Calibration Adjustment Procedures
DC Current Gain Calibration
This calibration procedure applies to the 34465A and 34470A only.
Review
Test Considerations
Configuration: DC Current
1. Configure each function and range in the order shown in the table below.
2. Apply the input signal shown in the Input column.
3. Enter the actual applied input current as the Cal Value (see
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4. Repeat steps 1 through 3 for each gain calibration point shown in the table.
5. Store the new calibration constants (see
Each range takes less than 4 seconds to complete.
Instrument Settings
Input
Function
1 mA
10 mA
100 mA
DC Current
1 A
2 A
566
and
Gain Calibration Considerations
Storing the Calibration
Range
1 mA
10 mA
100 mA
1 A
3 A
before beginning this procedure.
Entering Calibration
Constants).
Next Procedure
Keysight Truevolt Series Operating and Service Guide
Values).A Calibration