If a trigger occurs during a measurement, that measurement is considered to be a pretrigger sample.
l
After setting the sample count and pretrigger sample count, you must place the DMM in the "wait-
l
for-trigger" state using
for-trigger" state.
The pretrigger sample count must be less than the sample count set with the
l
mand, and there must be at least one post-trigger sample:
Pretrigger Sample Count ≤ Sample Count - 1
If any of the calculation functions are on
l
samples.
CONFigure
and
MEASure?
l
The instrument sets the pretrigger sample count to "0" after a Factory Reset ( *RST command) or a
l
Preset (SYSTem:PRESet command).
See Also
SAMPle:COUNt
SAMPle:SOURce {IMMediate|TIMer}
SAMPle:SOURce?
Determines sample timing, with
sample count is greater than one.
It is recommended that all triggered measurements be made using an appropriate
fixed manual range. That is, turn autorange off
OFF), or set a fixed range using the SENSe:<function>:RANGe, CONFigure, or
MEASure
Parameter
{IMMediate|TIMer}. Default: IMMediate.
Set sample source to timer. The INIT command sets the meter to the "wait-for-trigger" state. When a trig-
ger is received from the selected trigger source, the instrument starts taking periodic samples after the
trigger delay.
SAMP:SOUR TIM
SAMP:TIM 0.1
INIT
Operation
In all cases, the first sample is taken one trigger delay time after the trigger (the delay being set with the
TRIGger:DELay
command). Beyond that, the timing mechanism depends on whether you select IMMe-
diate (default) or TIMer as the source:
IMMediate - The first sample starts one trigger delay time after the trigger, and then the trigger delay
l
time is inserted between the first and second samples, the second and third, and so on:
308
INITiate
or
READ?
An external trigger is buffered until the DMM is in the "wait-
(CALC:STAT
set the pretrigger sample count to "0".
TRIGger:DELay
and
command.
ON), the pretrigger count is limited to 10,000
SAMPle:TIMer
to determine sample timing, when the
(SENSe:<function>:RANGe:AUTO
Typical Return
IMM or TIM