[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe] {ON|1|OFF|0}
[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe]?
Enables or disables low-power, 2-wire or 4-wire, RTD measurements.
This parameter is common to 2-wire and 4-wire RTD measurements. Using the FRTD version of this command
or query is identical to using the RTD version.
Parameter
{ON|1|OFF|0}. Default: OFF.
Configure offset-compensated 2-wire RTD measurements. Make and read two measurements:
CONF:TEMP RTD
TEMP:TRANS:RTD:POW:LIM ON
TEMP:TRANS:RTD:OCOM ON
SAMP:COUN 2
READ?
Typical Response: +1.05451008E+01,+1.07391062E+01
Applies only to the 34465A and 34470A.
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence] {<reference>|MIN|MAX|DEF}
[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence]? [{MIN|MAX|DEF}]
Selects the nominal resistance (R 0 ) for 2-wire and 4-wire RTD measurements. R 0 is the nominal res-
istance of an RTD at 0 °C.
Parameter
34460A/61A: 80 Ω to 120 Ω. Default: 100 Ω.
34465A/70A: 49 Ω to 2100 Ω. Default: 100 Ω.
Configure 4-wire RTD measurements using an RTD with an R
urements:
CONF:TEMP FRTD
TEMP:TRAN:FRTD:RES 100.1
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
Typical Return
0 (OFF) or 1 (ON)
Typical Return
+1.00100000E+02
of 100.1 Ω. Make and read two meas-
0
SCPI Programming Reference
365