Additional Digitize Default Settings
On entering digitize mode, the DMM configures these settings:
If trigger source is set to manual it is changed to auto. (external and level remain as is.)
l
Limits mode is turned off.
l
Scaling is turned off.
l
Statistic and histogram are put in post-processing mode (computed after digitize is complete).
l
On the selected function (DCV or DCI) and for the new function if changed:
l
Autorange is turned off.
l
Autozero is turned off.
l
NPLC and aperture are set to their minimum values.
l
If trigger source is external or level, the pretrigger count is set to the digitize pretrigger count setting
l
(default of 0).
Samples per trigger is set according to the digitize duration (time or samples).
l
Sample timer is put in timer (not immediate) mode and Sample timer is set according to the digitize
l
sample rate or sample interval.
Trend Chart mode is changed to bucketized when data logging to a file.
l
On return to Continuous mode, settings are left as done in data log or digitize mode except:
l
Sample source is set to immediate.
l
Pretrigger count is set to 0.
l
Samples per trigger is set to 1.
l
Trigger count is set to infinite.
l