operating and Service Guide...
  • Page 534 705 Cal:Aborted. It may also generate errors 740 through 742. If this occurs, you should not use the instrument until a complete re–calibration has been performed. See the list of calibration errors for more information. Keysight Truevolt Series Operating and Service Guide...
  • Page 535 If the firmware revision ends with 02 or greater, use Procedure A. If the firmware revision ends with 01, use Procedure If you encounter problems, contact Keysight Technologies for technical support. In the United States: (800) 829-4444 In Europe: 31 20 547 2111 In Japan: 0120-421-345 www.keysight.com/find/assist to contact Keysight worldwide, or contact your Keysight Technologies representative.
  • Page 536 Press [Shift] > [Utility] > Test / Admin > Security > New Code. b. Enter AT3446XA as the new passcode. c. Press Done. d. Verify the new code and press Yes to confirm the change. Keysight Truevolt Series Operating and Service Guide...
  • Page 537 Calibration Adjustment Procedures 5. Remove power, disassemble the instrument, and remove the short. Then reassemble the instrument to complete the security code (calibration password) reset procedure. Keysight Truevolt Series Operating and Service Guide...
  • Page 538 3. Reassemble the instrument, apply power and turn it on; the instrument will generate an error and beep (this is normal). Wait 30 to 40 seconds for the instrument to power up completely. Keysight Truevolt Series Operating and Service Guide...
  • Page 539 Verify the new code and press Yes to confirm the change. 5. Remove power, disassemble the instrument, and remove the short. Then reassemble the instrument to complete the security code (calibration password) reset procedure. Keysight Truevolt Series Operating and Service Guide...
  • Page 540 Frequency Accuracy Calibration DC Voltage Gain Calibration Ohms Gain Calibration DC Current Gain Calibration AC Current 10 A Gain Calibration DC Current 10 A Gain Calibration Capacitance Offset Calibration (Optional) Finishing Calibration First Procedure  Keysight Truevolt Series Operating and Service Guide...
  • Page 541 The following gain and flatness calibration procedures can be done using either the front or rear terminals (rear terminals not available on the 34460A). On the 34461A/65A/70A, make sure the Front/Rear switch matches the terminals in use. Keysight Truevolt Series Operating and Service Guide...
  • Page 542 Calibration Adjustment Procedures Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 543 5. Store the new calibration constants (see Storing the Calibration Constants). Each voltage and frequency takes approximately 2.5 seconds to complete. Input Instrument Vrms Frequency Range 10 Hz 10 V 40 Hz 10 V Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 544 5 kHz, 10 kHz, 20 kHz, 35 kHz, 50 kHz, 75 kHz, 100 kHz, 50 V 100 V 200 kHz, 300 kHz, 390 kHz, 400 kHz, 220 Hz, 1000 Hz 500 V 1000 Hz 750 V Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 545 100 mA 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz 100 mA 10 mA 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz 10 mA 1 mA 1 kHz 1 mA 100 µA 1 kHz 100 µA Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 546 Calibration Step Failed, check the input value, range, function, and entered calibration value and repeat the calibration step. 3. Store the new calibration constants (see Storing the Calibration Constants). This calibration takes approximately 10 seconds to complete. Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 547 Configuration: Frequency, 10 V range The frequency accuracy of the Fluke 5720A is insufficient to calibrate the DMM. Its frequency output needs to be calibrated against a more accurate reference. The Keysight 33500B Series Function/Arbitrary Waveform Generator is recommended for this calibration procedure.
  • Page 548 Instrument Settings Input Function Range 10 V 10 V -10 V 10 V 0.1 V 100 mV -0.1 V DC Voltage 100 mV 100 V 100 V 500 V 1000 V Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 549 100 Ω 100 Ω 1 kΩ 1 kΩ 10 kΩ 10 kΩ 4-wire Ohms 100 kΩ 100 kΩ 1 MΩ 1 MΩ 10 MΩ 10 MΩ 100 MΩ 2-wire Ohms 100 MΩ Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 550 Each range takes less than 4 seconds to complete. Instrument Settings Input Function Range 100 µA 100 µA 1 mA 1 mA 10 mA DC Current 10 mA 100 mA 100 mA Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 551 4. Store the new calibration constants (see Storing the Calibration Constants). This calibration takes approximately 2.5 seconds to complete. Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 552 4. Store the new calibration constants (see Storing the Calibration Constants). This calibration takes less than 4 seconds to complete. Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 553 4. Repeat steps 1 through 3 for the instrument's rear terminals (not applicable to the 34460A). 5. Store the new calibration constants (see Storing the Calibration Constants). Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 554 1. Remove all shorting blocks and connections from the instrument. 2. Reset the calibration message. Secure the instrument with the calibration password. 4. Record the new calibration count. 5. Perform the performance verification tests to verify the calibration. Keysight Truevolt Series Operating and Service Guide...
  • Page 555 DC Voltage Gain Calibration Ohms Gain Calibration DC Current Gain Calibration AC Current 10 A Gain Calibration DC Current 10 A Gain Calibration DC High Voltage Gain Calibration Capacitance Offset Calibration (Optional) Finishing Calibration First Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 556 The following gain and flatness calibration procedures can be done using either the front or rear terminals (rear terminals not available on the 34460A). On the 34461A/65A/70A, make sure the Front/Rear switch matches the terminals in use. Keysight Truevolt Series Operating and Service Guide...
  • Page 557 Calibration Adjustment Procedures Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 558 5. Store the new calibration constants (see Storing the Calibration Constants). Each voltage and frequency takes approximately 2.5 seconds to complete. Input Instrument Vrms Frequency Range 10 Hz 10 V 40 Hz 10 V Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 559 5 kHz, 10 kHz, 20 kHz, 35 kHz, 50 kHz, 75 kHz, 100 kHz, 50 V 100 V 200 kHz, 300 kHz, 390 kHz, 400 kHz, 220 Hz, 1000 Hz 500 V 1000 Hz 750 V Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 560 1 mA 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz 1 mA 100 µA 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz 100 µA Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 561 Calibration Step Failed, check the input value, range, function, and entered calibration value and repeat the calibration step. 3. Store the new calibration constants (see Storing the Calibration Constants). This calibration takes approximately 10 seconds to complete. Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 562 Configuration: Frequency, 10 V range The frequency accuracy of the Fluke 5720A is insufficient to calibrate the DMM. Its frequency output needs to be calibrated against a more accurate reference. The Keysight 33500B Series Function/Arbitrary Waveform Generator is recommended for this calibration procedure.
  • Page 563 10 V -10 V 10 V 0.1 V 100 mV -0.1 V 100 mV The 100V and 1000V ranges are adjusted later in the DC Voltage High Voltage Gain Cal- ibration. Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 564 100 Ω 100 Ω 1 kΩ 1 kΩ 10 kΩ 10 kΩ 4-wire Ohms 100 kΩ 100 kΩ 1 MΩ 1 MΩ 10 MΩ 10 MΩ 100 MΩ 2-wire Ohms 100 MΩ Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 565 Storing the Calibration Constants). Each range takes less than 4 seconds to complete. Instrument Settings Input Function Range 1 mA 1 mA 10 mA 10 mA 100 mA DC Current 100 mA Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 566 5. Store the new calibration constants (see Storing the Calibration Constants). Each range takes less than 4 seconds to complete. Instrument Settings Input Function Range 100 V 100 V DC Voltage 500 V 1000 V Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 567 4. Store the new calibration constants (see Storing the Calibration Constants). This calibration takes approximately 2.5 seconds to complete. Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 568 4. Store the new calibration constants (see Storing the Calibration Constants). This calibration takes less than 4 seconds to complete. Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 569 4. Repeat steps 1 through 3 for the instrument's rear terminals (not applicable to the 34460A). 5. Store the new calibration constants (see Storing the Calibration Constants). Next Procedure Keysight Truevolt Series Operating and Service Guide...
  • Page 570 1. Remove all shorting blocks and connections from the instrument. 2. Reset the calibration message. Secure the instrument with the calibration password. 4. Record the new calibration count. 5. Perform the performance verification tests to verify the calibration. Keysight Truevolt Series Operating and Service Guide...
  • Page 571 2-wire measurements 346-347, 352 ACALibration 393-394 2-wire Resistance 250, 313, 353-355 ACI 64 2-Wire Resistance 66 acquiring readings 95 2-wire RTD 251 ACV 60 2-Wire vs. 4-Wire Measurements 172 ADC 237, 542, 557 ADDRess 407 Keysight Truevolt Series Operating and Service Guide...
  • Page 572 Average, moving 234-235 calibration count 534 Calibration Errors 461 CALibration Subsystem 237 BANDwidth 320, 373 Capacitance 76 Bar Meter 121 CAPacitance 245, 282, 314-318 battery 493 Capacitance Measurement Con- BEEPer 394-395 siderations 176 BMP 263 Keysight Truevolt Series Operating and Service Guide...
  • Page 573 COMMunicate 406-415, 417 Data Logging Local/Remote Interaction 180, 186 CONDition 389-390 Data Logging, Trend Chart 127 CONFigure 245-252, 254 DATA Subsystem 256 Configure I/O 144 DATA2 334 CONFigure Subsystem 243 DATE 396 CONFigure? 245 Keysight Truevolt Series Operating and Service Guide...
  • Page 574 DESired 278 Device-Specific Errors 461 ENABle 277, 389, 391, 406, 418 Device Clear 199, 408 End-Or-Identify 197 DHCP 408 ERRor 397, 424 Digitize Mode 47, 86 Error Messages 461 Digitizing 86 Error Queue 461 Keysight Truevolt Series Operating and Service Guide...
  • Page 575 GPIB 194, 407 flatness 546, 561 GPIB option 498 flatness adjustments 530 Ground Loops, Noise 164 flatness calibration 546, 561 FNAMe 303 Handle 40 FORMat 261, 263, 303-304 HCOPy 263 FORMat Subsystem 261 Keysight Truevolt Series Operating and Service Guide...
  • Page 576 Keysight IO Libraries Suite 194 I/O Configuration 144 LAN 423, 425 LABel 400 IDENtify 277, 399 LAN 408-415, 417 IDN? 269 configuring 278 IEEE 488.2 Commands 197, 271 LAN configuration 27 IEEE 488.2 Common Commands 264 Keysight Truevolt Series Operating and Service Guide...
  • Page 577 LOWer 215, 220, 340 Measurement Considerations, Capa- LPOWer 353, 366 citance 176 LRN 271 Measurement limits 18 LXI 277-279 Measurement Tutorial 159 LXI Subsystem 277 Measurements 56 Measurements Below Full Scale 190 MEDium 234 Keysight Truevolt Series Operating and Service Guide...
  • Page 578 NPLC 327, 329, 346, 348, 359-360, 379, Period 79 PERiod 286, 313, 337-344 Null 102 Period and Frequency Measurement NULL 314-316, 321-323, 338-339, 349- Errors 171 351, 361-362, 374-376 PLC 327, 346, 359, 379, 382, 453 NULL Reading 173 Keysight Truevolt Series Operating and Service Guide...
  • Page 579 QUEStionable 390-391 repair 478, 528, 535, 544-545, 548-550, Questionable Data Register 273 552-553, 559-560, 563-565, 567- Quick Reference 435 Replaceable Parts 488 replacing fuses 495 R? 205 REQuest 421 rack 44 RESet 279 Keysight Truevolt Series Operating and Service Guide...
  • Page 580 SENSe 313-318, 320-333, 337-344, 346- 357, 359-368, 370-371, 373-385 Safety symbols 13 {FREQuency|PERiod} Subsystem 336 SAMPle 306-308, 310 {RESistance|FRESistance} Sample Rate and Interval 193 Subsystem 345 SAMPle Subsystem Introduction 306 CURRent Subsystem 319 SAV 273 Keysight Truevolt Series Operating and Service Guide...
  • Page 581 359-368, 370-371, 402 State Store and Recall 140 Temperature (34460 and 34461A) 69 Statistical functions Temperature (34465A and 34470A) 71 maximum value 196 Temperature Measurements 172 minimum value 196 Temperature Probe Type Choice 172 Keysight Truevolt Series Operating and Service Guide...
  • Page 582 Trigger 306, 308, 430 351, 361-362, 375-376 TRIGger 204, 428-430, 432-433 verification 511, 516-517, 522 Trigger delay 97 Verification Tests 504 Trigger Delay 429 VERSion 403 trigger delays 457 VIEW 260 TRIGger Subsystem 427 Keysight Truevolt Series Operating and Service Guide...
  • Page 583 517 Voltmeter Complete 460 WAI 276 Wait-for-trigger 275, 308-309, 311 Web Interface 152, 194 Web UI 152 WINS 417 WMESsage 404, 415 ZERO 332, 357, 371, 385 zero offset verification 507 Keysight Truevolt Series Operating and Service Guide...
  • Page 584 Keysight Technologies, Inc. 2013 - 2022 All © rights reserved. Part Number: 34460-90901 Printed in Malaysia: Edition 09, October 2022...
  • Page 585 Multimè tres numé riques Sé rie True volt Ce manuel contient des informations pour l'utilisation, la programmation et la maintenance des multimè tres numé riques (DMM) Keysight de la sé rie True volt. GUIDE D'UTILISATION ET DE MAINTENANCE...
  • Page 586 Avertissement : Ce document contient des références à Agilent Technologies. L’ancienne activité de test et de mesure d’Agilent est devenue Keysight Technologies. Pour de plus amples informations, rendez-vous sur www.keysight.com.
  • Page 587 Présentation de l’instrument Présentation générale de l’instrument Affichage – Affichez, enregistrez et documentez aisément vos résultats de mesure Mesures – Performances de mesure Truevolt de Keysight alliées à un accès de pointe aux E/S Langage de programmation Bref aperçu du panneau avant Touches du panneau avant Bref aperçu du panneau arrière...
  • Page 588 Math (Opération mathématique) - Introduction Math (Opération mathématique) - Null (Valeur de référence) Math (Opération mathématique) - dB/dBm Scaling (Réglage d’échelle dB/dBm) Math (Opération mathématique) - Scaling (Réglage d’échelle) Math (Opération mathématique) - Smoothing (Lissage) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 589 Réalisation de mesures en courant alternatif haut débit Réalisation de mesures en courant continu et de résistance haut débit Capacité Considérations relatives aux mesures de capacité Numérisation de mesures La fréquence d’échantillonnage Déclenchement par niveau Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 590 FETCh? INITiate[:IMMediate] OUTPut:TRIGger:SLOPe {POSitive|NEGative}OUTPut:TRIGger:SLOPe? R? [<lectures_max>] READ? ROUTe:TERMinals? TEST:ALL? UNIT:TEMPerature {C|F|K}UNIT:TEMPerature? Sous-système CALCulate - Présentation Sous-système CALibration Sous-système CONFigure Sous-système DATA Sous-système DISPlay Sous-système FORMat Sous-système HCOPy Commandes courantes IEEE 488.2 Sous-système LXI Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 591 Erreurs propres au dispositif (-300…) Erreurs de requête (-400…) Erreurs réseau (+100...) Erreurs de l’instrument (+200…) Erreurs diverses (+300... et +500..) Erreurs de licence et d’autotest (+600...) Erreurs d’étalonnage (+700…) Erreurs diverses (+800...) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 592 Connexions d’entrée Considérations relatives aux tests Équipement de test recommandé Procédure de réglage de l’étalonnage Présentation du réglage de l'étalonnage de gain Présentation du réglage de l’étalonnage de gain et de la planéité Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 593 Étalonnage du gain en courant continu Étalonnage du gain en tension continue Étalonnage du gain 10 A en courant alternatif Étalonnage du gain 10 A en courant continu Étalonnage du décalage de capacité (en option) Finalisation de l’étalonnage Index Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 594 à la fourniture, à l’utilisation ou à l’exactitude de ce docu- ment ou aux performances de tout produit Keysight auquel il se rapporte. Si Keysight a passé un contrat écrit avec l’utilisateur et que certains termes de ce contrat semblent contradictoires avec ceux de ce document, les termes et conditions de la garantie prévalent.
  • Page 595 Limitation des droits du Gouvernement des États-Unis. Les droits s’appliquant au logiciel et aux infor- mations techniques concédées au gouvernement fédéral incluent uniquement les droits concédés habi- tuellement aux clients utilisateurs. Keysight concède la licence commerciale habituelle sur le logiciel et les informations techniques suivant les directives FAR 12.211 (Informations techniques) et 12.212 (Logi- ciel informatique) et, pour le Ministère de la Défense, selon les directives DFARS 252.227-7015 (Infor-...
  • Page 596 Ce texte indique que l’instrument est un produit industriel, scientifique et médical de classe A et du groupe 1 (CISPER 11, clause 4). ICES/NMB- This ISM device complies with Canadian ICES-001. Cet instrument ISM est conforme à la norme NMB-001 du Canada. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 597 Pour mettre l’instrument hors tension, il convient de débrancher le cordon d’alimentation secteur. Assu- rez-vous de laisser un espace suffisant autour du cordon d’alimentation pour faciliter son débranchement du secteur. N’utilisez que le cordon d’alimentation indiqué par Keysight pour une utilisation dans le pays de destination ou un cordon présentant des caractéristiques équivalentes.
  • Page 598 à l’alimentation secteur pour les tensions secteur supérieures à 300 Vca. Pour de plus amples informations, reportez-vous à la section Instrument de mesure de classe de classe II conforme à la norme CEI. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 599 Conformité et réglementation des produits Les multimètres numériques de la série Truevolt sont conformes aux normes de sécurité et de CEM. Reportez-vous à la Déclaration de conformité http://www.keysight.com/go/conformity pour connaître la dernière révision. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 600 Pour garantir une utilisation sans danger de l’instrument, ne dépassez pas les limites de mesures indiquées sur le panneau avant et arrière, et définies comme suit : Zone de bornes du panneau avant 34460A Zone de bornes du panneau avant  34461A, 34465A, 34470A Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 601 L’unité comporte des fusibles de protection de courant de 3 A et 10 A (modèles 34461A, 34465A et 34470A). Pour obtenir des instructions de remplacement, contactez votre revendeur et le service après-vente Keysight ou consultez la documentation d’entretien du produit. Pour une protection optimale, remplacez les fusibles par des fusibles du type et du calibre spécifiés.
  • Page 602 Appareil de mesure de classe II conforme à la norme CEI Pour prévenir tout risque d’électrocution, le multimètre numérique Keysight série Truevolt protège l’utilisateur contre les événements de surtension de l’alimentation secteur. Lors de la mesure de l’alimentation secteur, les bornes d’entrée HI et LO peuvent être connectées à...
  • Page 603 Keysight 34139A. Si le jeu de cordons de test est utilisé d’une manière non recommandée par Keysight Technologies, la pro- tection assurée par ces cordons peut être inhibée. Aussi, n’utilisez pas de jeux de cordons de test endom- magés ou usés, sous peine de provoquer des blessures ou la mort.
  • Page 604 Bienvenue Ce Guide d’utilisation et de maintenance contient des informations sur l’utilisation, la programmation et l’entretien des multimètres numériques Keysight série Truevolt. Pour nous faire part de vos commentaires sur ce document, rendez-vous sur www.keysight.com/find/truevolt-docfeedback. Dans ce document, le terme « esclave » a été remplacé par « secondaire ».
  • Page 605 Installation de l’interface GPIB en option Pièces remplaçables par l’utilisateur Vérification du fonctionnement Vérification du fonctionnement Procédures d’étalonnage Procédures d’étalonnage Procédures d’étalonnage des modèles 34460A et 34461A Procédures d’étalonnage des modèles 34465A et 34470A Mise à jour du microprogramme Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 606 Interface USB, LAN (en option sur le modèle 34460A) et GPIB en option Connectivité USB sans pilote de type Glisser-déposer Mesures – Performances de mesure Truevolt de Keysight alliées à un accès de pointe aux Performances de niveau métrologique brevetées qui constituent la base de toutes les mesures...
  • Page 607 [Shift] et en la relâchant avant d’appuyer dessus. Par exemple, si vous appuyez sur la touche [Shift] et que vous la relâchez avant d’appuyer sur [Display], vous accéderez à la fonction [Utility] : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 608 Sortie complète de mesure du voltmètre Entrée de déclenchement externe Connecteur de réseau local (LAN) Connecteur d’interface USB Verrouillage du câble de l’instrument Borne à vis de mise à la terre du châssis Entrée secteur Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 609 Présentation de l’instrument Modèles et options Les multimètres numériques Keysight Truevolt suivants sont disponibles : 34460A - Multimètre numérique de base 6½ chiffres 34461A - Multimètre numérique de rechange 6½ chiffres 34401A 34465A - Multimètre numérique 6½ chiffres 34470A - Multimètre numérique 7½ chiffres Les options installées en usine et celles pouvant être installées par vous-même ou un distributeur sont répertoriées dans les tableaux ci-dessous.
  • Page 610 Présentation de l’instrument Référence 34460A 34461A 34465A 34470A Description produit 3446MEMU *S/O *S/O Option Option Mémoire de 2 Mo pour la licence des mul- timètres numériques 34465A et 34470A série Truevolt. *S/O = Sans objet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 611 L’instrument est livré avec le CD-ROM Keysight Automation-Ready. Ce CD-ROM contient le logiciel Keysight IO Libraries Suite qui doit être installé pour les opérations sur l’interface de commande à dis- tance. Il démarre automatiquement et contient des informations sur l’installation du logiciel. Il contient également le manuel Keysight Technologies USB/LAN/GPIB Connectivity Guide qui fournit...
  • Page 612 Pour configurer manuellement une adresse IP, un masque de sous-réseau ou une passerelle par défaut, désactivez le protocole DHCP, puis procédez comme suit pour modifier l’adresse IP. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 613 Si vous modifiez ce paramètre, vous devez appuyer sur la touche de fonction Apply Changes (pan- neau avant) ou envoyer la commande SYSTem:COMMunicate:LAN:UPDate (interface de commande à distance) pour que la modification soit prise en compte. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 614 Nom de domaine Un nom de domaine est un nom Internet enregistré qui est traduit en adresse IP. Vous pouvez le confi- gurer sur le panneau avant ou avec une commande SCPI. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 615 L’instrument intègre une interface Web pour l’accès et le contrôle à distance de l’instrument sur un réseau local via un navigateur Web. Pour de plus amples informations, reportez-vous à la section Inter- face Web. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 616 étant la valeur décimale d’un octet compris entre 0 et 255. Vous pouvez saisir une nouvelle adresse IP à l’aide des touches fléchées du panneau avant. Ne saisissez pas de zéro au début des nombres. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 617 à la notation décimale « 192.168.16.9 » car « .020 » est interprété comme « 16 » en base 8 et « .011 » comme « 9 ». Pour éviter toute confusion, utilisez uniquement des valeurs décimales comprises entre 0 et 255 sans zéro d’en-tête. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 618 *IDN? afin qu’elle corresponde au numéro de modèle réel, mettez le micro- programme à jour, puis réutilisez la commande SYSTem:IDENtify pour définir la réponse *IDN? sur l’autre numéro de modèle. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 619 Présentation de l’instrument Contacter Keysight Technologies Vous pouvez contacter Keysight Technologies pour toute information sur la garantie, les services ou l’assistance technique. Aux États-Unis : (800) 829-4444 En Europe : 31 20 547 2111 Au Japon : 0120-421-345 Pour contacter Keysight dans le monde entier ou votre représentant Keysight Technologies, rendez-vous www.keysight.com/find/assist...
  • Page 620 Keysight ou votre revendeur autorisé Keysight. Cordon d’alimentation (adapté au pays de destination) Certificat d’étalonnage (en option) CD Keysight Automation-Ready (en anglais) (contient le logiciel Keysight IO Libraries Suite) (en option sur le modèle 34460A) Dossier de documentation supplémentaire Câble USB 2.0 (en option sur le modèle 34460A)
  • Page 621 250 V, 0,25 A à action différée de 5x20 mm. Le numéro de référence Keysight est 2110-0817. Plage de tension d’alimentation Sélecteur de tension secteur nominale d’alimentation secteur 100 - 115 120 - 127 202 - 230 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 622 (terre de sécurité) au niveau de la prise de courant. Toute interruption du conducteur de pro- tection (mise à la terre) ou tout débranchement de la borne de terre de protection vous expose à un risque d’électrocution susceptible de provoquer des blessures graves ou la mort. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 623 500 ms. Cela évite d’éteindre accidentellement l’instrument en appuyant brièvement sur l’interrupteur. Si vous éteignez l’instrument en débranchant la prise d’alimentation (procédure déconseillée), il s’allume dès l’alimentation rétablie. Il n’est pas nécessaire d’appuyer sur l’interrupteur. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 624 Réglage de la poignée de transport La poignée comporte trois positions, illustrées ci-dessous. Pour régler la position de la poignée, saisissez les bords de la poignée, tirez vers l’extérieur et faites tour- ner la poignée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 625 Si le message contient plus d’informations que ne peut en afficher l’écran, appuyez sur la touche de fonc- tion fléchée vers le bas pour le parcourir. Appuyez sur Done pour quitter l’aide. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 626 Appuyez ensuite sur Select. Vous pouvez également appuyer sur la touche Demos pour exécuter des démonstrations interactives sur l’utilisation de l’instrument. Dans ce cas, la rubrique d’aide suivante apparaît : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 627 [Utility] > System Setup > User Settings > Help Lang. Sélectionnez ensuite la langue désirée. Les libellés des touches de fonction des menus et les messages des lignes d’état ne sont pas tra- duits. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 628 Vous pouvez installer l’instrument dans une armoire standard 19 pouces à l’aide d’un des deux kits dis- ponibles en option. Chacun de ces kits contient des instructions et le matériel nécessaire au montage. Vous pouvez également monter un autre instrument Keysight System II de hauteur et largeur identiques à côté de l’instrument.
  • Page 629 Assurez-vous de pouvoir utiliser les rails supports internes à l’armoire de baie. Étagère coulissante Pour installer un ou deux instruments sur une étagère coulissante, commandez l’étagère 5063-9255 et le kit de glissières 1494-0015. Pour un seul instrument, commandez également le panneau aveugle 5002- 3999. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 630 Menu Utility (Utilitaire) Interface Web Dans ce document, les états et les valeurs « par défaut » sont identifiés. Il s’agit des états par défaut à la mise sous tension lorsque l’instrument quitte l’usine. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 631 (lectures). La figure ci-dessous illustre le résultat de numérisation d’une onde sinusoïdale. Ce chapitre décrit les diverses méthodes de numérisation de signaux, l’importance de la fréquence d’échantillonnage et comment utiliser le déclenchement par niveau. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 632 Le délai de déclenchement est réglé sur Automatique. Le nombre de pré-déclenchements est défini à zéro. Le nombre d’échantillons par déclenchement est défini en fonction de la durée d’enregistrement des données (temps ou échantillons). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 633 La source d’échantillonnage est réglée sur Immédiate. Le nombre de pré-déclenchements est défini à 0. Le nombre d’échantillons par déclenchement est défini à 1. Le nombre de déclenchements est réglé sur Infini. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 634 Réglage automatique du zéro : Désactivé ou Activé (valeur par défaut) Configurer les mesures en tension alternative : Plage : Commutation automatique de plage (valeur par défaut), 100 mV, 1 V, 10 V, 100 V ou 750 V Filtre : >3 Hz, >20 Hz, >200 Hz Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 635 Réglage automatique du zéro : Désactivé ou Activé (valeur par défaut) OffstComp : OFF (valeur par défaut) ou ON. Concerne uniquement les modèles 34465A et 34470A. Low Power : Désactive (Off) ou active (On) les mesures de basse puissance. Concerne uniquement les modèles 34465A et 34470A. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 636 Plage : 1 nF, 10 nF, 100 nF, 1 µF, 10 µF, 100 µF or Auto (valeur par défaut) Configurer les mesures de continuité : Avertisseur sonore : Désactivé ou Activé (valeur par défaut) Configurer les mesures de diode : Avertisseur sonore : Désactivé ou Activé (valeur par défaut) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 637 (Mesures en 2 fils seulement ; non disponible pour les mesures en 4 fils) OffstComp : OFF (valeur par défaut) ou ON. (Mesures de RTD en 2 et 4 fils seule- ment) Vérification d’ouverture : Mesures TCouple uniquement.* Unités : °C, °F ou K Lancer et arrêter des mesures. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 638 Effectuer un étalonnage automatique (34465A/70A uniquement). Sélectionner la commutation de plage manuelle ou automatique. Appuyez sur pour monter manuellement d’une plage ou sur pour des- cendre manuellement d’une plage. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 639 Retourner au mode de commande local de l’instrument (en mode de commande à distance) ou indiquer que la touche suivante du panneau avant sera « basculée », par exemple [Probe Hold] à la place de [Single]. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 640 (modèles 34460A/61A uniquement) Configurer dB, dBm Scaling (modèles 34465A/70A uniquement) Configurer le réglage d’échelle : dB, dBm, %, Mx B Statistics Activer, désactiver et supprimer des statistiques. Limits Activer ou désactiver les limites supérieure et inférieure. Touche [Display] (Affichage) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 641 Effectuer des tâches d’autotest, d’étalonnage, de gestion de la sécurité et de la licence, et de mise à jour du microprogramme. System Définir les préférences de l’utilisateur, la date et l’heure et le message qui Setup s’affiche à la mise sous tension. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 642 Caractéristiques et fonctions Mesures Les multimètres numériques Keysight Truevolt prennent en charge de nombreuses mesures courantes : Tension continue Tension alternative Courant continu Courant alternatif Résistance Température Capacité Continuité Diode Fréquence et période Enregistrement de données Numérisation Déclenchement par niveau Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 643 (PLC) souhaités pour la mesure. Seuls 1, 10 et 100 PLC permettent une réjection (du bruit de fréquence réseau) en mode normal. La sélection de 100 PLC permet une réjection de bruit et une résolution optimales, mais ralentit les mesures. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 644 Dans la plupart des cas, la valeur 10 MΩ est suffisamment élevée pour ne pas charger la plupart des cir- cuits. Cependant, elle n’est pas suffisamment basse pour stabiliser les lectures des circuits à impédance Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 645 Configurez les mesures du rapport de tension continue comme indiqué ci-dessous : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 646 Si la polarisation en courant continu demeure constante, les mesures suivantes peuvent être réalisées avec une précision absolue sans retards de stabilisation sup- plémentaires. Étape 1 : Configurez les cordons de test comme indiqué. Étape 2 : Appuyez sur la touche [ACV] du panneau avant. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 647 Pour un affichage précis des statistiques des mesures en courant alternatif en mode Panneau avant, vous devez utiliser le retard de déclenchement manuel par défaut ([Acquire] > Delay Man). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 648 Étape 3 : Pour les modèles 34465A/70A, la touche de fonction Aperture NPLC est sélectionnée par défaut. Utilisez les touches fléchées vers le haut/bas pour spécifier le temps d’intégration en nombre de cycles de tension d’alimentation (PLC) souhaités pour la mesure. 1, 10 et 100 PLC permettent une Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 649 Il recueille une nouvelle mesure du décalage chaque fois que vous modifiez la fonc- tion, la plage ou le temps d’intégration. (Il n’existe aucun réglage du zéro automatique pour les mesures de résistance en 4 fils.) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 650 à 10 A, la plage de mesure devient automatiquement 10 A. Lors de la réalisation de mesures à l’aide de bornes 10 A, la présence d’un signal sur les bornes 3A peut provoquer de graves erreurs. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 651 Pour un affichage précis des statistiques des mesures en courant alternatif en mode Panneau avant, vous devez utiliser le retard de déclenchement manuel par défaut ([Acquire] > Delay Man). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 652 (PLC) souhaités pour la mesure. 1, 10 et 100 PLC permettent une réjec- tion (du bruit de fréquence réseau) en mode normal. La sélection de 100 PLC permet une réjection de bruit et une résolution optimales, mais ralentit les mesures. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 653 Range entraîne l’affichage du courant inférieur généré pour chaque plage : Les mesures de résistance de basse puissance s’appliquent uniquement aux plages 100 Ω-100 kΩ. Les plages 1 MΩ-1 GΩ génèrent le même courant, quel que soit le réglage de basse puissance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 654 Dans les mêmes conditions, l’instrument 34401A renvoie la valeur absolue de la mesure pour éviter toute confusion associée aux lectures négatives. Les multimètres numériques Keysight série Truevolt renvoient des valeurs négatives. Ce comportement permet d’obtenir des résultats extrêmement précis après une opération NULL.
  • Page 655 Cette section décrit la procédure à suivre pour configurer les mesures de température en 2 ou 4 fils à par- tir du panneau avant. Étape 1 : Configurez les cordons de test comme indiqué. Température 2 fils : Température 4 fils : Étape 2 : Appuyez sur la touche [Temp] du panneau avant. Le menu suivant s’affiche. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 656 La sélection de 100 PLC permet une réjection de bruit et une résolution optimales, mais ralentit les mesures : Étape 6 : Utilisez la touche de fonction Units pour afficher la température en degrés Celsius, degrés Fah- renheit ou Kelvin. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 657 Cette section contient des informations générales sur la configuration des mesures de tem- pérature. Pour un didacticiel détaillé sur les mesures de température, consultez la note d’application 290 de Keysight intitulée Mesures de température pratiques, disponible à l’adresse www.keysight.com. Étape 1 : Configurez les cordons de test comme indiqué.
  • Page 658 élimine les effets de faibles tensions continues dans le circuit mesuré. La méthode consiste à recueillir la différence entre deux mesures de résistance, une avec la source de courant réglée sur la Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 659 Cette méthode évite que des tensions de décalage pré- sentes sur le circuit d’entrée de l’instrument n’affectent la précision de mesure. Lorsque cette fonction est Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 660 à 20 °C pour une température mesurée de 100 °C. Pour un didacticiel détaillé sur les mesures de température, consultez la note d’application 290 de Key- sight intitulée Mesures de température pratiques, disponible à l’adresse www.keysight.com. Pour les sondes de type TCouple...
  • Page 661 Caractéristiques et fonctions +20,68 °C et le décalage spécifié est de +5 °C, les fonctions mathématiques de jonction froide artificielle utilisent la somme des deux valeurs. Keysight recommande de laisser le décalage défini à la valeur 0, sauf indication contraire de la métrologie.
  • Page 662 élevée pour l’algorithme à mesurer. Si vous appliquez une tension continue ou un court-circuit aux bornes d’entrée en mode de mesure de capacité, l’instrument rapporte une surcharge. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 663 ≤ 10 Ω affiche la résistance mesurée et émet un signal sonore (si l’avertisseur sonore est activé) 10 Ω à 1,2 kΩ affiche la résistance mesurée sans émettre de signal sonore > 1,2 kΩ affiche l’état OPEN sans aucun signal sonore Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 664 émet un signal sonore lorsque le signal et 5 V passe au seuil compris entre 0,3 et 0,8 V (si l’avertisseur sonore est activé) >5 V le panneau avant affiche le message « OPEN » et SCPI renvoie 9.9E37 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 665 Cette fonction est pratique, mais génère des mesures plus lentes par rapport à la fonction de com- mutation manuelle. Elle augmente d’une plage à 120 % de la plage actuelle et descend d’une plage sous 10 % de la plage actuelle. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 666 Pour un affichage précis des statistiques des mesures en courant alternatif en mode Panneau avant, vous devez utiliser le retard de déclenchement manuel par défaut ([Acquire] > Delay Man). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 667 Pour éviter tout accès à distance via le réseau local, vous pouvez connecter l’instrument der- rière un routeur. Vous pouvez également désactiver les différentes interfaces E/S des menus du panneau avant sous [Utility] > I/O Config. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 668 à la section Mesures). Par exemple, appuyez sur DCV et configurez les cordons de test comme indiqué. Étape 2 : Appuyez sur la touche [Acquire] du panneau avant, puis sur l’une des touches de fonction suivantes : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 669 HH.MM.SS. Pour utiliser l’heure de la journée, il faut que l’horloge en temps réel de l’instrument soit correctement réglée. Pour régler l’horloge, appuyez sur [Shift] > [Utility] > > System Setup.Date/Time. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 670 (voir l’étape 7 ci-dessous). Le nombre de mesures pouvant être enregistré en mémoire varie selon la quantité de mémoire de l’instrument. Avec l’option MEM, la limite est de 2 000 000 lectures. Sans l’option MEM, la limite est de 50 000 lectures. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 671 Metadata - Active le numéro de mesure, l’horodatage de la première lecture et l’intervalle d’échantillonnage (si celui-ci est disponible) dans le fichier. Separator - Indique le caractère (virgule, tabulation ou point-virgule) à utiliser pour séparer les Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 672 Le diagramme de tendance est particulièrement utile pour la visualisation des mesures d’enregistrement de données. Pour plus de détails, reportez-vous à la section Diagramme de tendance (Modes Numé- risation et Enregistrement de données). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 673 à distance. Pour empêcher l’accès à distance, vous pouvez décon- necter les câbles d’interface LAN, GPIB et USB de l’instrument avant de lancer les mesures. Pour éviter tout accès à distance via le réseau local, vous pouvez connecter l’instrument der- Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 674 Cependant, ces données ne sont mises à jour qu’une fois la numérisation terminée. Étapes détaillées de la numérisation Étape 1 : Pour numériser la tension continue, appuyez sur DCV et configurez les cordons de test comme indiqué. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 675 à 1 A : Étape 2 : Appuyez sur la touche [Acquire] du panneau avant, puis sur la touche de fonction Acquire : Appuyez sur la touche de fonction Digitize : Étape 3 : Le menu Numérisation s’ouvre : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 676 Ext - L’instrument se déclenche lorsqu’un bord de la pente appropriée parvient au connecteur Ext Trig du panneau arrière. Vous pouvez spécifier la pente dans le menu de touches de fonction qui apparaît lorsque Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 677 Les curseurs X et Y (indiqués ci-dessous), les curseurs de suivi et les commandes de pano- ramique/zoom vous permettent d’analyser les données numérisées. Pour plus de détails, reportez-vous à la section Diagramme de tendance (Modes Numérisation et Enregistrement de données). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 678 Caractéristiques et fonctions Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 679 Pour une latence minimale et une sensibilité maximale, utilisez une plage fixe lors de l’utilisation du déclenchement par niveau. Cependant, la précision du niveau de déclenchement est réduite lorsque le détecteur de matériel est utilisé. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 680 La précision du niveau de déclenchement et la latence augmentent et la sensibilité diminue car le temps de porte augmente. La plage de tension fixe élimine les incertitudes (en raison du changement de plage) en matière de latence de déclenchement. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 681 Résistance à 2 et 4 fils Opération mathématique préalable I CC I CA I CA , Crête, Opération mathématique préalable I CA Fréquence I CC , Fréquence, Opération mathé- matique préalable Fréquence Période Période, V CA, Opération mathématique préalable Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 682 4 secondes, le multimètre numérique procède à une mesure secondaire. Pour la mesure secondaire de crête, la touche de fonction Clear Peaks vous permet d’effacer l’historique cumulé de la fonction crête à crête (réinitialise l’enregistrement des valeurs de crête). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 683 Le menu (Trg Src) vous permet de sélectionner l’une des trois sources de déclenchement suivantes : Auto - L’instrument effectue des mesures en continu, et émet automatiquement un nouveau déclen- chement dès que la mesure est terminée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 684 [Single] permet de faire basculer l’instrument en mode de déclenchement unique. Dans les modes Unique et Ext, la touche [Run/Stop] permet d’arrêter les lectures en cours ou de basculer en mode Auto si les lectures sont arrêtées. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 685 Dans cette configuration, la fréquence d’échantillonnage est déterminante, car le début de chaque échantillon est déterminé par l’intervalle d’échantillonnage spécifié (le retard de déclenchement n’affecte que le début du premier échantillon). Le temps d’intégration et la Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 686 Pour Max, la limite correspond au nombre d’octets autorisés par le système de fichiers (2 4,294967296 Go). Cela représente environ 252 millions de mesures avec le paramètre Metadata désactivé ou 159 millions de mesures avec le paramètre Metadata activé. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 687 Separator - Indique le caractère (virgule, tabulation ou point-virgule) à utiliser pour séparer les infor- mations présentes sur chaque ligne. Une fois la configuration du stockage de lectures terminé, appuyez sur Done > Save Readings pour sau- vegarder les lectures en mémoire dans un fichier. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 688 Modification de la polarité de sortie VM Comp Modification du masque de chiffres Modification de l’échelle de l’histogramme, de la barre de mesure ou du diagramme de tendance Modification des préférences de l’utilisateur Exécution de l’autotest Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 689 à partir de l’interface de commande à distance. Cependant, vous pouvez effectuer une capture d’écran de l’affichage Probe Hold. Pour plus de détails, reportez-vous à la section Menu Utility (Utilitaire) - Manage Files (Gérer les fichiers). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 690 Scaling (Réglage d’échelle dB/dBm) Statistics (Statistiques) Limits (Limites) Ces fonctions mathématiques sont disponibles pour les modèles 34465A/70A : Null (Valeur de référence) Smoothing Filter (Filtre de lissage) Scaling (Réglage d’échelle) Statistics (Statistiques) Limits (Limites) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 691 à nouveau sur [Null] ou sur [Math] et réglez la première touche de fonction sur Off. Voir également Math (Opération mathématique) - Introduction dB/dBm Scaling (Réglage d’échelle dB/dBm) Statistics (Statistiques) Limits (Limites) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 692 La valeur relative doit être comprise entre -200 et +200 dBm (0 par défaut). Vous pouvez mesurer cette valeur en appuyant sur la touche Measure Ref Value ou bien saisir une valeur spécifiée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 693 600 (valeur par défaut), 800, 900, 1 000, 1 200 ou 8 000 Ω. Appuyez sur Ref R et utilisez les touches flé- chées vers le haut et bas du panneau avant pour effectuer votre sélection. Voir également Math (Opération mathématique) - Introduction Null (Valeur de référence) Statistics (Statistiques) Limits (Limites) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 694 Measure Ref Value ou bien utiliser la touche de fonction dB Ref Value pour saisir une valeur. Le réglage d’échelle dB concerne uniquement les fonctions de mesure en tension alternative et continue. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 695 (par exemple, RPM ou °C) pour les mesures Mx-B à l’échelle. Le réglage d’échelle Mx-B s’applique à toutes les fonctions de mesure, à l’exception de Rapport, Con- tinuité et Diode. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 696 Caractéristiques et fonctions Voir également Math (Opération mathématique) - Introduction Null (Valeur de référence) Statistics (Statistiques) Limits (Limites) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 697 ( Si toutes les mesures sont comprises dans la variation autorisée par rapport aux mesures présentes dans le filtre, l’icône de filtre est surlignée en vert ( Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 698 Appuyez sur Clear Readings pour effacer la mémoire de mesures et relancer le calcul des sta- tistiques. Voir également Math (Opération mathématique) - Introduction Null (Valeur de référence) dB/dBm Scaling (Réglage d’échelle dB/dBm) Limits (Limites) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 699 (elle active ou désactive également l’avertisseur sonore pour les autres fonctions qui utilisent ce dernier : maintien de la sonde, diode, continuité et erreurs). La touche Clear Condition redéfinit les marqueurs de limite au vert comme décrit ci-dessous. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 700 La bordure demeure rouge, même si la ligne de tendance quitte la zone limite. Si la ligne de tendance est comprise dans les limites, vous pouvez redéfinir les bordures au vert en appuyant sur la touche Clear Con- dition. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 701 Dans l’image ci-dessous, la bordure limite (gauche) inférieure est rouge, ce qui signifie que la limite infé- rieure a été dépassée. (La valeur affichée dans l’angle supérieur gauche (-01.68487 Vcc) est comprise dans les limites. Elle n’est donc pas affichée en rouge.) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 702 La couleur rouge clair (illustrée ci-dessous) indique que la mesure affichée dépasse les limites. L’écran Number indique également le nombre de fois que les limites ont été dépassées. Voir également Math (Opération mathématique) - Introduction Null (Valeur de référence) dB/dBm Scaling (Réglage d’échelle dB/dBm) Statistics (Statistiques) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 703 Pour de plus amples informations, reportez-vous à la section Mesures secondaires. Sélection de l’affichage Le fait d’appuyer sur la touche , puis sur la touche de fonction Display vous permet de sélec- tionner le type d’affichage : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 704 à la section Mode Numérisation. Pour plus d’informations sur l’un des types d’affichages suivants, cliquez sur un lien ci-dessous : Nombre Barre de mesure Trend Chart (Diagramme de tendance) (34461A/65A/70A uniquement) Histogramme Mesures secondaires Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 705 Caractéristiques et fonctions Nombre Par défaut, l’instrument affiche les mesures sous la forme d’un nombre : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 706 (Tension alternative), Peak (Crête) ou Pre-Math (Opération mathématique préalable). Lorsque la fonction V CA (Tension alternative) est sélectionnée comme mesure secondaire, l’écran affiche la mesure en ten- sion continue dans la partie supérieure et la mesure en tension alternative dans la partie inférieure : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 707 Caractéristiques et fonctions Pour de plus amples informations sur les mesures secondaires disponibles pour chaque fonction de mesure, reportez-vous à la section Mesures secondaires. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 708 La touche de fonction Auto indique que le nombre de chiffres affichés dépend d’autres paramètres spé- cifiques à la fonction, tels que l’ouverture de mesure, définie en nombres de cycles de tension d’alimentation (NPLC). Les mesures sont toujours arrondies ; elles ne sont jamais tronquées. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 709 Caractéristiques et fonctions Voir également Display (Affichage) - Introduction Barre de mesure Trend Chart (Diagramme de tendance) (34461A/65A/70A uniquement) Histogramme Mesures secondaires Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 710 Caractéristiques et fonctions Barre de mesure La barre de mesure (ci-dessous) ajoute une barre de déplacement sous l’affichage Number (Nombre) stan- dard. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 711 V CA (Tension alternative) est sélectionnée comme mesure secondaire, l’écran affiche la mesure en ten- sion continue sous forme numérique dans la partie supérieure, la tension continue dans la barre de mesure et la mesure en tension alternative au-dessus de cette barre : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 712 Pour de plus amples informations sur les mesures secondaires disponibles pour chaque fonction de mesure, reportez-vous à la section Mesures secondaires. Voir également Display (Affichage) - Introduction Nombre Trend Chart (Diagramme de tendance) (34461A/65A/70A uniquement) Histogramme Mesures secondaires Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 713 Time Window n’est disponible. La modification de ce paramètre entraîne l’effacement du dia- gramme de tendance, mais pas celui de la mémoire de mesures ou des données statistiques ou d’histogramme. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 714 Si la fréquence de lecture est inférieure au temps pour chaque colonne de pixels, certaines colonnes peuvent ne représenter aucune lecture. Dans ce cas, le diagramme de tendance se poursuit avec une ligne horizontale en travers du pixel : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 715 (Low) de 0 V et une limite élevée (High) de 5 V équivalent à une valeur centrale (Center) de 2,5 V et à un intervalle (Span) de 5 V. Si vous avez activé des limites, la touche de fonction (Limits) apparaît également. Cette dernière définit l’échelle verticale en fonction des limites. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 716 Caractéristiques et fonctions Voir également Display (Affichage) - Introduction Nombre Barre de mesure Histogramme Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 717 1 %, 2 %, 5 %, 10 %, 20 %, 50 % ou 100 %. Par exemple, le graphique ci-dessus affiche un zoom de 5 % alors que le graphique ci-dessous affiche le même signal avec un zoom de 100 % : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 718 Cursors - Affiche et contrôle les curseurs X1, X2, Y1, Y2 et de suivi (indiqués sous forme de lignes) sur le diagramme de tendance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 719 X Only – Affiche uniquement les curseurs Y1 et Y2. Dans ce mode, les touches de fonction suivantes sont disponibles : X1 - Utilisez les touches fléchées pour sélectionner le curseur Y1. Y2 - Utilisez les touches fléchées pour sélectionner le curseur Y2. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 720 Dans le graphique ci-dessus : Le curseur Y1 suit la position du curseur X1. Le curseur Y2 suit la position du curseur X2. Informations sur le temps de curseur et l’amplitude ∆X, ∆Y. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 721 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 722 Les commandes Zoom, Pan et Cursors ne sont pas disponibles lors de l’enregistrement des données dans des fichiers. Pour de plus amples informations, reportez-vous à la section Diagramme de tendance (Mode de mesure continue). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 723 Caractéristiques et fonctions Voir également Display (Affichage) - Introduction Diagramme de tendance (Mode de mesure continue) Nombre Barre de mesure Histogramme Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 724 Dans le cas d’un binning automatique (Binning Auto), l’algorithme commence par réajuster conti- nuellement l’étendue de l’histogramme sur les mesures reçues, entraînant ainsi un nouveau binning des données chaque fois qu’une nouvelle valeur est située en dehors de la plage actuelle. Après avoir acquis Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 725 Si les cases externes ne sont pas visibles, les données correspondantes ne sont pas représentées sur la ligne. La ligne de distribution cumulative est toujours comprise entre 0 et 100 % sur l’échelle ver- ticale, quelle que soit l’échelle de l’histogramme. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 726 # Bins 20 - Nombre total de cases. 0, 2.5k, 5k - (Sous l’histogramme) Plage de valeurs de fréquence. Curseurs (modèles 34465A/70A uniquement) Appuyez sur la touche de fonction Cursors (modèles 34465A/70A uniquement) pour afficher les curseurs de l’histogramme. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 727 Histogramme avec statistiques L’affichage des statistiques (Shift > Math > Statistics) est particulièrement utile pour l’affichage de l’histogramme. Par exemple, dans le graphique ci-dessous, la ligne bleue épaisse correspond à la Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 728 Caractéristiques et fonctions moyenne, chaque ligne bleue épaisse représentant un écart-type de la moyenne. Voir également Display (Affichage) - Introduction Nombre Barre de mesure Trend Chart (Diagramme de tendance) (34461A/65A/70A uniquement) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 729 Configurer les interfaces LAN (en option sur le modèle 34460A), GPIB (en option sur tous les modèles) et Exécuter des autotests et des fonctions d’administration de l’instrument Configurer la date et l’heure de l’instrument, ainsi que divers paramètres utilisateur Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 730 Une fois l’opération terminée, appuyez sur la touche de fonction Done. Étape 5 : Appuyez sur la touche de fonction Perform Autocal. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 731 Store State (illustrée ci-dessus) ou Store Pref (libellé de la touche de fonction en cas de sauvegarde des préférences) pour sauvegarder le fichier spécifié. La touche Action vous permet de sauvegarder un fichier ou de créer un nouveau dossier. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 732 Defined) ou de la configuration par défaut définie en usine (Factory Defaults). Set to Defaults (Rétablir les réglages par défaut) La touche Set to Defaults permet de charger les paramètres de configuration ou de préférences par défaut d’usine de l’instrument. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 733 à renommer, puis appuyez sur Select. Appuyez sur New Name, saisissez un nouveau nom et appuyez sur Done. Appuyez sur Perform Rename > Done. Browse (Parcourir) La touche Browse permet de sélectionner le fichier ou le dossier sur lequel l’action sera effectuée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 734 Previous Char et Next Char pour déplacer le curseur dans la zone où le nom est saisi. L’image ci-dessous n’illustre aucune touche de fonction Next Char car le curseur est en fin de ligne. Appuyez sur [Done] ou sur [Cancel] pour terminer. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 735 Elle permet également d’activer la configuration des paramètres réseau d’après le protocole (DHCP ou Manuelle) sélectionné. Par exemple, appuyez sur Host Name ou Service mDNS pour modifier le nom d’hôte de l’instrument ou le nom du service mDNS, illustré ci-dessous. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 736 Après avoir modifié l’état de l’interface, vous devez éteindre, puis rallumer l’instrument pour que la modi- fication soit prise en compte. Lorsque cette option est désactivée, l’interface ne peut pas être configurée par l’utilitaire Keysight IO Libraries Connection Expert. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 737 Pour utiliser la touche Easy File Access sur un ordinateur exécutant le système d’exploitation Windows XP, assurez-vous que Microsoft Windows Media Player 11 pour Windows XP SP2, SP3 ou une nouvelle version de Windows sont installés. Vous pouvez télécharger ce logiciel à l’adresse www.- microsoft.com/en-us/download/details.aspx?id=8163. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 738 GPIB Settings (Paramètres GPIB) (en option) La touche GPIB Settings permet d’activer ou de désactiver l’interface GPIB de votre instrument. Lorsque cette option est désactivée, l’interface ne peut pas être configurée par l’utilitaire Keysight IO Libraries Connection Expert. Vous pouvez également définir l’adresse GPIB à une valeur comprise entre 0 et 30. Après avoir activé ou désactivé...
  • Page 739 Calibrate (Étalonner) La touche Calibrate permet d’accéder à la procédure d’étalonnage de l’ínstrument. Pour de plus amples informations, reportez-vous à la section Étalonnage. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 740 La touche Install License active les fonctions sous licence de l’instrument. Pour de plus amples infor- mations sur l’obtention de licences, rendez-vous sur www.keysight.com/find/truevolt. Après avoir reçu un fichier de licence de Keysight, effectuez la procédure suivante pour installer la licence : 1.
  • Page 741 Mise à jour du microprogramme La touche Firmware Update met à jour le microprogramme de l’instrument vers une nouvelle version. Pour de plus amples informations, reportez-vous à la section Mise à jour du microprogramme. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 742 Permet également d’activer ou de désactiver la tonalité audible (Beeper On ou Off) associée aux fonc- tionnalités suivantes : Limits (Limites) : une limite de mesure est dépassée (si les limites sont activées) Probe Hold (Maintien de la sonde) : le signal mesuré enregistre une lecture stable Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 743 (Keysight), votre instrument continue de répondre avec le nom de fabri- cant « Agilent » et non « Keysight » tant que vous n’avez pas défini l’ID SCPI du panneau avant sur 34460A Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 744 SYST:IDEN DEF ou réinitialisé les préférences utilisateur. Une fois cette opération terminée, l’instrument répond avec le nom de fabricant « Keysight ». Important : Pour mettre à jour le microprogramme de l’instrument à partir de l’interface de commande à distance, le numéro de modèle de la réponse *IDN? doit correspondre à...
  • Page 745 Caractéristiques et fonctions Interface Web Les multimètres numériques Keysight série Truevolt intègrent une interface Web conçue pour surveiller et contrôler l’instrument à partir d’un navigateur Web. Pour commencer, connectez votre instrument au réseau local et saisissez l’adresse IP de l’instrument dans la barre d’adresse située en haut du navigateur Web de l’ordinateur, ou sélectionnez l’instrument dans Connection Expert et cliquez sur le bouton Ins-...
  • Page 746 En mode Allow Full Control (Autoriser le contrôle intégral), cette fenêtre permet de configurer l’instrument et d’effectuer des mesures. Sélectionnez simplement les paramètres du multimètre numé- rique, puis cliquez sur Apply. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 747 Get Display Image pour mettre à jour l’affichage graphique : Visualisation des données Pour les modes Observe Only et Allow Full Control, cette fenêtre maintient un affichage continu des mesures en mémoire (Readings in Memory) et active les commandes suivantes : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 748 à supprimer (Readings to Remove) et cliquez sur Get Data pour afficher ces mesures au bas de la fenêtre. Les mesures affichées SONT effacées de la mémoire de l’instrument. Delete All Readings - Supprime toutes les mesures de la mémoire de mesure de l’instrument et n’affiche aucune mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 749 En mode Allow Full Control, cette fenêtre vous permet d’envoyer des commandes au multimètre numé- rique de façon interactive et de lire les réponses. Vous pouvez utiliser ce dernier pour vous familiariser Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 750 La page de configuration vous permet de visualiser les paramètres de configuration réseau de l’instrument, qui permettent de configurer l’instrument afin qu’il fonctionne dans votre environnement. Pour modifier un paramètre, cliquez sur Modify Configuration. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 751 Caractéristiques et fonctions Aide Une aide est disponible pour chacun des onglets décrits ci-dessus. Par exemple : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 752 Didacticiel de mesure Didacticiel de mesure Pour un niveau de précision optimal sur les multimètres Keysight série Truevolt, suivez les recom- mandations fournies dans les sections ci-dessous. Métrologie Considérations relatives aux mesures en courant continu Réjection du bruit Considérations relatives aux mesures de résistance Mesures de valeurs efficaces vraies en courant alternatif Capacité...
  • Page 753 Didacticiel de mesure Considérations relatives aux mesures Pour atteindre un niveau de précision optimal sur les multimètres Keysight série Truevolt, vous devez éli- miner les erreurs de mesure potentielles. Ce chapitre décrit les erreurs courantes et préconise des méthodes pour les éviter.
  • Page 754 Métrologie Le multimètre numérique 7½ chiffres 34470A convient idéalement aux applications de métrologie de haute précision. Keysight prévoit de nombreuses notes d’application destinées à maximiser la précision de mesure, par exemple : Conseils pratiques pour l’optimisation de la précision de mesure avec votre multimètre numérique : cette...
  • Page 755 Pour limiter les effets des erreurs dues à la charge et minimiser le captage de bruit, réglez la résistance d’entrée du multimètre sur 10 GΩ (impédance élevée) pour les plages 100 mVcc, 1 Vcc et 10 Vcc. La résis- tance d’entrée est maintenue à 10 MΩ pour les plages 100 Vcc et 1 000 Vcc. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 756 LO d’entrée et la masse, comme illustré ci-dessous. Cela peut provoquer des erreurs lors de la mesure de tensions faibles qui flottent par rapport à la masse. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 757 à la masse. Si le multimètre doit être relié à la terre, connectez-le avec le dis- positif testé à un point de masse commun. Connectez également le multimètre et le dispositif testé à la même prise électrique chaque fois que possible. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 758 Pour éliminer les erreurs de décalage associées aux mesures de résistance en 2 fils des cordons de test, procédez comme suit : Court-circuitez simplement les cordons de test ensemble et relevez la résistance des cordons affichée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 759 Courant de test standard Courant de test en mode basse puissance 1 GΩ 500 nA 500 nA 100 MΩ 500 nA 500 nA 10 MΩ 500 nA 500 nA 1 MΩ 5 µA 5 µA 100 kΩ 10 µA 5 µA 10 kΩ 100 µA 10 µA 1 kΩ 1 mA 100 µA 100 Ω 1 mA 100 µA Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 760  Ω), comparés aux isolants PTFE Teflon (10  Ω). La résis- tance de fuite des isolants en nylon ou en PVC peut engendrer une erreur de 0,1 % lors de la mesure d’une résistance de 1 MΩ dans un environnement humide. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 761 à la bande passante efficace du multimètre. Notez que le Keysight série Truevolt utilise les mêmes techniques pour mesurer la tension et le courant efficaces vrais. La bande passante de tension alternative efficace est de 300 kHz, alors que la bande pas- sante de courant alternatif est de 10 kHz.
  • Page 762 En résumé, les erreurs de mesure de valeur efficace vraie se produisent lorsqu’une quantité d’énergie non négligeable est présente dans le signal d’entrée à des fréquences supérieures à la bande passante du mul- timètre. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 763 à f = 1/t Cela donne une impression immédiate du contenu haute fréquence en identifiant l’emplacement où se produit ce passage comme fonction du facteur de crête : f =(CF )(prf). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 764 3 V (c’est-à-dire, un facteur de crête 3) et une durée de 111 µs. Le PRF peut être établi à 1 000 Hz, comme suit : Ainsi, dans le tableau ci-dessus, ce signal de courant alternatif peut être mesuré avec une erreur sup- plémentaire de 0,18 pour cent. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 765 10 A provoque généralement des erreurs. Les bornes de mesure Hi et Lo ne sont pas utilisées pour de nombreuses mesures. L’application de signaux ici lorsque cela n’est pas nécessaire peut également engendrer des erreurs. Les tensions Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 766 à 20 °C pour une température mesurée de 100 °C. Pour un didacticiel détaillé sur les mesures de température, consultez la note d’application 290 de Key- sight intitulée Mesures de température pratiques, disponible à l’adresse www.keysight.com. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 767 (s) (lectures/s) partielle de lectures/s V CA I CA V CA I CA V CA/I CA V CA/I CA Lent 3 Hz 1,67 Moyen 20 Hz 0,63 0,25 Rapide 200 Hz 0,025 0,025 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 768 Désactivez le réglage automatique du zéro Désactivez la compensation de décalage (mesures de résistance) Pour plus d’informations sur les fonctions ci-dessus, reportez-vous au type de mesure en question dans la section Mesures. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 769 (Dt). Cette mesure est répétée à deux périodes différentes pendant la montée exponentielle qui se produit. Un algorithme sélectionne les données de ces quatre points, et en linéarisant cette montée expo- nentielle sur ces « ouvertures courtes », il calcule avec précision la valeur de capacité. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 770 Un conden- sateur à couche de haute qualité affiche généralement la valeur la plus basse et un condensateur élec- trolytique la valeur la plus élevée, avec habituellement des condensateurs céramiques entre les deux. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 771 Sample Interval. La figure ci-dessous illustre une onde sinusoïdale échantillonnée à une fréquence légèrement inférieure à 2F. Comme illustré par la ligne pointillée, le résultat est une fréquence de repliement considérablement Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 772 [Run/Stop]. La numérisation démarre lorsque l’événement de déclenchement indiqué se produit. Le mode Numérisation est uniquement disponible à partir du panneau avant et ne peut être utilisé que pour les fonctions de mesure V CC ou I CC. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 773 Le nombre maximal de mesures pouvant être capturées varie selon l’espace disponible de la mémoire volatile. La banque de lecture est effacée au début d’une nouvelle acquisition. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 774 Une fois l’acquisition terminée, le diagramme de tendance, les fonctions de pano- ramique/zoom et les curseurs peuvent être utilisés pour examiner les données. Les statistiques et les données de l’histogramme sont calculées une fois la numérisation terminée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 775 à distance. *Si l’accès est effectué à distance, l’instrument poursuit la numérisation ou l’enregistrement des données jusqu’à la fin et vous pouvez récupérer les lectures depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 776 Web de l’instrument ne permet pas de paramétrer l’instrument en mode de commande à dis- tance. *Si l’accès est effectué à distance, l’instrument poursuit la numérisation ou l’enregistrement des données jusqu’à la fin et vous pouvez récupérer les lectures depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 777 La durée maximale d’enregistrement de données est de 100 heures, la durée minimale est de 1 seconde. Par défaut, le mode de déclenchement par défaut est Automatique. Les sources de déclenchement par niveau et externe ne sont pas prises en charge pour l’enregistrement de données. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 778 être créés pour conserver les données, le deuxième nom de fichier portera le suffixe _ 0001, le troisième fichier portera le suffixe _00002, etc. En cas d’enregistrement des données dans des fichiers, le nombre maximal de mesures est de 100 heures x 1 000 lectures/s =360 000 000. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 779 Vous pouvez spécifier une virgule, un onglet ou un point-virgule comme délimiteur pour séparer les valeurs. Lorsque le paramètre Metadata est désactivé, seules les données de lecture sont enregistrées : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 780 Didacticiel de mesure Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 781 En effet, le nombre de lectures affichées par colonne de pixels varie selon la vitesse de lecture et la fenêtre de temps sélectionnée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 782 à distance. *Si l’accès est effectué à distance, l’instrument poursuit la numérisation ou l’enregistrement des données jusqu’à la fin et vous pouvez récupérer les lectures depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 783 Pour une latence minimale et une sensibilité maximale, utilisez une plage fixe lors de l’utilisation du déclenchement par niveau. Cependant, la précision du niveau de déclenchement est réduite lorsque le détecteur de matériel est utilisé. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 784 La précision du niveau de déclenchement et la latence augmentent et la sensibilité diminue car le temps de porte augmente. La plage de tension fixe élimine les incertitudes (en raison du changement de plage) en matière de latence de déclenchement. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 785 Fréquence d’entrée Résistance d’entrée (kΩ) 100 Hz 1 kHz 10 kHz 100 kHz 15,7 Pour les basses fréquences, l’erreur due à la charge est la suivante : En haute fréquence, l’erreur additionnelle due à la charge est la suivante : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 786 La chute de tension augmente avec la fréquence d’entrée. Certains circuits peuvent se mettre à osciller lors de la réalisation de mesures de courant en raison de l’inductance série du multimètre et de vos connexions de mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 787 0,06 % pour une entrée inverse de 100 V, 100 kHz. Vous pouvez utiliser les techniques de mise à la masse décrites dans les problèmes de mode commun de courant continu pour minimiser les tensions de mode commun de courant alternatif. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 788 Les signaux inutiles appliqués aux bornes de mesure Hi et Lo peuvent également provoquer des erreurs. Les tensions alternatives ou continues supérieures à 15 volts sur les bornes de mesure peuvent provoquer des erreurs de mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 789 En mode Numérisation, le temporisateur/l’intervalle d’échantillonnage, le NPLC et l’ouverture sont contrôlés de la même manière ; la modification d’un paramètre peut avoir un impact sur les autres paramètres. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 790 État à la mise sous tension/réinitialisation Informations connexes Bibliothèques d’E/S et pilotes de l’instrument Le logiciel Keysight IO Libraries Suite et les instructions d’installation respectives sont disponibles sur le CD-ROM Keysight Automation Ready CD-ROM fourni avec votre instrument (en option pour le modèle 34460A).
  • Page 791 Les paramètres facultatifs sont entourés de crochets ([ ]). Les crochets ne sont pas envoyés avec la chaîne de caractères des commandes. Si vous ne spécifiez pas de valeur pour un paramètre facultatif, l’instrument utilise une valeur par défaut. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 792 À la place d’une valeur donnée pour le paramètre <plage>, vous pouvez remplacer MIN pour définir la plage à sa valeur minimale, MAX pour définir la plage à sa valeur maximale ou DEF pour définir la plage à sa valeur par défaut. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 793 (*), sont composées de trois caractères et peuvent inclure un ou plusieurs paramètres. Le mot- clé d’une commande est séparé du premier paramètre par un espace. Utilisez un point virgule (;) pour séparer plusieurs commandes, comme indiqué ci-dessous : *RST; *CLS; *ESE 32; *OPC? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 794 « OFF » ou « 0 ». Pour une condition vraie, l’instrument accepte « ON » ou « 1 ». Lorsque vous effectuez une requête sur un paramètre logique, l’instrument renvoie tou- jours « 0 » ou « 1 ». La commande suivante requiert un paramètre logique : DISPlay:STATe {ON|1|OFF|0} Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 795 Une commande superposée, le cas échéant, sera terminée sans message « Operation Complete » (Opération terminée) (s’applique à INIT). L’utilisation de la commande ABORt est la méthode recommandée pour terminer une mesure. Commandes par sous-système ABORt CALCulate:CLEar[:IMMediate] Sous-système CALCulate:AVERage Sous-système CALCulate:LIMit Sous-système CALCulate:SCALe Sous-système CALCulate:SMOothing Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 796 Sous-système [SENSe:]CAPacitance Sous-système [SENSe:]CURRent Sous-système [SENSe:]DATA2 Sous-système [SENSe:]{FREQuency|PERiod} Sous-système [SENSe:]{RESistance|FRESistance} Sous-système [SENSe:]TEMPerature Sous-système [SENSe:]VOLTage Sous-système STATus Sous-système SYSTem - Commandes de base Sous-système SYSTem - Configuration des E/S Sous-système SYSTem:LICense Sous-système SYSTem:LOCK Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 797 Référence de programmation SCPI TEST:ALL? Sous-système TRIGger UNIT:TEMPerature Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 798 (aucun) (aucun) Interrompre une mesure en cours : ABORt Utilisez cette commande pour interrompre une mesure lorsque l’instrument attend un déclen- chement ou pour interrompre une mesure longue ou une série de mesures. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 799 (voir la section Présentation du système d’état). L’instrument efface toutes les mesures de la mémoire lorsque la configuration de mesure change ou lorsque l’une des commandes suivantes est exécutée : INITiate, MEASure:<fonction>?, READ?, *RST, SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 800 Pour récupérer les mesures de la mémoire de mesures, utilisez la commande FETCh?. Utilisez la com- mande DATA:REMove? pour lire et effacer la totalité ou une partie des mesures disponibles. Utilisez la commande ABORt pour renvoyer l’instrument à l’état inactif. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 801 INIT Sur le modèle 34460A, l’option 34460A-LAN ou l’option 3446LANU est requise. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 802 (voir la section Présentation du système d’état). L’instrument efface toutes les mesures de la mémoire lorsque la configuration de mesure change ou lorsque l’une des commandes suivantes est exécutée : INITiate, MEASure:<fonction>?, READ?, *RST, SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 803 (voir la section Présentation du système d’état). L’instrument efface toutes les mesures de la mémoire lorsque la configuration de mesure change ou lorsque l’une des commandes suivantes est exécutée : INITiate, MEASure:<fonction>?, READ?, *RST, SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 804 être endommagé par des tensions ou des courants élevés qui risqueraient de compromettre les fonctions de sécurité de l’instrument. Paramètre Renvoi type (aucun) FRON ou REAR Renvoyer l’état du commutateur Front/Rear : ROUT:TERM? Sur le modèle 34460A, cette requête renvoie toujours FRON. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 805 Si un ou plusieurs test échouent, la requête renvoie +1 et enregistre une erreur dans la file d’erreurs. Pour une liste complète des messages d’erreur relatifs aux échecs d’autotest, reportez-vous à la sec- tion Messages d'erreur d’autotest. Une fois le test terminé, l’instrument retourne à l’état précédant l’autotest. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 806 La commande accepte également CEL ou FAR, mais la requête renvoie C ou F. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Voir également CONFigure:TEMPerature MEASure:TEMPerature? [SENSe:]FUNCtion[:ON] Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 807 De plus, il réalise éventuellement les opérations mathématiques suivantes, comme indiqué dans le diagramme ci-dessous : Lissage Réglage d’échelle Statistiques Vérification des limites Histogramme Seuls les calculs activés par CALC:SCAL:STAT, CALC:TRAN:HIST:STAT, CALC:LIM:STAT CALC:AVER:STAT sont effectués. Composantes du sous-système CALCulate CALCulate:CLEar[:IMMediate] Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 808 Référence de programmation SCPI Sous-système AVERage Sous-système HISTogram Sous-système LIMit Sous-système SCALing Sous-système SMOothing Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 809 Effacer toutes les limites, données d’histogramme, statistiques et mesures : CALC:CLE:IMM Les éléments sont supprimés de façon synchrone ; les données d’histogramme, statistiques et de limite redémarrent toutes en même temps que les mesures. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 810 CONF:VOLT 10,.001 SAMP:COUN 100 CALC:LIM:LOW 3.2 CALC:LIM:UPP 3.4 CALC:LIM:STAT ON INIT *WAI STAT:QUES? Réponse type : +4096 (au moins une mesure était supérieure à la limite maximale) Récapitulatif des commandes CALCulate:LIMit:CLEar[:IMMediate] CALCulate:LIMit:{LOWer|UPPer}[:DATA] CALCulate:LIMit[:STATe] Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 811 12 dans le registre des données douteuses lorsque la fonction de mesure change ou lorsque l’une des commandes suivantes est exécutée : CALCulate:LIMit:STATe ON INITiate MEASure:<fonction>? READ? CALCulate:LIMit:CLEar *RST SYSTem:PRESet Pour effacer les statistiques, les limites, les données d’histogramme et les données de mesure, uti- lisez la commande CALCulate:CLEar[:IMMediate]. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 812 La commande CONFigure réinitialise les deux limites à 0. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 813 CALCulate:LIMit:STATe ON INITiate MEASure:<fonction>? READ? CALCulate:LIMit:CLEar *RST SYSTem:PRESet L’instrument désactive ce paramètre après un changement de la fonction de mesure ou l’exécution de la commande *RST ou SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 814 CALCulate:TRANsform:HISTogram:CLEar[:IMMediate] CALCulate:TRANsform:HISTogram:POINts CALCulate:TRANsform:HISTogram:RANGe:AUTO CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer} CALCulate:TRANsform:HISTogram[:STATe] INITiate[:IMMediate] MEASure:<fonction>? READ? *RST SYSTem:PRESet Récapitulatif des commandes CALCulate:TRANsform:HISTogram:ALL? CALCulate:TRANsform:HISTogram:CLEar[:IMMediate] CALCulate:TRANsform:HISTogram:COUNt? CALCulate:TRANsform:HISTogram:DATA? CALCulate:TRANsform:HISTogram:POINts CALCulate:TRANsform:HISTogram:RANGe:AUTO CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer} CALCulate:TRANsform:HISTogram[:STATe] Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 815 Le nombre de mesures supérieures à la valeur de plage supérieure Les valeurs de plage sont des nombres réels renvoyés au format +1.00000000E+00. Le nombre de mesures et de données de cases sont des entiers positifs signés renvoyés au format +100. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 816 CALCulate:TRANsform:HISTogram:RANGe:AUTO est activée. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 817 Les valeurs de plage inférieure et supérieure sont calculées automatiquement si la commande CALCu- late:TRANsform:HISTogram:RANGe:AUTO est activée. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 818 Paramètre Renvoi type {ON|1|OFF|0}. Valeur par défaut : OFF. 0 (OFF) ou 1 (ON) Voir l’ exemple L’instrument désactive ce paramètre après un changement de la fonction de mesure ou l’exécution de la commande *RST ou SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 819 L’instrument définit la valeur de référence à 0,0 avec la sélection automatique de référence activée après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou un changement de fonction de mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 820 à une résistance de référence (CALCulate:SCALe:DBM:REFerence), par rapport à 1 milliwatt. (dBm = 10 × log 10 (mesure 2 / résistance de référence / 1 mW)). Le réglage d’échelle dBm concerne uniquement les fonctions de mesure en tension alternative et continue. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 821 CALC:SCAL:OFFS 5 CALC:SCAL:STAT ON Concerne uniquement les modèles 34465A et 34470A. L’instrument définit la valeur de gain à 0.0 après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 822 L’instrument définit la valeur de référence à 0.0 avec la sélection automatique de référence activée après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument () ou un changement de fonction de mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 823 Le réglage d’échelle est paramétré sur OFF lorsque vous changez de fonction de mesure (par exemple, lors du passage d’une tension continue à une tension alternative). Vous devez réactiver le réglage d’échelle après avoir changé de fonction de mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 824 ( ° ). Si vous définissez le libellé de mesure à °C, °F ou K, cela n’a aucun effet sur les unités de température définies à l’aide de la commande UNIT:TEMPerature. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 825 Concerne uniquement les modèles 34465A et 34470A. La commande CALCulate:SCALe:UNIT spécifie la chaîne d’unités à afficher. L’instrument désactive l’affichage de la chaîne d’unités après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 826 Référence de programmation SCPI Sous-système CALCulate:AVERage Ce sous-système calcule les statistiques de mesure. Récapitulatif des commandes CALCulate:AVERage:ALL? CALCulate:AVERage:AVERage? CALCulate:AVERage:CLEar[:IMMediate] CALCulate:AVERage:COUNt? CALCulate:AVERage:MAXimum? CALCulate:AVERage:MINimum? CALCulate:AVERage:PTPeak? CALCulate:AVERage:SDEViation? CALCulate:AVERage[:STATe] Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 827 Les six autres requêtes répertoriées ci-dessus renvoient des valeurs individuelles. Paramètre Renvoi type (aucun) (voir ci-dessous) Renvoyer les statistiques de 100 mesures de fréquence : CONF:FREQ 1.0E3,.001 SAMP:COUN 100 CALC:AVER:STAT ON INIT *WAI CALC:AVER:ALL? Réponse type : +1.00520000E+03,+1.00512000E+03,+1.00527000E+03,+4.13500000E-01 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 828 MEASure:<fonction>? READ? *RST SYSTem:PRESet Toutes les valeurs sont renvoyées au format +1.23450000E+01. Lorsque la mise à l’échelle dB ou dBm est utilisée, les requêtes CALC:AVER:AVER et CALC:AVER:SDEV renvoient +9.91000000E+37 (et non un nombre). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 829 Active ou désactive le calcul des statistiques. Paramètre Renvoi type {ON|1|OFF|0}. Valeur par défaut : OFF. 0 (OFF) ou 1 (ON) Renvoyer les statistiques de 100 mesures de fréquence : CONF:FREQ 1.0E3,.001 SAMP:COUN 100 CALC:AVER:STAT ON INIT *WAI CALC:AVER:ALL? Réponse type : +1.00520000E+03,+1.00512000E+03,+1.00527000E+03,+4.13500000E-01 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 830 Les statistiques sont effacées lorsque la fonction de mesure change ou lorsque l’une des commandes suivantes est exécutée : CALCulate:AVERage:STATe ON CALCulate:AVERage:CLEar INITiate MEASure:<fonction>? READ? *RST SYSTem:PRESet L’instrument désactive ce paramètre après un changement de la fonction de mesure ou l’exécution de la commande *RST ou SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 831 à la moyenne flottante des 10, 50 ou 100 dernières mesures. Une pondération uniforme est appli- quée à toutes les mesures de la moyenne. L’instrument définit le délai de réponse sur FAST après une réinitialisation de la configuration d’usine (*RST) ou après un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 832 L’instrument désactive le filtre (à moyenne flottante) de lissage après une réinitialisation de la confi- guration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 833 Les données du diagramme de tendance doivent être désactivées (OFF) pour effectuer des mesures à la vitesse maximale (50 kHz). Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 834 Renvoi type (aucun) 0 (réussite) ou 1 (échec) Étalonner le CAN : CAL:ADC? Première étape de la séquence de réétalonnage, cette action doit être effectuée avant de procéder à un étalonnage du décalage. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 835 Vous pouvez exécuter cette requête, que la sécurité de l’instrument soit ou non verrouillée. Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension ou l’exécution de la com- mande *RST ou SYSTem:PRESet. Voir également SYSTem:SECurity:COUNt? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 836 À la livraison de l’usine, l’instrument est verrouillé par le code de sécurité défini sur AT3446XA. Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension ou l’exécution de la com- mande *RST ou SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 837 Vous devez déverrouiller l’instrument pour l’étalonner ou modifier la chaîne de caractères ou le code de sécurité d’étalonnage. Paramètre Renvoi type (aucun) (aucun) Enregistrer les constantes d’étalonnage en mémoire non volatile : CAL:STOR Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 838 Vous pouvez exécuter cette requête, que la sécurité de l’instrument soit ou non verrouillée. Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension ou l’exécution de la com- mande *RST ou SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 839 Numérique, 0,0 par défaut +2.37000000E-02 Spécifier la valeur d’étalonnage 0.0237 : CAL:VAL 2.37E-2 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 840 1 échantillon Nombre de déclenchements 1 déclenchement Retard de déclenchement Retard automatique Source de déclenchement Immédiate Pente de déclenchement NEGative Fonctions mathématiques Désactivées. Les autres paramètres ne changent pas. État Null par fonction Désactivé Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 841 à l’aide de la commande INITiate et la sau- vegarde de la mesure dans la mémoire de mesures. La plage de 10 kΩ est sélectionnée avec une réso- lution de 100 Ω. CONF:RES 10000,100 INIT FETC? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 842 Renvoie une chaîne de caractères entre guillemets indiquant la fonction, la plage et la résolution actuelles. La forme abrégée du nom de la fonction (CURR:AC, FREQ) est toujours renvoyée. Paramètre Renvoi type (aucun) "VOLT +1.00000000E+01,+3.00000000E-06" Renvoyer la fonction, la plage et la résolution actuelles : CONF? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 843 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Utilisez la commande READ? INITiate pour lancer la mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 844 Au-delà de 1,2 kΩ, l’instrument affiche le message « OPEN » sans émettre de signal sonore. Les requêtes FETCh?, READ? MEASure:CONTinuity? renvoient la résistance mesurée, quelle que soit sa valeur. Utilisez la commande READ? INITiate pour lancer la mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 845 DEFault pour la <résolution> ou omettez complètement la <résolution>. La commutation automatique de plage descend d’une plage à moins de 10 % de la plage et augmente d’une plage à plus de 120 % de la plage. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 846 Si le signal dépasse 5,05 V, le panneau avant affiche « OPEN » et la valeur ren- voyée par SCPI est 9.9E37. Les requêtes FETCh?, READ? MEASure:DIODe? renvoient la tension mesurée, quelle que soit sa valeur. Utilisez la commande READ? INITiate pour lancer la mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 847 Sans objet 100 ppm × <plage> (MAXimum) 1 ms 100 ppm × <plage> (MAXimum) 10 ppm × <plage> (DEFault) 10 ms 10 ppm × <plage> (DEFault) 1 ppm × <plage> 100 ms 1 ppm × <plage> (MINimum) 0,1 ppm × <plage> (MINimum) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 848 Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. La commutation automatique de plage peut être activée pour la tension d’entrée. Utilisez la commande READ? INITiate pour lancer la mesure. Voir également CONFigure? MEASure:{FREQuency|PERiod}? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 849 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Utilisez la commande READ? INITiate pour lancer la mesure. Voir également CONFigure? MEASure:{RESistance|FRESistance}? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 850 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Utilisez la commande READ? INITiate pour lancer la mesure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 851 DEFault pour la <résolution> ou omettez complètement la <résolution>. La commutation automatique de plage descend d’une plage à moins de 10 % de la plage et augmente d’une plage à plus de 120 % de la plage. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 852 Pour contrôler la vitesse des mesures en courant alternatif, modifiez le retard de déclenchement ou la bande passante du filtre de courant alternatif. Utilisez la commande READ? INITiate pour lancer la mesure. Voir également CONFigure? MEASure[:VOLTage]:{AC|DC}? [SENSe:]CURRent:AC:BANDwidth Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 853 DEFault pour la <résolution> ou omettez complètement la <résolution>. La commutation automatique de plage descend d’une plage à moins de 10 % de la plage et augmente d’une plage à plus de 120 % de la plage. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 854 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Utilisez la commande READ? INITiate pour lancer la mesure. Voir également CONFigure? MEASure[:VOLTage][:DC]:RATio? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 855 Vous pouvez enregistrer jusqu’à 1 000 mesures dans la mémoire de lecture du modèle 34460A, 10 000 mesures dans celle du modèle 34461A, 50 000 mesures sur le modèle 34465A/70A (sans l’option MEM) ou 2 000 000 mesures sur le modèle 34465A/70A (avec l’option MEM). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 856 à la section Présentation du sous-système STATus. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 857 Aucune erreur n’est générée. Toutefois, le bit Reading Mem Ovfl (bit 14) est défini dans le registre de conditions du registre des données douteuses (voir la section Présentation du système d’état). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 858 Z), des chiffres (0-9) et des caractères spéciaux comme @, %, *, etc. gress..." Valeur par défaut : "" Afficher un message à l’écran : DISP:TEXT "Test in progress..." L’envoi de la commande DISPlay:TEXT <chaîne> entraîne l’affichage du texte, même avec DISP:STAT OFF. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 859 L’écran n’est pas activé si son état est OFF. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 860 Définit le format de données sur ASCII ou REAL. Affecte le format de données des commandes MEASure?, READ?, FETCh?, DATA:REMove? uniquement. Paramètre Renvoi type <longueur> ASCii : "9" ou REAL : "64" ASC,9 Spécifier le format de données REAL 64 bits : FORM:DATA REAL,64 ou FORM:DATA REAL Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 861 REAL au format de bloc de longueur définie IEEE 488.2. Le paramètre <longueur> est facul- tatif, mais « 64 » est la valeur par défaut et la seule valeur autorisée. L’ordre des octets est contrôlé par la commande FORMat:BORDer. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 862 (PNG ou BMP). HCOPy:SDUMp:DATA:FORMat {PNG|BMP} HCOPy:SDUMp:DATA:FORMat? Spécifie le format des images renvoyées par la commande HCOPy:SDUMp:DATA?. Paramètre Renvoi type {PNG|BMP}. Valeur par défaut : PNG PNG ou BMP Voir l’ exemple Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 863 Commandes courantes IEEE 488.2 Ces commandes et requêtes sont associées à la norme IEEE 488.2. Récapitulatif des commandes *CAL? *CLS *ESE *ESR? *IDN? *LRN? *OPC *OPC? *OPT? *PSC *RCL *RST *SAV *SRE *STB? *TRG *TST? *WAI Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 864 (Request for Service - RQS). Les bits doivent être activés avec la com- *SRE mande Récapitulatif du Un ou plusieurs bits sont définis dans le registre de fonctionnement standard (les bits doivent fonctionnement STATus:OPERation:ENABle être activés, voir standard Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 865 Commande d’effacement de l’état. Efface les registres des événements dans tous les groupes de registres. Efface également la file d’erreurs. Paramètre Renvoi type (aucun) (aucun) Effacer les bits du registre des événements et la file d’erreurs : *CLS Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 866 Le paramètre <activation> est non volatile. Il ne change pas après une réinitialisation de la confi- guration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet), un préréglage d’état (STATus:PRESet) ou un effacement d’état (*CLS). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 867 *ESE. Lorsqu’un bit est activé, il reste défini jusqu’à son effacement en lisant le registre des événements ou en envoyant la commande *CLS (effacement d’état). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 868 - Révision du microprogramme de la carte de mesure gg.gg - Révision FPGA de la carte de mesure bb - Révision de la carte de mesure fp - Révision de la carte du panneau avant Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 869 - Révision du microprogramme de la carte de mesure gg.gg - Révision FPGA de la carte de mesure bb - Révision de la carte de mesure fp - Révision de la carte du panneau avant Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 870 Les autres commandes peuvent être exécutées avant que le bit Opération terminée ne soit posi- tionné. *OPC diffère de la commande *OPC? en ce sens qu’elle active un bit d’état et génère la valeur « 1 » lorsque l’opération en cours est terminée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 871 La valeur 0 est renvoyée pour les options qui ne sont pas installées. Par exemple, un échantillon 34461A avec l’option GBIP installée renvoie GPB,0. Un modèle 34460A avec uniquement les options GPB et SEC installées renvoie GPB,0,SEC. Les options varient en fonction du modèle, comme indiqué ici. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 872 Les paramètres d’activation d’état de l’instrument sont stockés dans la mémoire volatile. Ils ne changent pas lorsque l’instrument est remis sous tension, après réinitialisation de la configuration d’usine (*RST) ou après un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 873 à la mise sous tension. Par exemple, la commande *PSC 0 conserve le contenu du registre d’activation lorsque l’instrument est éteint puis rallumé. La commande *RST n’efface pas le registre d’activation de l’octet d’état. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 874 Après avoir défini la source de déclenchement, vous devez envoyer la commande INITiate pour placer le multimètre à l’état « en attente de déclenchement ». La commande *TRG n’est acceptée que si le multimètre numérique est à l’état « en attente de déclenchement ». Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 875 Patienter jusqu’à la fin de toutes les opérations en attente : *WAI Les changements de configuration et les mesures sont considérés comme des opérations en attente. Par conséquent, la commande *WAI provoque l’application de ces changements. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 876 Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension, une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Ce paramètre est défini sur la valeur par défaut lorsque l’instrument est expédié de l’usine et après l’exécution de la commande SYSTem:SECurity:IMMediate. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 877 éventuel de « (<N>) ». La lettre N est un entier ajouté au besoin afin que le nom soit unique. Si nécessaire, le nom désiré peut être tronqué pour faire de la place à l’entier ajouté. Paramètre Renvoi type (aucun) "Multimètre numérique Keysight 34461A - 00123" Renvoyer le nom du service mDNS résolu : LXI:MDNS:SNAMe:RESolved? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 878 Si l’interface LAN ou des services LAN spécifiques (VXI-11, sockets, etc.) sont désactivés par la com- mande SYSTem:COMMunicate:ENABle, vous devez réactiver séparément l’interface ou les services, et remettre l’instrument sous tension pour que le réseau local soit opérationnel. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 879 1 échantillon Nombre de déclenchements 1 déclenchement Retard de déclenchement Retard automatique Source de déclenchement Immédiate Pente de déclenchement NEGative Fonctions mathématiques Désactivées. Les autres paramètres ne changent pas. État Null par fonction Désactivé Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 880 L’exemple suivant illustre la configuration de l’instrument pour des mesures de résistance en 2 fils, le déclenchement de l’instrument pour effectuer une mesure et la lecture de la mesure. La plage 1 kΩ est sélectionnée avec une résolution de 0,1 Ω. MEAS:RES? 1000,0.1 Réponse type : +3.27150000E+02 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 881 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 882 Entre 10 Ω et 1,2 kΩ, l’instrument affiche la mesure de résistance réelle sans aucun signal sonore. Au-delà de 1,2 kΩ, l’instrument affiche le message « OPEN » sans émettre de signal sonore. Les requêtes FETCh?, READ? MEASure:CONTinuity? renvoient la résistance mesurée, quelle que soit sa valeur. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 883 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 884 Si le signal dépasse 5,05 V, le panneau avant affiche « OPEN » et la valeur ren- voyée par SCPI est 9.9E37. Les requêtes FETCh?, READ? MEASure:DIODe? renvoient la tension mesurée, quelle que soit sa valeur. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 885 Le signal d’entrée pour les mesures de fréquence ou de période inclut un composant de tension alter- native. Par défaut, cette commande utilise la commutation automatique de plage pour sélectionner la plage de tension. Si aucun signal n’est appliqué, la valeur 0 est renvoyée. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 886 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 887 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 888 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 889 Si le signal d’entrée est supérieur au signal pouvant être mesuré dans la plage manuelle spécifiée, l’instrument affiche l’indication Overload (Surcharge) sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 890 Les commandes de gestion de fichiers de base du sous-système MMEMory sont répertoriées ci-dessous. Le sous-système MMEMory inclut également deux autres types de commandes : Gestion des fichiers STATe et PREFerence Transfert de données Commandes et requêtes MMEMory:CATalog[:ALL]? MMEMory:CDIRectory MMEMory:COPY MMEMory:DELete MMEMory:MDIRectory MMEMory:MOVE MMEMory:RDIRectory Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 891 Les noms de dossiers et de fichiers ne doivent pas contenir les caractères suivants : \ / : * ? " < > | La combinaison du nom de dossier et de fichier ne doit pas dépasser 240 caractères. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 892 S’il n’y a pas de fichier, seules les valeurs <mém_utilisée> et <mém_disponible> sont renvoyées. L’instrument utilisant une faible quantité d’espace dans le système de fichiers flash à des fins d’utilisation interne, la valeur <mém_utilisée> n’est jamais égale à zéro. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 893 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Vous pouvez supprimer uniquement un dossier vide. Sinon, l’instrument génère l’erreur « Directory not empty » (Répertoire non vide). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 894 Le paramètre <spéc_fichier> peut inclure le caractère générique * : *.bmp, *.sta, abc*.*, etc. Le dossier spécifié doit exister et ne peut pas être marqué comme un dossier masqué ou système. La commande SYSTem:SECurity:IMMediate permet de supprimer et de nettoyer tous les fichiers. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 895 Pour renommer un fichier, spécifiez le même dossier pour <fichier1> et <fichier2>. Pour déplacer un fichier dans un fichier du même nom dans un autre dossier, vous pouvez spécifier uniquement <lecteur>:<chemin> pour <fichier2>. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 896 En règle générale, les fichiers de configuration enregistrent les paramètres volatiles asso- ciés aux mesures. Les préférences sont des paramètres non volatiles associés à l’instrument, mais à aucune mesure spécifique. Le tableau suivant récapitule les informations contenues dans chaque fichier. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 897 Réglage du décalage de la jonction de référence à thermocouple Le sous-système MMEMory inclut également deux autres types de commandes : Gestion de fichiers de base Transfert de données Récapitulatif des commandes MMEMory:LOAD:PREFerences Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 898 Les noms de dossiers et de fichiers ne doivent pas contenir les caractères suivants : \ / : * ? " < > | La combinaison du nom de dossier et de fichier ne doit pas dépasser 240 caractères. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 899 Pour une liste des informations contenues dans chaque fichier, reportez-vous au tableau Fichier de configuration/Fichier de préférences. Le dossier spécifié doit exister et ne peut pas être marqué comme un dossier masqué ou système. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 900 Le fichier de configuration créé par la commande *SAV 0, appelé STATE_0.sta, est situé dans le dos- sier racine du système de fichiers flash interne. Il est remplacé par l’état d’extinction de l’instrument lorsque l’alimentation est rétablie. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 901 Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension, une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 902 Le fichier spécifié ne peut pas être vide ni marqué comme un fichier masqué ou système. Utilisez cette requête avant d’envoyer la commande *RCL MMEMory:LOAD:STATe pour déterminer si un état a été enregistré dans le fichier. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 903 Le sous-système MMEMory inclut également deux autres types de commandes : Gestion de fichiers de base Gestion des fichiers STATe et PREFerence Récapitulatif des commandes MMEMory:DOWNload:DATA MMEMory:DOWNload:FNAMe MMEMory:FORMat:READing:CSEParator MMEMory:FORMat:READing:INFormation MMEMory:FORMat:READing:RLIMit MMEMory:STORe:DATA MMEMory:UPLoad? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 904 Les noms de dossiers et de fichiers ne doivent pas contenir les caractères suivants : \ / : * ? " < > | La combinaison du nom de dossier et de fichier ne doit pas dépasser 240 caractères. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 905 Spécifie le caractère utilisé pour séparer les colonnes dans les données stockées. Paramètre Renvoi type {COMMa|SEMicolon|TAB}. Valeur par TAB, COMM or SEM défaut : COMMa. Spécifier des onglets comme séparateur de colonnes : MMEMory:FORMat:READing:CSEParator TAB Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 906 34465A/34470A. Cette fonction vous permet de prendre en charge des applications d’analyse de feuilles de calcul, de bases de données et de données communes limitées à 1 million de lignes par fichier. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 907 Tout nom de fichier valide Bloc IEEE 488.2 de longueur définie Télécharger le fichier de configuration « Myfile.sta » situé dans le répertoire racine du système de fichiers flash interne vers l’ordinateur hôte : MMEM:UPL? "INT:\Myfile.sta" Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 908 (TRIGger:COUNt), qui définit le nombre de déclenchements à accepter avant de renvoyer l’état de déclenchement « inactif ». Le nombre total de mesures renvoyées correspond au produit du nombre d’échantillons et du nombre de déclenchements. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 909 VOLT:AC:RANG 100 SAMP:COUN 10000 SAMP:COUN:PRET 5000 TRIG:SOUR INT TRIG:LEV 0.75 INIT Concerne uniquement les modèles 34465A et 34470A. Le pré-déclenchement s’applique à toutes les fonctions de mesure, à l’exception du test CONTinuity et DIODe. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 910 à 0. L’instrument définit la valeur de décalage à 0 après une réinitialisation de la configuration d’usine (commande*RST) ou un préréglage de l’instrument (commande SYSTem:PRESet). Voir également SAMPle:COUNt Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 911 TIMer - Le premier échantillon commence avec un temps de retard après le déclenchement. Cepen- dant, le deuxième échantillon commence à l’intervalle d’échantillonnage suivant le début du premier échantillon, etc. : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 912 Le premier échantillon est prélevé avec un temps de retard après le déclenchement (le retard actuel- lement défini avec la commande TRIGger:DELay). Le deuxième échantillon commence à l’intervalle d’échantillonnage suivant le début du premier échantillon, etc., comme indiqué ci-dessous : Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 913 Vous devez configurer entièrement la mesure avant de définir le tem- porisateur d’échantillonnage sur MIN, sous peine de générer une erreur. Une configuration complète inclut des éléments, tels que les statistiques mathématiques ou le réglage d’échelle. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 914 3 600 secondes. Cependant, la valeur est arrondie au pas le plus proche. Pour les mesures en courant continu, la taille du pas est de 1 µs. Pour les mesures en courant alternatif, elle dépend de la bande passante de courant alternatif. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 915 Référence de programmation SCPI L’instrument définit le temporisateur d’échantillonnage à 1 seconde après une réinitialisation de la configuration d’usine (commande *RST) ou un préréglage de l’instrument (commande SYSTem:PRESet). Voir également TRIGger:COUNt TRIGger:DELay TRIGger:SOURce SAMPle:COUNt SAMPle:SOURce Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 916 (commandes CALC:SCAL:STAT, CALC:LIM:STAT, CALC:TRAN:HIST:STAT, CALC:AVER:STAT CALC:TCH:STAT désactivées). Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 917 [SENSe:]CAPa- citance:NULL:VALue. L’instrument désactive la fonction de mesure par rapport à une référence après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 918 Pour utiliser la valeur de référence, activez l’état Null à l’aide de la commande [Sen- se:]CAPacitance:NULL[:STATe]. Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 919 à une référence AUTO, utilisez la fonction de mesure par rapport à une référence AUTO. Effectuez deux mesures, puis envoyez-les au tampon de sortie de l’instrument : CONF:CAP CAP:RANG:AUTO ON CAP:NULL ON CAP:NULL:VAL:AUTO ON SAMP:COUN 2 READ? Réponse type : +3.01963841E-10,+3.01566277E-10 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 920 Overload sur le panneau avant et renvoie « 9.9E37 » depuis l’interface de commande à distance. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 921 Utilisez la commande [SENSe:]DATA2? pour récupérer la mesure secondaire. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 922 Sous-système [SENSe:]CURRent Ce sous-système configure les mesures en courant alternatif et en courant continu. Récapitulatif des commandes [SENSe:]CURRent:AC:BANDwidth [SENSe:]CURRent:{AC|DC}:NULL[:STATe] [SENSe:]CURRent:{AC|DC}:NULL:VALue [SENSe:]CURRent:{AC|DC}:NULL:VALue:AUTO [SENSe:]CURRent:{AC|DC}:RANGe [SENSe:]CURRent:{AC|DC}:RANGe:AUTO [SENSe:]CURRent:{AC|DC}:TERMinals [SENSe:]CURRent:AC:SECondary [SENSe:]CURRent[:DC]:APERture [SENSe:]CURRent[:DC]:APERture:ENABled [SENSe:]CURRent[:DC]:NPLC [SENSe:]CURRent[:DC]:RESolution [SENSe:]CURRent[:DC]:SECondary [SENSe:]CURRent[:DC]:ZERO:AUTO [SENSe:]CURRent:SWITch:MODE Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 923 20 Hz - 300 kHz (moyen) 0,25 s/mesure 200 Hz - 300 kHz (rapide) 0,025 s/mesure Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 924 [SENSe:]CURRent: {AC|DC}:NULL:VALue. L’instrument désactive la fonction de mesure par rapport à une référence après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 925 Pour utiliser la valeur de référence, l’état Null doit être activé ([SENSe:]CURRent: {AC|DC}:NULL:STATe ON). Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 926 à une référence est activée ([SENSe:]CURRent:{AC|DC}:NULL:STATe ON). Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 927 à la plage de 1 A produira une résolution de 10 µA. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 928 à utiliser la plage de 10 A, mais n’affecte pas les valeurs de la com- mande [SENSe:]CURRent:{AC|DC}:RANGe ou [SENSe:]CURRent:{AC|DC}:RANGe:AUTO. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 929 1 µA sur la plage de 1 A, le passage aux bornes de 10 A produira une résolution de 10 µA. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 930 (Sans l’option DIG) 200 µs à 1 s (résolution de 2 µs), valeur par défaut : 100 ms. +1.00000000E-01 (Avec l’option DIG) 20 µs à 1 s (résolution de 2 µs), valeur par défaut : 100 ms. Activer le mode temps d’ouverture, définir le temps d’ouverture à 300 ms : CURR:APER:ENAB ON CURR:DC:APER 300E-03 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 931 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 932 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Voir également [SENSe:]CURRent[:DC]:RESolution Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 933 à un temps d’intégration représentant un nombre entier de cycles de tension d’alimentation. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Voir également [SENSe:]CURRent[:DC]:NPLC Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 934 "PTPeak" - Valeurs crête à crête, maximale et minimale du signal d’entrée. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 935 à 1 PLC à l’aide de ces commandes. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 936 Le mode de commutation s’applique uniquement aux bornes de courant 3 A. Les bornes 10A sont tou- jours effectivement en mode FAST. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 937 L’instrument efface toutes les mesures de la mémoire lorsque la configuration de mesure change ou lorsque l’une des commandes suivantes est exécutée : INITiate, MEASure:<fonction>?, READ?, *RST, SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 938 Ultérieurement, effacez les données de mesure secondaires crête à crête. Effectuez et récupérez des mesures. DATA2:CLE READ?;DATA2? Réponse type : +1.29177114E-03;+6.07221096E-04,+3.43677658E-03,+2.82955548E-03 (Données renvoyées dans l’ordre suivant : Lecture de tension continue, Crête min., Crête max., Crête à crête) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 939 Référence de programmation SCPI Sous-système [SENSe:]{FREQuency|PERiod} Ce sous-système configure les mesures de fréquence et de période. Récapitulatif des commandes [SENSe:]{FREQuency|PERiod}:APERture [SENSe:]{FREQuency|PERiod}:NULL[:STATe] [SENSe:]{FREQuency|PERiod}:NULL:VALue [SENSe:]{FREQuency|PERiod}:NULL:VALue:AUTO [SENSe:]{FREQuency|PERiod}:RANGe:LOWer [SENSe:]{FREQuency|PERiod}:TIMeout:AUTO [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe:AUTO [SENSe:]FREQuency:SECondary [SENSe:]PERiod:SECondary Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 940 100 ms 1 ppm × <plage> (MINimum) 0,1 ppm × <plage> (MINimum) Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 941 [SENSe:]{FREQuen- cy|PERiod}:NULL:VALue. L’instrument désactive la fonction de mesure par rapport à une référence après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 942 Les paramètres de mesure par rapport à une référence sont distincts pour les mesures de fréquence et de période. Paramètre Renvoi type {ON|1|OFF|0}. Valeur par défaut : OFF. 0 (OFF) ou 1 (ON) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 943 à une référence est activée ([SENSe:]{FREQuency|PERiod}:NULL:STATe ON). Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 944 20 Hz - 300 kHz (moyen) 0,25 s/mesure 200 Hz - 300 kHz (rapide) 0,025 s/mesure Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 945 Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension ou l’exécution de la com- mande *RST ou SYSTem:PRESet. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 946 à distance. La commutation automatique de plage peut être activée pour la tension d’entrée. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 947 La sélection d’une plage fixe ([SENSe:]<fonction>:RANGe) entraîne la désactivation de la com- mutation automatique de plage. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 948 « VOLTage:AC » - Mesure en tension alternative du signal d’entrée. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 949 Référence de programmation SCPI Sous-système [SENSe:]{RESistance|FRESistance} Ce sous-système configure les mesures de résistance en deux et quatre fils. Récapitulatif des commandes [SENSe:]{RESistance|FRESistance}:APERture [SENSe:]{RESistance|FRESistance}:APERture:ENABled [SENSe:]{RESistance|FRESistance}:NPLC [SENSe:]{RESistance|FRESistance}:NULL[:STATe] [SENSe:]{RESistance|FRESistance}:NULL:VALue [SENSe:]{RESistance|FRESistance}:NULL:VALue:AUTO [SENSe:]{RESistance|FRESistance}:OCOMpensated [SENSe:]{RESistance|FRESistance}:POWer:LIMit[:STATe] [SENSe:]{RESistance|FRESistance}:RANGe [SENSe:]{RESistance|FRESistance}:RANGe:AUTO [SENSe:]{RESistance|FRESistance}:RESolution [SENSe:]{FRESistance|RESistance}:SECondary [SENSe:]RESistance:ZERO:AUTO Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 950 [SENSe:]<fonction>:APERture:ENABled sur 1 (Activé) ; l’envoi de la commande [SENSe:]<fonction>:NPLC entraîne son réglage sur 0 (Désactivé). Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 951 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 952 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 953 [SENSe:]{RESis- tance|FRESistance}:NULL:VALue. L’instrument désactive la fonction de mesure par rapport à une référence après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 954 Pour utiliser la valeur de référence, l’état Null doit être activé ([SENSe:]{RESis- tance|FRESistance}:NULL:STATe ON). Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 955 à une référence est activée ([SENSe:]{RESistance|FRESistance}:NULL:STATe ON). Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 956 Cette méthode ne s’applique qu’aux mesures de résistances en 2 et 4 fils sur les plages 100 Ω, 1 kΩ, 10 kΩ et 100 kΩ. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 957 RES:RANG 10E3 READ? Réponse type : +6.27530000E+03 La plage de 1 GΩ concerne uniquement les modèles 34465A et 34470A. La sélection d’une plage fixe ([SENSe:]<fonction>:RANGe) entraîne la désactivation de la com- mutation automatique de plage. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 958 La sélection d’une plage fixe ([SENSe:]<fonction>:RANGe) entraîne la désactivation de la com- mutation automatique de plage. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 959 à un temps d’intégration représentant un nombre entier de cycles de tension d’alimentation. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Voir également [SENSe:]{RESistance|FRESistance}:NPLC Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 960 Utilisez la commande [SENSe:]DATA2? pour récupérer la mesure secondaire. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 961 à 1 PLC à l’aide de ces com- mandes. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 962 Référence de programmation SCPI Sous-système [SENSe:]TEMPerature Ce sous-système configure les mesures de température. Récapitulatif des commandes [SENSe:]TEMPerature:APERture [SENSe:]TEMPerature:APERture:ENABled [SENSe:]TEMPerature:NPLC [SENSe:]TEMPerature:NULL[:STATe] [SENSe:]TEMPerature:NULL:VALue [SENSe:]TEMPerature:NULL:VALue:AUTO [SENSe:]TEMPerature:SECondary [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:OCOMpensated [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe] [SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence] [SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:POWer:LIMit[:STATe] [SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:TYPE [SENSe:]TEMPerature:TRANsducer:TCouple:CHECk [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:OFFSet:ADJust [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:TYPE [SENSe:]TEMPerature:TRANsducer:TCouple:TYPE [SENSe:]TEMPerature:TRANsducer:TYPE [SENSe:]TEMPerature:ZERO:AUTO Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 963 TEMP:APER:ENAB ON TEMP:APER 300E-03 Concerne uniquement les modèles 34465A et 34470A. Les commandes CONFigure:TEMPerature, MEASure:TEMPerature? [SENSe:]TEMPerature:NPLC désactivent le mode temps d’ouverture et sélectionnent un temps d’intégration en nombre de cycles de tension d’alimentation. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 964 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 965 Pour utiliser la valeur de référence, l’état Null doit être activé ([SENSe:]TEMPerature:NULL:STATe ON). Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 966 à une référence est activée ([SENSe:]TEMPerature:NULL:STATe ON). Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 967 Utilisez la commande READ? INITiate pour lancer la mesure. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 968 Concerne uniquement les mesures de résistance en 2 et 4 fils. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 969 CONF:TEMP FRTD TEMP:TRAN:FRTD:RES 100.1 SAMP:COUN 2 READ? Réponse type : +1.04530000E+02,+1.04570000E+02 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 970 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 971 à +20 °C. Activez la vérification des thermocouples. Effectuez et lisez deux mesures : CONF:TEMP TC,K TEMP:TRAN:TC:RJUN:TYPE FIX TEMP:TRAN:TC:RJUN 20.0 TEMP:TRAN:TC:CHEC ON SAMP:COUN 2 READ? Réponse type : +3.31785468E+00,+3.61554515E+00 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 972 (voir la commande UNIT:TEMPerature). Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 973 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 974 Vous pouvez lire la température de jonction de référence interne à l’aide la commande SYSTem:TEMPerature. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 975 THERmistor ou FTHermistor importe peu, car les deux commandes concernent le même para- mètre. Paramètre Renvoi type 5000 (seul choix disponible) +5000 Configurez les mesures de thermistance en 4 fils. Effectuez et lisez deux mesures : CONF:TEMP FTH TEMP:TRAN:FTH:TYPE 5000 SAMP:COUN 2 READ? Réponse type : +1.04530000E+02,+1.04570000E+02 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 976 à 1 PLC à l’aide de ces commandes. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 977 Ce sous-système configure les mesures en tension alternative, en tension continue et du rapport de ten- sion continue. Récapitulatif des commandes [SENSe:]VOLTage:AC:BANDwidth [SENSe:]VOLTage:{AC|DC}:NULL[:STATe] [SENSe:]VOLTage:{AC|DC}:NULL:VALue [SENSe:]VOLTage:{AC|DC}:NULL:VALue:AUTO [SENSe:]VOLTage:{AC|DC}:RANGe [SENSe:]VOLTage:{AC|DC}:RANGe:AUTO [SENSe:]VOLTage:AC:SECondary [SENSe:]VOLTage[:DC]:APERture [SENSe:]VOLTage[:DC]:APERture:ENABled [SENSe:]VOLTage[:DC]:IMPedance:AUTO [SENSe:]VOLTage[:DC]:NPLC [SENSe:]VOLTage[:DC]:RATio:SECondary [SENSe:]VOLTage[:DC]:RESolution [SENSe:]VOLTage[:DC]:SECondary [SENSe:]VOLTage[:DC]:ZERO:AUTO Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 978 20 Hz - 300 kHz (moyen) 0,6250 s/mesure 200 Hz - 300 kHz (rapide) 0,0250 s/mesure Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 979 L’instrument désactive la fonction de mesure par rapport à une référence après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 980 La fonction de mesure par rapport à une référence n’est pas disponible pour les mesures du rapport de tension continue. Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 981 La fonction de mesure par rapport à une référence n’est pas disponible pour les mesures du rapport de tension continue. Ce paramètre est défini sur sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST), un préréglage de l’instrument (SYSTem:PRESet) ou l’exécution de la fonction CONFigure. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 982 à distance. L’instrument est défini sur la plage par défaut et la commutation automatique est activée ([SENSe:]VOLTage:{AC|DC}:RANGe:AUTO ON), après un rétablissement des réglages d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 983 La sélection d’une plage fixe ([SENSe:]<fonction>:RANGe) entraîne la désactivation de la com- mutation automatique de plage. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 984 (Sans l’option DIG) 200 µs à 1 s (résolution de 2 µs), valeur par défaut : 100 ms. +1.00000000E-01 (Avec l’option DIG) 20 µs à 1 s (résolution de 2 µs), valeur par défaut : 100 ms. Activer le mode temps d’ouverture, définir le temps d’ouverture à 300 ms : VOLT:APER:ENAB ON VOLT:DC:APER 300E-03 Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 985 Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 986 10 MΩ. Les commandes CONFigure et MEASure?sélectionnent automatiquement « AUTO OFF ». Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 987 Voir [SENSe:]VOLTage [:DC]:RESolution. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 988 à un temps d’intégration représentant un nombre entier de cycles de tension d’alimentation. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 989 "PTPeak" - Valeurs crête à crête, maximale et minimale du signal d’entrée. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 990 à 1 PLC à l’aide de ces commandes. Ce paramètre est défini à sa valeur par défaut après une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 991 Référence de programmation SCPI Sous-système STATus Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 992 Dépassement de capa- 16384 La mémoire de mesures est saturée. Une ou plusieurs mesures (les plus anciennes) ont cité de la mémoire été perdues. Non utilisé 32768 (Réservé à une utilisation ultérieure) Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 993 Activé si la file d’erreurs d’une interface de commande à distance comporte une erreur ; sinon, il est effacé. Non utilisé 16384 (Réservé à une utilisation ultérieure) Non utilisé 32768 La valeur « 0 » est renvoyée. Récapitulatif des commandes STATus:OPERation:CONDition? STATus:OPERation:ENABle STATus:OPERation[:EVENt]? STATus:PRESet STATus:QUEStionable:CONDition? STATus:QUEStionable:ENABle STATus:QUEStionable[:EVENt]? Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 994 (*RST), un préréglage de l’instrument (SYSTem:PRESet), un préréglage d’état (STATus:PRESet) ou un effacement d’état (*CLS). La commande *PSC détermine si le registre d’activation est effacé à la mise sous tension. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 995 Lire le registre de conditions (bit 12 actif) : STAT:QUES:COND? Les bits du registre de conditions reflètent l’état actuel. Si une condition n’est pas remplie, le bit cor- respondant est effacé dans le registre de conditions. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 996 Lire le registre des événements (bit 10 actif) : STAT:QUES? Lorsqu’un bit est activé, il reste défini jusqu’à son effacement en lisant le registre des événements ou en envoyant la commande *CLS (effacement d’état). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 997 à distance. Récapitulatif des commandes SYSTem:ACALibration:DATE? SYSTem:ACALibration:TEMPerature? SYSTem:ACALibration:TIME? SYSTem:BEEPer[:IMMediate] SYSTem:BEEPer:STATe SYSTem:CLICk:STATe SYSTem:DATE SYSTem:ERRor[:NEXT]? SYSTem:HELP? SYSTem:IDENtify SYSTem:LABel SYSTem:LFRequency? SYSTem:LOCal SYSTem:PRESet SYSTem:SECurity:COUNt? SYSTem:SECurity:IMMediate SYSTem:TEMPerature? SYSTem:TIME SYSTem:UPTime? SYSTem:VERSion? SYSTem:WMESsage Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 998 Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension ou l’exécution de la com- mande *RST ou SYSTem:PRESet. Utilisez la commande *CAL? pour effectuer un étalonnage automatique. Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 999 Les signaux sonores programmés peuvent être utiles pour le développement et la mise au point de programmes. Vous pouvez émettre un signal sonore à l’aide de cette commande, quel que soit l’état actuel de l’avertisseur sonore (SYSTem:BEEPer:STATe). Guide d’utilisation et de maintenance Keysight série Truevolt...
  • Page 1000 N’affecte pas l’avertisseur sonore associé aux tests de diodes et de continuité, aux erreurs, etc. Ce paramètre est non volatile ; il n’est pas altéré par une remise sous tension, une réinitialisation de la configuration d’usine (*RST) ou un préréglage de l’instrument (SYSTem:PRESet). Guide d’utilisation et de maintenance Keysight série Truevolt...

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a