[SENSe:]CURRent:{AC|DC}:RANGe:AUTO {OFF|ON|ONCE}
[SENSe:]CURRent:{AC|DC}:RANGe:AUTO?
Disables or enables autoranging for AC or DC current measurements. Autoranging is convenient because
it automatically selects the range for each measurement based on the input signal.
Specifying ONCE performs an immediate autorange and then turns autoranging off.
This parameter setting is not shared between AC and DC measurements. The parameters are independent for
AC and DC measurements.
On the 34461A, 34465A, or 34470A, use the
which current terminals to use for the measurement.
Parameter
{ON|1|OFF|0}. Default: ON.
Configure AC current measurements and perform an immediate autorange. Make and read two meas-
urements:
CONF:CURR:AC
CURR:AC:RANG:AUTO ONCE
SAMP:COUN 2
READ?
Typical Response: +1.04530000E-01,+1.04570000E-01
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
With autoranging enabled, the instrument selects the range based on the input signal.
l
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
l
Selecting the 10 A terminal with
l
current measurements to use the 10 A range, but it does not affect the values of
{AC|DC}:RANGe
or [SENSe:]CURRent:{AC|DC}:RANGe:AUTO.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
[SENSe:]CURRent:{AC|DC}:TERMinals
0 (OFF) or 1 (ON)
[SENSe:]CURRent:{AC|DC}:TERMinals
SCPI Programming Reference
command to specify
Typical Return
forces the specified type of
[SENSe:]CURRent:
325