Performance Verification
Input Function
Open
DC Current
Short
DC Volts
Short 4-wire Ohms
Short 2-wire Ohms 100 MΩ
[1] Select 10 PLC.
[2] These values are typical and these ranges are calibrated internally.
Q
: Quick performance verification test point.
510
Values for 34470A DMM
Quick
[1]
Range
24 hours 90 days 1 year 2 years
Check
1 µA
10 µA
100 µA
Q
1 mA
± 0.030 µA
10 mA
100 mA
± 4 µA
1 A
± 60 µA
3 A
10 A
100 mV
Q
± 3 µV
1 V
10 V
100 V
1000 V
100 Ω
± 3 mΩ
1 kΩ
10 kΩ
Q
100 kΩ
1 MΩ
10 MΩ
Error from Nominal
[2]
± 50 pA (50E-12 A)
± 200 pA (± 200.0E-12 A)
[2]
± 0.001 µA
± 0.050 µA
± 2 µA
± 5 µA
± 100 µA
± 600 µA
± 1 ma
± 3.5 µV
± 4 µV
± 20 µV
± 600 µV
± 6 mV
± 4 mΩ
± 5 mΩ
± 50 mΩ
± 500 mΩ
± 5 Ω
± 100 Ω
1 kΩ
Keysight Truevolt Series Operating and Service Guide
[2]