SCPI Programming Reference
TRIGger:DELay:AUTO {ON|1|OFF|0}
TRIGger:DELay:AUTO?
Disables or enables automatic trigger delay. If enabled, the instrument determines the delay based on
function, range, and integration time or bandwidth.
Parameter
{ON|1|OFF|0}. Default: ON.
Return 5 DC voltage measurements, with an automatic delay between each measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL 2
READ?
Typical Response:
+4.27230000E+00,+4.27150000E+00,+4.27190000E+00,+4.27170000E+00,+4.27200000E+00
Selecting a specific trigger delay using
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
TRIGger:LEVel {<level>|MIN|MAX|DEF}
TRIGger:LEVel? [{MIN|MAX|DEF}]
Sets the level on which a trigger occurs when level triggering is enabled
INTernal).
It is recommended that all triggered measurements be made using an appropriate
fixed manual range. That is, turn autorange off
OFF), or set a fixed range using the SENSe:<function>:RANGe, CONFigure, or
MEASure
Parameter
<level> (see bullet points below). Default: 0.
Configures dc voltage measurements, 1 V range, 0.1 mV resolution. Select internal trigger source. Set
the trigger level to 0.75 volts, positive slope. INIT puts the instrument in the "wait for trigger" mode.
CONF:VOLT:DC 1,0.0001
TRIG:SOUR INT
TRIG:LEV 0.75
TRIG:SLOP POS
INIT
Applies only to the 34465A and 34470A, and requires the DIG option.
l
Supported only for ac and dc voltage, ac and dc current, frequency, period, 2-wire resistance, 4-wire
l
resistance, RTD and thermistor temperature measurements. Resistance and temperature meas-
urements must have offset compensation and low power disabled.
430
TRIGger:DELay
command.
Typical Return
0 (OFF) or 1 (ON)
disables the automatic trigger delay.
(TRIGger:SOURce
(SENSe:<function>:RANGe:AUTO
Typical Return
+2.00000000E+01
Keysight Truevolt Series Operating and Service Guide