SYSTem:ERRor[:NEXT]?
Reads and clears one error from the error queue. See
message list.
Parameter
(none)
Read and clear the first error in the error queue:
SYST:ERR?
Up to 20 errors can be stored in the instrument's error queue. Each remote interface I/O session
l
(GPIB, USB, VXI-11, Telnet/Sockets) has its own interface-specific error queue. Errors appear in the
error queue of the I/O session that caused the error. For example, if an error was generated by a com-
mand sent over GPIB, send
instrument hardware are broadcast to all I/O sessions.
Error retrieval is first-in-first-out (FIFO), and errors are cleared as you read them. The instrument
l
beeps each time an error is generated (unless disabled by
If more than 20 errors have occurred, the most recent error stored in the queue is replaced with -
l
350,"Queue overflow". No additional errors are stored until you remove errors from the queue. If no
errors have occurred when you read the error queue, the instrument responds with +0,"No error".
Errors have the following format (the error string may contain up to 255 characters).
l
<error code>,<error string>
Where:
<error code> = a positive or negative integer
<error string> = a quoted ASCII string up to 255 characters
See Also
*SRE
Keysight Truevolt Series Operating and Service Guide
Typical Return
-113,"Undefined header"
SYSTem:ERRor?
from GPIB to read the error queue. Errors generated by
SCPI Programming Reference
SCPI Error Messages
for the complete SCPI error
SYSTem:BEEPer:STATe
OFF).