SCPI Programming Reference
CALCulate:LIMit[:STATe] {ON|1|OFF|0}
CALCulate:LIMit[:STATe]?
Enables or disables limit testing.
Parameter
{ON|1|OFF|0}. Default: OFF. 0 (OFF) or 1 (ON)
See
.
Example
The instrument clears front panel indications of limits being exceeded and clears bits 11 and 12 in the
l
Questionable Data Register when the measurement function changes, or when any of the following
commands is executed:
CALCulate:LIMit:STATe ON
INITiate
MEASure:<function>?
READ?
CALCulate:LIMit:CLEar
*RST
SYSTem:PRESet
The instrument turns this setting OFF when the measurement function is changed or after
l
SYSTem:PRESet.
216
Typical Return
Keysight Truevolt Series Operating and Service Guide
*RST