Télécharger Imprimer la page

Keysight Truevolt Serie Mode D'emploi page 359

Masquer les pouces Voir aussi pour Truevolt Serie:

Publicité

Les langues disponibles

Les langues disponibles

SCPI Programming Reference
[SENSe:]TEMPerature:NPLC {<PLC>|MIN|MAX|DEF}
[SENSe:]TEMPerature:NPLC? [{MIN|MAX|DEF}]
Sets the integration time in number of power line cycles (PLC) for temperature measurements. Integ-
ration time is the period that the instrument's analog-to-digital (A/D) converter samples the input signal
for a measurement. A longer integration time gives better measurement resolution but slower meas-
urement speed.
0.02, 0.2, 1, 10, 100. Default: 10 (34460A/61A)
0.02, 0.06, 0.2, 1, 10, 100. Default: 10 (34465A/70A without
DIG option)
0.001, 0.002, 0.006, 0.02, 0.06, .2, 1, 10, 100. Default: 10
(34465A/70A with DIG option)
See
Range, Resolution and NPLC
Configure 4-wire RTD measurements using a 10 PLC integration time. Then make and read one meas-
urement:
CONF:TEMP FRTD
TEMP:NPLC 10
READ?
Typical Response: +6.27530000E+01
Selecting an integration time in power line cycles overrides an aperture set with the
l
ature:APERture
command.
Only integration times of 1, 10, or 100 PLC provide normal mode (line frequency noise) rejection.
l
Setting the integration time also sets the measurement
l
tionship between integration time and resolution.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
360
Parameter
for more information.
Typical Return
+1.00000000E+01
resolution. Resolution Table
Keysight Truevolt Series Operating and Service Guide
[SENSe:]TEMPer-
shows the rela-

Publicité

loading

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a